DOI QR코드

DOI QR Code

Development of Dose Verification Method for In vivo Dosimetry in External Radiotherapy

방사선치료에서 투과선량을 이용한 체내선량 검증프로그램 개발

  • Hwang, Ui-Jung (Department of Radiation Oncology, National Medical Center) ;
  • Baek, Tae Seong (Department of Bio-convergence Engineering, Korea University) ;
  • Yoon, Myonggeun (Department of Bio-convergence Engineering, Korea University)
  • 황의중 (국립중앙의료원 방사선종양학과) ;
  • 백태성 (고려대학교 바이오융합공학과) ;
  • 윤명근 (고려대학교 바이오융합공학과)
  • Received : 2014.02.28
  • Accepted : 2014.03.13
  • Published : 2014.03.31

Abstract

The purpose of this study is to evaluate the developed dose verification program for in vivo dosimetry based on transit dose in radiotherapy. Five intensity modulated radiotherapy (IMRT) plans of lung cancer patients were used in the irradiation of a homogeneous solid water phantom and anthropomorphic phantom. Transit dose distribution was measured using electronic portal imaging device (EPID) and used for the calculation of in vivo dose in patient. The average passing rate compared with treatment planning system based on a gamma index with a 3% dose and a 3 mm distance-to-dose agreement tolerance limit was 95% for the in vivo dose with the homogeneous phantom, but was reduced to 81.8% for the in vivo dose with the anthropomorphic phantom. This feasibility study suggested that transit dose-based in vivo dosimetry can provide information about the actual dose delivery to patients in the treatment room.

방사선치료에서는 환자 체내에 전달된 선량이 원래 의도한 데로 분포되는 지 확인하기 위하여 균질한 팬텀을 이용한 정도 관리를 치료 전에 주로 시행하고 있다. 하지만 균질한 팬텀을 이용한 정도 관리는 표면이 불규칙적이고 불균질한 인체에 대한 선량분포를 완전히 보증해 주지는 못하고 있다. 본 연구에서는 환자를 투과하는 선량의 분포를 측정하여 역으로 환자체내 선량 분포를 계산하는 투과선량 기반 체내선량 검증프로그램을 개발하였다. 투과선량은 주방사선과 산란방사선으로 이루어져 있는데, 본 연구에서는 전자포탈영상장치로 측정한 선량분포로부터 주방사선만을 이용한 간단한 식으로 환자체내선량분포를 계산하는 프로그램과 감마값 분포를 평가하여 두 선량분포를 서로 비교할 수 있는 프로그램을 개발하였다. 개발된 프로그램을 이용하여 계산한 팬텀의 등중심점을 지나는 관상면의 체내선량 분포는 치료계획시스템에서 제공하는 동일 평면의 선량분포와의 비교결과 균질팬텀에서 평균 95%, 비균질팬텀에서 81.8%의 감마통과율을 보였다.

Keywords

References

  1. Boellaard R, M. van Herk, Mijnheer BJ: A convolution model to convert transmission dose images to exit dose distributions. Medical Physics 24:189-200 (1997) https://doi.org/10.1118/1.598066
  2. Ronald Boellaard, Marion Essers, Marcel van Herk, Ben J. Mijnheer: New method to obtain the midplane dose using portal in vivo dosimetry. Int J Rad Oncol Biol Phys 41(2):465-474 (1998) https://doi.org/10.1016/S0360-3016(98)00048-0
  3. Louwe RJW, Damen EMF, M. van Herk, Minken AWH, Torzsok O, Mijnheer BJ: 3D dose reconstruction of breast cancer treatment using portal imaging. Medical Physics 30: 2376-2398 (2003) https://doi.org/10.1118/1.1589496
  4. WJC van Elmpt, Nijsten SMJJG, Mijnheer BJ, Minken AWH: Experimental verificaiton of a portal dose prediction model. Medical Physics 32:2805-2818 (2005) https://doi.org/10.1118/1.1987988
  5. Markus Wendling, Robert J. W. Louwe, Leah N. McDermott, Jan-Jakob Sonke, Marcel van Herk, Ben J. Mijnheer: Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method. Medical Physics 33(2): 259-336 (2006) https://doi.org/10.1118/1.2147744
  6. Wouter J. C. van Elmpt, Sebastiaan M. J. J. G. Nijsten, Robert FH, et al: Monte Carlo based 3D dose reconstruction method derived from portal dose images. Medical Physics 33(7):2426-2478 (2006) https://doi.org/10.1118/1.2207315
  7. Louwe RJW, Wendling M, MB van Herk, Mijnheer BJ: 3D heart dose reconstruction to estimate normal tissue complication probability after breast irradiation using portal dosimetry. Medical Physics 34(4):1354-1371 (2007) https://doi.org/10.1118/1.2713216
  8. Markus Wendling, Leah N. McDermott, Anton Mans, Jan-Jakob Sonke, Marcel van Herk, Ben J. Mijnheer: Simple backprojection algorithm for 3D in vivo EPID dosimetry of IMRT treatments. Medical Physics 36(7):3310-3321 (2009) https://doi.org/10.1118/1.3148482
  9. Faiz M. Khan: The Physics of Radiation Therapy. 4th ed, Williams & Wilkins, Baltimore, MD (2010), pp. 146-148
  10. Daniel a. Low, James F. Dempsey: Evaluation of the gamma dose distribution comparison method. Medical Physics 30(9):2455-2464 (2003) https://doi.org/10.1118/1.1598711
  11. Janusz Winiecki, kTomasz Morgas, Karolina Majewaska, Barbara Drzewiecka: The gamma evaluation method as a routine QA procedure of IMRT. Reports of Practical Oncology & Radiotherapy 14(5):162-168 (2009)