Quality Assurance on Dose Distribution of Ir-192 Line Source

Ir-192 선 선원의 선량분포에 관한 품질보증

  • Kim, Jong-Eon (Department of Biomedical Engineering, College of Biomedical Sciences and Engineering, Inje University)
  • 김종언 (인제대학교 의용공학과)
  • Published : 2007.03.31

Abstract

The propose of this study is a verification of the correct calculation of the dose around source and the prescription dose of Ir-192 source in the plato treatment planning system. The source and orthogonal coordinates for lateral direction and those for the anterior posterior direction were drawn on a A4 paper and then input into the system. The prescription dose was prescribed to two points with radius 1 cm in the direction of polar angle $90^{\circ} and $270^{\circ} from the center of the source. The doses of prescription point and dose points acquired from the treatment planning system were compared with those from manual calculation using the geometry function formalism derived by Paul King et al. In this analysis, the doses of prescription point were exactly consistent with each other and those of dose points were obtained within the error point of 1.85%. And the system of accuracy was evaluated within 2% of tolerance error. Therefore, this manual dose calculation used for the geometry function formalism is considered to be useful in clinics due to its convenience and high quality assurance.

이 연구는 플라토 치료계획시스템의 치료계획에서 Ir-192 선원에 대한 처방점의 처방선량과 선원 주위의 선량분포상의 선량점들의 선량이 정확하게 계산되는지를 확인하는데 그 목적이 있다. 선원의 중심축의 전후방향에서의 평면의 직교좌표계와 측면방향에서의 평면의 직교좌표계 및 선원을 A4 용지 위에 그려서 치료계획시스템에 입력하였다. 처방선량은 선원중심으로부터 극각 $90^{\circ}, $270^{\circ}의 방향으로 반경 1 cm인 두 지점에 400 cGy를 처방하였다. 처방점과 선량점들의 선량은 치료계획시스템에서 출력된 선량과 파울 킹 등이 유도한 기하학 함수식으로 계산된 선량을 분석하였다. 본 실험의 분석에서 처방 점의 선량은 오차 없이 정확하게 일치하였고, 선량 점들의 선량은 1.85% 이내의 오차를 얻었다. 그리고 플라토 치료계획시스템의 선량계산은 허용오차 ${\pm}2%$ 범위 이내의 정확성으로 분석되었다. 파울 킹 등이 유도한 기하학 함수식을 사용하여 손으로 계산한 선량은 높은 정확성의 품질보증과 편리성에 기인하여, 임상에서 사용하는데 유용할 것으로 생각된다.

Keywords