A Study on Error Characteristics of Large Size Electromagnetic Flowmeter in the Range of Low Velocity

저유속 영역에서 대구경 전자기유량계의 오차특성 연구

  • 이동근 (한국수자원공사 수자원연구원) ;
  • 박종호 (충남대학교 기계공학과)
  • Published : 2008.03.01

Abstract

The large size electromagnetic flowmeter was tested to investigate the variation of its error characteristics in the range of low velocity under 0.6 m/s using flowmeter calibration system. For the two case of valve opening rate 100 % and 50 %, these tests were undertaken three times each for twelve velocity condition from $0.05\;^m/s\;to\;0.6\;^m/s$ with increment of $0.05\;^m/s$. It is shown that error characteristic of electromagnetic flowmeter was stabilized within ${\pm}0.4%$ of rate both higher than $0.25^m/s$ of velocity condition and 50 % of valve opening position. But, measurement deviation of flowmeter for ${\Phi}400mm\;and\;{\Phi}600mm$ was out of expected deviation range. It is necessary to correction with calibration. In conclusion, error characteristic of electromagnetic flowmeter wasn't changed proportion to its size.

Keywords

References

  1. Shercliff, J. A., 1954, 'Relation Between the Velocity Profile and the Sensitivity of Electromagnetic Flowmeters,' J. Applied Physics, Vol. 25, pp. 817-818 https://doi.org/10.1063/1.1721749
  2. KRISS, 1994, 'Development of Calibration System and Characterization Techniques of Electromagnetic Flowmeter'
  3. KRISS, Industrial Advancement Administration, 1991, 'Fluid Flow Measuremen'
  4. EA 4/02, 1993, 'Guide to the Expression of Uncertainty in Measurement'
  5. ISO 4185, 1980, ' Measurement of Liquid Flow in Closed Conduits - Weighing Method'
  6. KRISS, 1998, 'Guide to the Expression of Uncertainty in Measurement'
  7. Lim Ki Won , 2002, 'An Experimental Study on Installation Effects of Pipe Elbow on the Electromagnetic Flowmeter Characteristics (Turbulent Flow),' J. of the KSME(B), Vol. 26, No. 11, pp. 1613-1621 https://doi.org/10.3795/KSME-B.2002.26.11.1613

Cited by

  1. Error Characteristics of Clamp-on Ultrasonic Flowmeters Depending on Location of Sensors and Downstream Straight Run of Bent Pipe vol.35, pp.8, 2011, https://doi.org/10.3795/KSME-B.2011.35.8.861