Development of 3-Dimensional Polyimide-based Neural Probe with Improved Mechanical Stiffness and Double-side Recording Sites

증가된 기계적 강도 및 양방향 신호 검출이 가능한 3차원 폴리이미드 기반 뉴럴 프로브 개발

  • Published : 2007.11.01

Abstract

A flexible but implantable polyimide-based neural implant was fabricated for reliable and stable long-term monitoring of neural activities from brain. The developed neural implant provides 3-dimensional (3D) $3{\times}3$ structure, avoids any hand handling, and makes the insertion more efficient and reliable. Any film curvature caused by residual stress was not observed in the electrode. The 3D flexible polyimide electrode penetrated a dense gel whose stiffness is close to live brain tissue, because a ${\sim}1{\mu}m$ thick nickel was electroplated along the edge of the shank in order to improve the stiffness. The recording sites were positioned at both side of the shank to increase the probability of recording neural signals from a target volume of tissue. Impedance remained stable over 72 hours because of extremely low moisture uptake in the polyimide dielectric layers. At electrical recording test in vitro, the fabricated electrode showed excellent recording performance, suggesting that this electrode has the potential for great recording from neuron firing and long-term implant performance.

Keywords

References

  1. Y. Hanein, C. Schabmueller, G. Holman, P. Lucke, D. Denton and K. Bohringer, 'High aspect ratio submicrometer needles for intracellular applications', J. Micromech. Microeng. vol. 13, pp. 91-95, 2003 https://doi.org/10.1088/0960-1317/13/4/315
  2. L. Lin and A. Pisano, 'Silicon processed microneedles', IEEE J. Microelectromech. Syst. vol. 8, pp. 78-84, 1999 https://doi.org/10.1109/84.749406
  3. K. Wise, D. Anderson, J. Hetke, D. Kipke and K. Najafi, 'Wireless implantable microsystcrns: High density electronic interfaces to the nervous system', Proceedings of the IEEE vol. 92, pp. 76-97, 2004
  4. Q. Bai and K. D. Wise, 'Single unit neural recording with active microelectrode arrays', IEEE Trans. Biomed. Eng. vol. 48, pp. 911-920, 2001 https://doi.org/10.1109/10.936367
  5. P. Rousche and R. Normann, 'Chronic intracortical micro stimulation of cat sensory cortex using the utahintracortical electrode array', IEEE Trans. Rehabilitation Eng. vol. 7, pp. 56-68, 1999 https://doi.org/10.1109/86.750552
  6. Q. Bai, K. Wise and D. Anderson, 'A high yield microassembly structure for three dimensional microelectrode arrays', IEEE Trans. Biomed. Eng. vol.47, pp. 281-289, 2000 https://doi.org/10.1109/10.827288
  7. J. Chen, K. Wise, J. Hetke and S. Bledsoe, 'A multichannel neural probe for selective chemical delivery at the cellular level', IEEE Trans. Biomed. Eng. vol. 44, pp. 760-769, 1997 https://doi.org/10.1109/10.605435
  8. S. Massia, M. Holecko and G. Ehteshami, 'In vitro assessment of bioactive coatings for neural implant applications', J. Biomed Mater Res. vol. 68, pp. 177 -186, 2003
  9. T. Stieglitz, M. Gross, 'Flexible BIOMEMS with electrode arrangements on front and back side as key component in neural prostheses and biohybrid systems,' Sensors Actuator B, Vol. 83, pp. 8-14, 2002 https://doi.org/10.1016/S0925-4005(01)01021-8
  10. P.J. Rousche, D.S. Pellinen, D.P. Pivis, J.C. Williams, D. Kipke, 'Flexible polyimide-based intracortical electrode arrays with bioactive capability,' IEEE Trans. Biomed. Eng, Vol. 48, pp. 361-371, 2001 https://doi.org/10.1109/10.914800
  11. T. Shoji, S. Takafumi, M. Kunihiko, and F. Hiroyuki, '3D flexible multichannel neural probe array,' J. Micromech. Microeng, Vol. 14, pp. 104-107, 2004 https://doi.org/10.1088/0960-1317/14/1/014
  12. K. K Lee, J. He, A. Singh, S. Massia, G. Ehteshami, B. Kim and G. Raupp, 'Polyimide-base introcortical neural implant with improbed structural stiffness,' J. Micromech. Microeng, Vol. 14, pp. 32-37, 2004 https://doi.org/10.1088/0960-1317/14/1/305
  13. Zen-Jei Wei, Yung-Yun Wang, Chi-Chao Wan, Chen-He Huang, 'Study of wetters in nickel electroforming of 3D microstructures'. Materials Chemistry and Physics Vol. 63, pp. 235-239, 2000 https://doi.org/10.1016/S0254-0584(99)00228-X
  14. Y. Hanein, K. Bohringer, R. Wyeth and A. Willows, 'Towards MEMS probes for intracellular recording', Sensors update Vol. 10, pp. 47-75 2002 https://doi.org/10.1002/1616-8984(20021)10:1<47::AID-SEUP47>3.0.CO;2-A