DOI QR코드

DOI QR Code

Unexpected Detachment of Solitaire Stents during Mechanical Thrombectomy

  • Kim, Sung Tae (Department of Neurosurgery, Busan Paik Hospital, Inje University) ;
  • Jin, Sung-Chul (Department of Neurosurgery, Haeundae Paik Hospital, Inje University) ;
  • Jeong, Hae Woong (Department of Diagnostic Radiology, Busan Paik Hospital, Inje University) ;
  • Seo, Jung Hwa (Department of Neurology, Busan Paik Hospital, Inje University) ;
  • Ha, Sam Yeol (Department of Neurology, Haeundae Paik Hospital, Inje University) ;
  • Pyun, Hae Wook (Department of Diagnostic Radiology, S Pohang Hospital)
  • Received : 2014.07.16
  • Accepted : 2014.10.25
  • Published : 2014.12.28

Abstract

Objective : Unexpected Solitaire stent detachment can occur during mechanical Solitaire thrombectomy. The purpose of this study was to retrospectively evaluate the influencing factors causing unexpected Solitaire stent detachment and the clinical outcomes. Methods : Between October 2011 to December 2013, 232 cases of mechanical Solitaire thrombectomy for acute ischemic stroke were performed in 3 stroke centers. During this period, we encountered unexpected Solitaire stent detachments during mechanical Solitaire thrombectomies in 9 cases. Results : Solitaire stents unexpectedly detached in 9 cases (3.9%) during the retrieval of Solitaire stents. The median patient age was 76 years. The occlusion sites of the unexpected stent detachment were the proximal middle cerebral artery (MCA) in 7 cases and the internal carotid artery in 2 cases. The sizes of the stents that unexpectedly detached were $6{\times}30$ mm in 7 cases, $5{\times}30$ mm in 1 case, and $4{\times}20$ mm in 1 case. Four patients had unexpected detachment at the first retrieval, 1 patient at the second, 3 patients at the third, and 1 patient at the fifth. In all of the cases of unexpected detachment at the first retrieval, the stent deployment site was the proximal MCA. After detachment, a proximal marker of the Solitaire stent was observed in 3 patients. However, no marker was visible in the remaining 6 patients. Conclusion : Unexpected Solitaire stent detachment should be considered in the first instance of stent retrieval for a relatively large-diameter stent, especially in elderly patients with MCA occlusions.

Keywords

References

  1. Davalos A, Pereira VM, Chapot R, Bonafe A, Andersson T, Gralla J; Solitaire Group : Retrospective multicenter study of Solitaire FR for revascularization in the treatment of acute ischemic stroke. Stroke 43 : 2699-2705, 2012 https://doi.org/10.1161/STROKEAHA.112.663328
  2. Dorn F, Stehle S, Lockau H, Zimmer C, Liebig T : Endovascular treatment of acute intracerebral artery occlusions with the solitaire stent : single-centre experience with 108 recanalization procedures. Cerebrovasc Dis 34 : 70-77, 2012 https://doi.org/10.1159/000338903
  3. Kang DH, Park J, Hwang YH, Kim YS : Inadvertent self-detachment of Solitaire AB stent during the mechanical thrombectomy for recanalization of acute ischemic stroke : lessons learned from the removal of stent via surgical embolectomy. J Korean Neurosurg Soc 53 : 360-363, 2013 https://doi.org/10.3340/jkns.2013.53.6.360
  4. Kwon HJ, Chueh JY, Puri AS, Koh HS : Early detachment of the Solitaire stent during thrombectomy retrieval : an in vitro investigation. J Neurointerv Surg, 2014 JAN 16 [Epub]. http://dx.doi.org/10.1136/neurintsurg-2013-010942
  5. Mokin M, Dumont TM, Veznedaroglu E, Binning MJ, Liebman KM, Fessler RD 2nd, et al. : Solitaire Flow Restoration thrombectomy for acute ischemic stroke : retrospective multicenter analysis of early postmarket experience after FDA approval. Neurosurgery 73 : 19-25; discussion 25-26, 2013 https://doi.org/10.1227/01.neu.0000429859.96652.57
  6. Nayak S, Ladurner G, Killer M : Treatment of acute middle cerebral artery occlusion with a Solitaire AB stent : preliminary experience. Br J Radiol 83 : 1017-1022, 2010 https://doi.org/10.1259/bjr/42972759
  7. Pereira VM, Gralla J, Davalos A, Bonafé A, Castaño C, Chapot R, et al. : Prospective, multicenter, single-arm study of mechanical thrombectomy using Solitaire Flow Restoration in acute ischemic stroke. Stroke 44 : 2802-2807, 2013 https://doi.org/10.1161/STROKEAHA.113.001232
  8. Saver JL, Jahan R, Levy EI, Jovin TG, Baxter B, Nogueira RG, et al. : Solitaire flow restoration device versus the Merci Retriever in patients with acute ischaemic stroke (SWIFT) : a randomised, parallel-group, non-inferiority trial. Lancet 380 : 1241-1249, 2012 https://doi.org/10.1016/S0140-6736(12)61384-1

Cited by

  1. Unwanted detachment of the Solitaire device during mechanical thrombectomy in acute ischemic stroke vol.8, pp.12, 2014, https://doi.org/10.1136/neurintsurg-2015-012156
  2. Revolution in acute ischaemic stroke care: a practical guide to mechanical thrombectomy vol.17, pp.4, 2014, https://doi.org/10.1136/practneurol-2017-001685
  3. Complications of endovascular treatment for acute ischemic stroke: Prevention and management vol.13, pp.4, 2014, https://doi.org/10.1177/1747493017743051
  4. Solitaire stentectomy: ‘deploy and engage’ and ‘loop and snare’ techniques vol.10, pp.5, 2018, https://doi.org/10.1136/neurintsurg-2016-012547.rep
  5. Accidental guide catheter fracture in mechanical thrombectomy vol.25, pp.3, 2014, https://doi.org/10.1177/1591019918814011
  6. Solitaire Stentectomy Using a Stent-Retriever Technique in a Porcine Model vol.31, pp.2, 2021, https://doi.org/10.1007/s00062-020-00906-1