Development of Asbestos Quality Contral Samples for Proficiency Analytical Testing II - Evaluation of Amosite and Chrysotile Quality Contral Samples -

석면분석 정도관리용 표준시료 개발연구 II - Amosite 및 Chrysotile 표준시료 평가 -

  • Yi, Gwang Yong (Occupational Safety & Health Research Institute, KOSHA) ;
  • Yoon, Chung sik (School of Public Health, Seoul National University) ;
  • Han, Jin Ku (Catholic Industrial Medical Center, St. Mary's Hospital, The Catholic University of Korea) ;
  • Yun, Moonjong (Korea Environmental Research & Consulting) ;
  • Jung, Sijeong (Occupational Safety & Health Research Institute, KOSHA) ;
  • Lee, Jong Han (Occupational Safety & Health Research Institute, KOSHA) ;
  • Lee, In Sub (Occupational Safety & Health Research Institute, KOSHA) ;
  • Park, Doo Yong (Department of Mechanical Systems Engineering, Hansung University)
  • 이광용 (한국산업안전보건공단 산업안전보건연구원) ;
  • 윤충식 (서울대학교 보건대학원) ;
  • 한진구 (가톨릭대학교 성모병원 가톨릭산업의학선터) ;
  • 윤문종 (한국환경 리서치 & 컨설팅) ;
  • 정시정 (한국산업안전보건공단 산업안전보건연구원) ;
  • 이종한 (한국산업안전보건공단 산업안전보건연구원) ;
  • 이인섭 (한국산업안전보건공단 산업안전보건연구원) ;
  • 박두용 (한성대학교)
  • Received : 2009.08.05
  • Accepted : 2009.09.10
  • Published : 2009.09.30

Abstract

We conducted validation for asbestos quality control (QC) samples made by the developed method which was presented in our published paper. The main results of the validation study are as follows. The pooled coefficient variations of analytical results in three different concentrations (low, medium, and high concentration) for amosite and chrysotile were less than 20 %, which met the sample homogenicity criteria of NIOSH. Also we confirmed the homogenicity of asbestos samples by using the relocatable field slide. To evaluate the field applicability of the developed asbestos QC samples, the field validation was performed with four proficient asbestos analysts by using the statistical methods of AIHA asbestos PAT program. All analytical results from four asbestos analysts were located in the acceptable range.

Keywords

References

  1. 이광용, 이종한, 정시정, 박두용. 석면분석 정도관리용 표준시료 개발연구 I. 한국산업위생학회. 2009;19(2):81-87
  2. Niemeyer I.C, Gerchman LL. Preparation of asbestos filter samples for reference standard using a wet generation technique. American Industrial Hygiene Association Journal. 1981;42:757-759 https://doi.org/10.1080/15298668191420657
  3. Ortiz LW, Ettinger HJ, Fairchild CI. Calibration Standard for Counting Asbestos. American Industrial Hygiene Association Journal. 1975;36:104-112 https://doi.org/10.1080/0002889758507217
  4. Schlecht PC, Shulman SA. Performance of Asbestos fiber counting laboratories in the NIOSH Proficiency Analytical Testing(PAT) Program. American Industrial Hygiene Association Journal. 1986;47(5):259-269 https://doi.org/10.1080/15298668691389748
  5. Schlecht PC, Shulman SA. Phase contrast microscopy asbestos fiber counting performance in the Proficiency Analytical Testing Program. American Industrial Hygiene Association Journal. 1995;56:480-489
  6. Pang TWS, Dicker WL, Nazer MA. An evaluation of the precision and accuracy of the direct transfer method for the analysis of asbestos fibers with comparison to the NIOSH method. American Industrial Hygiene Association Journal. 1984;45(5):329-335
  7. Taylor DG, Baron PA. Shulman SA, Carter JW. Identification and counting of asbestos fibers. American Industrial Hygiene Association Journal. 1984;45(2):84-88 https://doi.org/10.1080/15298668491399415
  8. Baron PA, and Deye GJ. Generation of replicate asbestos aerosol samples for quality assurance. Applied Industrial Hygiene. 1994;2(3):114-118
  9. Skogstad A, Eduard W, Huser PO. A laboratory method for generation of replicate fiber samples of asbestos fibers in air. American Industrial Hygiene Association Journal. 1996;57:741-745 https://doi.org/10.1080/15428119691014602
  10. Timbrell V, Rendall REG. Preparation of the UICC Standard Reference samples of asbestos. Powder Technology 1971- 1972;5:279-287 https://doi.org/10.1016/0032-5910(72)80032-9
  11. Timbrell V, Skidmore JW, Hyett AW, Wagner JC. Exposure chambers for inhalation experiments with standard reference samples of asbestos of the International Union Against Cancer(UICC). Aerosol Science. 1970;Vol 1:215-223 https://doi.org/10.1016/0021-8502(70)90035-2
  12. National Institute of Occupational Safety and Health(NIOSH). NIOSH Manual of Analytical Method. 4th ed. NIOSH;1994