Abstract

In this paper, we studied additional filter and Ion chamber combinations to reduce patient dose without decreasing image quality in digital chest x-ray projection. The experiment set 125 kVp, 320 mA, AEC mode. Ion chamber sensors was divided by 4 cases of combinations, then, we measured patient dose and calculated organ dose using PCXMC. Also, physical image assessment using MTF was performed. As a results, The surface entrance dose and organ dose were the lowest when selecting both left and right Ion chamber sensors under the same conditions of additional filter. In image quality assessment, The spatial frequency scored 2.494 lp/mm which was highest when selecting both right and left Ion-chambers and 0.1 mmCu filter. And to conclude, to select both right and left Ion chamber sensors and 0.1 mmCu filter will help for acquiring good quality image as well as reducing patient dose.

본 연구는 디지털 흉부엑스선 검사에서 화질의 저하 없이 환자선량을 감소시키기 위한 부가필터와 Ion chamber 센서 조합을 알아보고자 하였다. 실험은 관전압 125 kVp, 관전류 320 mA, AEC모드로 하여 부가필터와 Ion chamber의 센서를 네 가지 조합으로 나누어 선량을 측정하고, PCXMC를 이용하여 장기선량을 산출하였다. 또한 MTF로 물리적 화질을 평가하였다. 그 결과 동일 부가필터의 조건하에서 Ion chamber의 좌우 양쪽 센서 모두를 선택했을 때 입사표면 선량과 장기선량이 가장 낮게 나타났으며, 화질평가에서는 좌우 Ion chamber의 선택과 0.1 mmCu 필터를 선택했을 때 공간주파수 값이 2.494 lp/mm로 가장 높게 나타났다. 결론적으로 디지털 흉부촬영 시 Ion chamber의 좌우 양쪽 센서와 0.1 mmCu 필터를 선택하는 것이 우수한 화질의 영상을 획득하고 환자선량 저감에 도움이 될 것이다.

Keywords

References

  1. Kwak JH, Jeong JB, "A Study on the Guardan's Perception of Attending Patient in Pediatric Radiology", J. Korean Soc. Radiology, Vol. 8, No. 4, pp.189-201, 2014. https://doi.org/10.7742/jksr.2014.8.4.189
  2. Lee IJ, Kim HY, Kim NC, Lee YC, Park YK, "Evaluation of image According to Exposure Conditions using Contrast-Detail Phantom for Chest Digital Radiography", J. Radiological Science and Technology, Vol. 32, No. 1, pp.25-32, 2009.
  3. Jo KH, Kang YH, Kim BS, "A Study on the Exposure Parameter and the Patient Dose for Digital Radiography System in Dae Goo", J. Korean Soc. Radiol. Science, Vol. 31, No. 2, pp.177-182, 2008.
  4. Kim YH, "Method for Measurement of Entrance Surface Dose and Role of Radiation technologist", J. Radiological Science and Technology, Vol. 28, No. 3, pp.173-191, 2005.
  5. Ullman G, Svendborg M, Dance DR, Hunt R, Alm Carlsson G, "The influence of patient thickness and imaging system on patient dose and physical image quality in chest image", Radiat. Prot. Dosimetry, Vol.114, No. 1-3, pp.294-297, 2005. https://doi.org/10.1093/rpd/nch542
  6. Jung JS, Choi BU, Kim SH, Kim YM, Shim JN, Ahn HS, Jin DE, Lim JS, Kang SH, "Effect of Automatic Exposure Control Marker with Chest Radiography in Radon Reduction", J. Radiological Science and Technology, Vol. 32, No. 1, pp.25-32, 2009.
  7. Choi NG, Seong HJ, Jeon JS, Kim YH, Seong DO, "A Comparative Study of Image Quality and Radiation Dose according to Variable Added Filter and Radiation Exposure in Diagnostic X-ray Radiography", J. Radiation protection, Vol. 37, No. 1, pp.25-34, 2012. https://doi.org/10.14407/jrp.2012.37.1.025
  8. Huh J, Kim CK, LEE IJ, Shin WS, Park JC, Kang HS, Ahn BS, "A study on the effect of rare earth screen Filter", J. Korean Soc. Radiological technology, Vol. 11, No. 1, pp.17-23, 1988.
  9. Shin SG, "Decreased of Patient dose by Built-in Filter in Pelvis A-P Projection", J. Korean Contents Association, Vol. 12, No. 8, pp.233-239, 2012.
  10. Seok DW, "Evaluation of Imaging Performance of the a-Se based Flat-panel X-ray Detector System using MTF, NPS and DQE measurements", The graduate school of Inje University, 2004.
  11. Lee JS, Ko SJ, Kang SS, Kim JH, Kim DH, Kim CS, "Quantitative Evaluation of Image Quality using Automatic Exposure Control & Sensitivity in the Digital Chest Image", J. Korean Contents Association, Vol. 13, No. 8, pp.275-283, 2013.
  12. Choi KM, Shin SI, Yoon JM, Kim SC, Lee SS, Huh J, "The Reduction of Radiation Dose using Key-Filter in Chest Radiography", J. Radiological Science and Technology, Vol. 19, No. 2, pp.173-191, 1996.
  13. Schaefer-Prokop C, Neitzel U, Venema HW, Uffmann M, Prokop M; "Digital chest radiography: an update on modern technology, dose containment and control of image quality", European Radiology, Vol. 18, Issue 9, pp.1818-1830, 2008. https://doi.org/10.1007/s00330-008-0948-3
  14. Kim YS, Park HS, Park SJ, Kim HJ, "Effective Detective Quantum Efficiency (eDQE) Evaluation for the Influence of Focal Spot Size and Magnification on the Digital Radiography System", Korean journal of medical physis, Vol. 23, No. 1, pp.26-32, 2012.
  15. Kwon SM, Park CH, Park JK, Song UH, Jung JE, "The Effect of Source to Image-Receptor Distance(SID) on Radiation Dose for Digital Chest Radiography", J. Korean Soc. Radiology, Vol. 8, No. 4, pp.203-210, 2014. https://doi.org/10.7742/jksr.2014.8.4.203
  16. "ICRP, 2007 Recommendations of the International Commission on Radiological Protection Publication 103", Annals of the ICRP Vol. 137, No. 2-4, Pergamon Press, Oxford, 2007.

Cited by

  1. A Convergence Study on Evaluation of Usefulness of Copper Additional Filter in the Digital Radiography System vol.13, pp.9, 2015, https://doi.org/10.14400/JDC.2015.13.9.351
  2. The Study of Forward Scattering Dose according to the Thickness of Filter in General Radiography vol.9, pp.7, 2015, https://doi.org/10.7742/jksr.2015.9.7.445
  3. The Study of Affecting Image Quality according to forward Scattering Dose used Additional Filter in Diagnostic Imaging System vol.10, pp.8, 2016, https://doi.org/10.7742/jksr.2016.10.8.597
  4. 인터벤션 시 방사선조사 조건에 따른 선량감소 : 면적선량과 영상화질 변화를 중심으로 vol.40, pp.3, 2015, https://doi.org/10.17946/jrst.2017.40.3.06
  5. 자동노출제어장치의 채광창 차폐정도와 농도, 감도의 변화가 관전류량과 영상품질에 미치는 영향 vol.14, pp.5, 2015, https://doi.org/10.7742/jksr.2020.14.5.635
  6. 미세 대조도 영상을 기반한 디지털 방사선 영상 시스템의 자동노출제어 조절인자 간의 상관관계 분석 vol.44, pp.1, 2015, https://doi.org/10.17946/jrst.2021.44.1.1
  7. 부비동 디지털 엑스선검사에서 자동노출제어 조절인자의 최적화를 통한 조사선량 감소 효과 vol.44, pp.3, 2015, https://doi.org/10.17946/jrst.2021.44.3.173