Hexokinase 방법을 이용한 Heparin 혈장과 NaF 혈장 시료의 혈당 분석치 비교

Comparison of Li-heparinized Versus Na-fluorinated Plasma for Routine Blood Glucose Determination Using Hexokinase Procedure

  • 문인경 (제주한라대학 임상병리과)
  • Moon, In-Kyung (Department of Clinical Pathology, CheJu Halla College)
  • 발행 : 2007.04.30

초록

In this study, the two sampling methods for blood glucose measurement were evaluated in order to reduce pre-analytical error. For this purpose, glucose levels of Li-heparinized plasma (LHP) and Na-fluorinated plasma (NFP) in blood collected from healthy volunteers were determined by using routine hexokinase procedures. The percentage range of pre-analytical error was quantitated by comparing LHP glucose values with NFP glucose values according to delaying analytical time (0, 60, and 120 min). LHP glucose values were decreased by 6.6% in 60 min., 17.7% in 120 min, whereas NFP glucose values decreased by 1.1% in 60 min, 2.0% in 120 min. Therefore it may be recommended that the NFP sampling method should be used for routine blood glucose determination in diabetes mellitus diagnosis.

키워드