Evaluate the Possibility of MT Pulse at 3T CE-TOF-MRA in Patients with Cerebral Infarction

뇌경색 환자의 3Tesla CE-TOF-MRA에서 MT 펄스의 유용성

  • Bae, Sung-Jin (Dept. of Radiology, Dongsan hospital, Keimyung University)
  • 배성진 (계명대학교 동산병원 영상의학과)
  • Published : 2007.09.29

Abstract

The purpose of this study was to evaluate the possibility of utilizing MT pulse at CE-TOF-MRA in patients with cerebral infarction. MRA using time-of-flight(TOF) technique with varying offset frequencies (0, 600, 1,200, and 1,800 Hz) magnetization transfer were performed in 10 patients with cerebral infarction at 3.0T MR scanner. CE-TOF-MRA and TOF-SPGR in normal vessel shown decreased SNR and increased CNR. The highest CNR in narrowing vessel shown at CE-TOF-MRA using 600 and 1,200 Hz offset frequencies. CNR in stenosis vessel increased dependent on using offset frequencies. The occlusion was clearly shown, and the highest CNR in occlusion shown at CE-TOF-MRA using 1,800 Hz offset frequencies. There was no shape variation in narrowing vessel or no visualizing vessel.

뇌경색환자들에 있어서 CE-TOF-MRA에서 MT 펄스사용의 유용성을 알아보고자 하였다. 2007년 5월에서 6월까지 뇌경색 증상 발현 후 최소 4시간에서 최대 5일 이내에 환자 10명을 대상으로 3Tesla 자기공명영상기기를 사용하여 조영제 주입 후 offset frequency 0, 600, 1,200, 1,800 Hz의 MT 펄스를 사용한 CE-TOF-MRA을 획득하였다. TOF-SPGR기법인 CE-TOF-MRA영상에서 정상혈관의 경우 SNR이 감소되었지만 CNR은 증가를 보였다. 좁아진 혈관부위는 MT 펄스 600 Hz, 1,200 Hz에서 최고의 CNR을 보였고, 협착된 혈관에서는 MT 펄스 강도와 비례하여 CNR이 증가 하였다. 혈관이 폐색된 경우에는 폐색범위가 명확하게 나타났고 MT 펄스 1,800 Hz에서 최고의 CNR로 평가되었다. 좁아지거나 보이지 않는 혈관에서는 형태변화가 없었고, 혈관이 폐색되어 평가가 불가능한 혈관에서 MT 펄스 후 혈관과 배후조직 SNR은 다소 감소하였지만 두 조직간 CNR은 증가되었다.