DOI QR코드

DOI QR Code

휴대형 심전계 장치의 RR 간격의 정확도 개선을 위한 보간법 개발

Interpolation Technique to Improve the Accuracy of RR-interval in Portable ECG Device

  • 이은미 (충북대학교 의과대학 의공학교실) ;
  • 홍주현 (충북대학교 의과대학 의공학교실) ;
  • 차은종 (충북대학교 의과대학 의공학교실) ;
  • 이태수 (충북대학교 의과대학 의공학교실)
  • Lee, Eun-Mi (Department of Biomedical Engineering, College of Medicine, Chungbuk National University) ;
  • Hong, Joo-Hyun (Department of Biomedical Engineering, College of Medicine, Chungbuk National University) ;
  • Cha, Eun-Jong (Department of Biomedical Engineering, College of Medicine, Chungbuk National University) ;
  • Lee, Tae-Soo (Department of Biomedical Engineering, College of Medicine, Chungbuk National University)
  • 투고 : 2010.05.12
  • 심사 : 2010.06.09
  • 발행 : 2010.09.30

초록

HRV(Heart rate variability) analysis parameter is widely used as an index to evaluate the autonomic nervous system and cardiac function. For reliable HRV analysis, we need to acquire the accurate ECG signals. Most of commercially available portable ECG devices have low sampling rate because of low power consumption and small size issues, which make it difficult to measure RR-interval accurately. This study is to improve the accuracy of RR-interval by developing R-wave interpolation technique, based on the morphological characteristics of the QRS complex. When the developed method was applied to ECG obtained at 200 Hz and the results were compared with 1000 Hz reference device, the error range decreased by 1.33 times in sitting and by 2.38 times in cycling exercise. Therefore, the proposed interpolation technique is thought to be useful to improve the accuracy of R-R interval in the portable ECG device with low sampling rate.

키워드

참고문헌

  1. Statistics Korea, "Number of death(1983-) according to the causes of death(236 items)/gender/age(5 years), Death rate(2000-)," Available at http://www.kostat.go.kr, Accessed Feb 19, 2010.
  2. T.S. Lee, and J.H. Hong, "Current Developments of Biomedical Mobile Devices for Ubiquitous Healthcare," J. of Biomed. Eng. Res., vol. 30, no. 3, pp.185-190, 2009.
  3. Task Force of The European Society of Cardiology and The North American Society of Pacing and Electrophysiology, "Heart rate variability - Standards of measurement, physiological interpretation, and clinical use," European Heart Journal, vol. 17, no. 3, pp.354-381, 1996. https://doi.org/10.1093/oxfordjournals.eurheartj.a014868
  4. M. Merri, D.C. Farden, J.G. Mottley, and E.L. Titlebaum, "Sampling frequency of the electrocardiogram for spectral analysis of the heart rate variability," IEEE Transactions on Biomedical Engineering, vol. 37, no. 1, pp.99-106, 1990. https://doi.org/10.1109/10.43621
  5. F. Badilini, AJ. Moss, and E.L. Titlebaum, "Cubic Spline Baseline Estimation in Ambulatory ECG Recordings for the Measurement of ST Segment Displacements," in Proc. the 13th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Orlando, FL, USA, Oct. 1991, pp.584-585.
  6. P. Castiglioni, L. Piccini, and M.Di. Milan, "Interpolation Technique for Extracting Features from ECG Signals Sampled at Low Sampling Rates," Computers in Cardiology, vol. 30, pp.481-484, 2003.
  7. G.M. Frieson, T.C. Jannett, M.A. Jadallah, S.L. Yates, S.R. Quint, and H.T. Nagle, "A Comparison of the Noise Sensitivity of Nine QRS Detection Algorithms," IEEE Transactions on Biomedical Enginnering, vol. 37, no. 1, pp.85-98, 1990. https://doi.org/10.1109/10.43620
  8. K. Dewitte, C. Fierens, D. Stockl, and L.M. Thienpont, "Application of the Bland–Altman Plot for Interpretation of Method-Comparison Studies: A Critical Investigation of Its Practice," Clinical Chemistry, vol. 48, no. 1, pp.799-801, 2002.
  9. G. Nollo, L. Faes, B. Pellegrini, A. Porta, and R. Antolini, "Synchronization Index for Quantifying Nonlinear Causal Coupling between RR Interval and Systolic Arterial Pressure after Myocardial Infarction," Computers in Cardiology, vol. 27, pp.143-146, 2000.
  10. J. Allen, "Photoplethysmography and Its Application in Clinical Physiological Measurement," Physiological Measurement, vol. 28, no. 3, pp. R1-R39, 2007. https://doi.org/10.1088/0967-3334/28/3/R01