Abstract

This study was conducted to minimize the imaging distortion by reducing the differences in susceptibility between the tissue and air surrounding it while performing a thyroid diffusion imaging procedure. The study group was composed of 23 healthy adults. Thyroids with many distortions near the air, larynx, and trachea were chosen to test and evaluate the diffusion imaging difference between before and after an application of silicon. As a result, there was reduced distortion with silicon application, and the differences decreased from 30% to 10%. According to One-way ANOVA and Duncan's post-hoc test, there were no significant differences between imaging with a silicon application and T2 imaging of the surface area, which was the standard image. In conclusion, this study presented a radical improvement in reducing the distortions in imaging by compensating for an uneven tissue surface near air without affecting the magnetic resonance contrast and complicating the imaging processes.

본 연구는 갑상선의 확산강조영상 검사 시 조직과 공기 간 자화율차이를 줄임으로서 영상의 뒤틀림을 감소시키고자 하였다. 연구대상은 정상인 23명을 대상으로 하였으며, 공기와 인접해 있고 주변에 후두와 기관이 자리 잡고 있어 뒤틀림이 많이 발생하는 갑상선을 대상으로, 실리콘 적용 전 후 확산강조영상의 뒤틀림과 오차율을 비교평가 하였다. 연구결과 실리콘 적용 전보다 적용 후가 뒤틀림이 감소된 것을 볼 수 있으며, 오차율 또한 30%대에서 10%대로 감소하는 것을 볼 수 있었다. 또한 일원배치분산분석과 Duncan의 사후분석을 통해서도 실리콘 적용 후가 기준영상인 T2 강조영상의 면적과 차이가 없음을 알 수 있었다. 결론적으로 본 연구는 선행연구들의 복잡한 과정 없이 자기공명신호의 대조도에 영향을 주지 않으면서 공기와 맞닿은 굴곡진 부분을 보상함으로써 영상의 뒤틀림을 감소시킬 수 있는 근본적인 개선 방안을 제시하였다는 데에 커다란 의의가 있다.

Keywords

References

  1. National Cancer Information Center, Available From: http://www.cancer.go.kr/ncic/cics_f/01/012/index.html, (accessed Jun, 23, 2013)
  2. Miyakoshi. A, Dalley. R.W, Anzai. Y, "Magnetic Resonance Imaging of Thyroid Cancer", Top Magn Reson Imaging, v.18, n.4, pp.293-302, 2007. DOI: http://dx.doi.org/10.1097/RMR.0b013e318572b76
  3. Olivier. T, Dominique G, Rachel M, "Comparison between clinical, ultrasound, CT, MRI and pathology findings in dogs presested for suspected thyroid carcinoma" : Veterinary radiology & ultrasound :Vet Radiol Ultrasound, Associationv.54, no.1, pp.61-70, 2013. https://doi.org/10.1111/j.1740-8261.2012.01966.x
  4. Ying. Y, Yue. X.H, Tao. X.F, "The diagnostic value of dynamic contrast-enhanced MRI for thyroid tumors", Eur J Radiol, v.81, no.11, pp.3313-3318, 2012. DOI: http://dx.doi.org/10.1016/j.ejrad.2012.04.029
  5. Guan. Y, Xie. B, Yuan. X, Yuan XP, Li. HG, "Diagnosis of thyroid carcinoma by magnetic resonance imaging", Chin J Cancer, v.22, no.7, pp.739-744, 2003.
  6. Mutlu, H. Sivrioglu, A.K. Sonmez, G, Veliglu HO, Basekim CC, Kizikaya E, "Role of apparent diffusion coefficent values and diffusion weighted magnetic resonance imaging in differentiation between benign and malignant thyroid nodules", Clin Imaging, v.36, n.1 ,pp.1-7, 2012. DOI: http://dx.doi.org/10.1016/j.clinimag.2011.04.001
  7. Razek, A. A. K, Sadek, A. G, Kombar. O. R, "Role of apparent diffusion coefficeint vales in differentiation between malignant and benign solitary thyroid nodules", AJNR Am J Neuroradiol. v.29, no.3, pp.563-68, 2003. DOI: http://dx.doi.org/10.3174/ajnr.A0849
  8. Kato. H, Kanematsu. M, Tanaka. O, "Head and neck squamous cell carcinoma : usefulness of diffusion weighted MR imaging in the prediction of a neoadjuvant therapeutic effect", Eur J Radiol, v.19, n.1, pp.103-109, 2009. DOI: http://dx.doi.org/10.1007/s00330-008-1108-5
  9. Schueller-Weidekamm. C, Kaserer. K, Schueller. G, "Can quantitative diffusion weighted mr imaging differentiate benign and malignant cold thyroid nodules? initial results in 25 patients", AJNR Am J Neuroradiol, v.30, n.2, pp.417-422, 2009.
  10. ZH Cho, Ro. YM,. "Reduction of Susceptibility Artifact in Gradient-Echo Imaging", Magn Reson Med, 23, pp.193, 1992. DOI: http://dx.doi.org/10.1002/mrm.1910230120
  11. KW Choi, SY Son, "A research on improving signal to noise ratio for magnetic resonance imaging through increasing filling factor inside surface coil", Journal of the Korea Academia-Industrial cooperation Society, v.13, n.11, pp.5299-5304, 2012. DOI: http://dx.doi.org/10.5762/KAIS.2012.13.11.5299
  12. KW Choi, HB Lee, "A study on the reduction on magnetic susceptible artifacts through the usage of silicon", Journal of the Korea Academia-Industrial cooperation Society, v.13, n.12, pp.5937-5942, 2012. DOI: http://dx.doi.org/10.5762/KAIS.2012.13.12.5937
  13. Chung. J. Y, Han. Y, Park. H, "An analysis and correction method of artifacts in gradient-echo EPI Using spin-echo EPI Reference", MAGMA, 2007.
  14. Heid O, "Robust EPI Phase Correction", J Magn Reson Imaging, pp.2014, 1997.
  15. Buonocore M. H, Gao L, "Ghost Artifact Reduction for Echo Planar Imaging Using Image Phase Correction", Magn Reson Med, v.38, no.1, pp.89-100, 1997. DOI: http://dx.doi.org/10.1002/mrm.1910380114
  16. Sohn. C.H, Kim .I.S, "PROPELLER(periodically rotated overlapping parallel lines with enhanced reconstruction) and EPI Diffusion-weighte d MR Imaging at 3.0 T: Pontine Magnetic Susceptibility Artifacts Depend on Pneumatization of the Sphenoid Sinus", J. korea radiol soc, v.55, pp.321-326, 2006. https://doi.org/10.3348/jkrs.2006.55.4.321
  17. Van den bergen B, Stolk CC, Berg JB, Lagendijik JJ, "ultra fast electromagnetic computations for RF Multi-transmit techiqu in hih field MRI", Phys Med Biol, no.54, pp.1253-1364, 2009.
  18. Yamashita. K, Yoshiura. T, Hiwatashi. A, Kamano. H, Dashjamts. T, Shibata. S, Tamae. A, Honda. H, "Detection of middle earcholesteatoma by diffusion weighted mr imaging :multishot echo plan an imaging compared with single shot echo planar imaging", AJNR Am J Neuroradiol, v.32, pp.1915-918, 2011. DOI: http://dx.doi.org/10.3174/ajnr.A2651
  19. Zhuang J, Hrabe. J, Kangarlu. A, "Correction of eddy-current distortions in diffusion tensor images using the known directions and strengths of diffusion gradients", J Magn Reson Imaging, v.24, no.5, pp.1188-1193, 2006. DOI: http://dx.doi.org/10.1002/jmri.20727

Cited by

  1. The Study on Reduction of Image Distortion by using Single-Shot Turbo Spin Echo in Brain Stem Diffusion MRI vol.10, pp.4, 2016, https://doi.org/10.7742/jksr.2016.10.4.279