DOI QR코드

DOI QR Code

A Study on the Thyroid Dose High-Energy Radiation Therapy of Lung Cancer

폐암 고에너지 방사선치료 시 갑상선 피폭에 관한 연구

  • 양오남 (목포과학대학교 방사선과/한서대학교 보건의료학과) ;
  • 임청환 (한서대학교 보건의료학과)
  • Received : 2015.03.23
  • Accepted : 2015.04.15
  • Published : 2015.06.28

Abstract

High-energy medical linear accelerator on the dose to the thyroid cancer during radiotherapy were evaluated using optical stimulation luminescence dosimeters(OSLD) using. Scattered's influence in the case of 3D-CRT 25.4 mSv, 28.8 mSv, 31.3 mSv, 26.5 mSv, 27.4 mSv 5 times with an average 27.9 mSv, in the IMRT 46.8 mSv, 43.2 mSv, 42.3 mSv, 41.5 mSv, 44.1 mSv to five times the average of 43.6 was the result of mSv. In the case of light neutron dosimetry results 3D-CRT 3 mSv, 3 mSv, 3.4 mSv, 3.5 mSv, 3.1 mSv to five times the average 3.2 mSv, in the IMRT 5.1 mSv, 4.8 mSv, 4.2 mSv, 4.8 mSv, 4.9 mSv, to five times the average of 4.7 was the result of mSv. Both parties and the light scattered neutrons were significantly appreciated compared to IMRT 3D-CRT. Treatment of cancer using radiation workers, as in this study, and that a significant amount of scattered rays in the adjacent normal tissues during radiation therapy using energy assessment to influence by fully aware of this information is necessary for the exposure reduction efforts the feed.

고에너지 의료용 선형가속기를 이용한 폐암 방사선치료 시 갑상선에 미치는 선량을 광자극발광선량계(OSLD)를 이용하여 평가하였다. 산란광자의 영향은 3D-CRT의 경우 25.4 mSv, 28.8 mSv, 31.3 mSv, 26.5 mSv, 27.4 mSv로 5회 평균은 27.9 mSv, IMRT에 있어서는 46.8 mSv, 43.2 mSv, 42.3 mSv, 41.5 mSv, 44.1 mSv로 5회 평균은 43.6 mSv의 결과 값을 보였다. 광중성자 선량 평가 결과는 3D-CRT의 경우 3 mSv, 3 mSv, 3.4 mSv, 3.5 mSv, 3.1 mSv로 5회 평균은 3.2 mSv, IMRT에 있어서는 5.1 mSv, 4.8 mSv, 4.2 mSv, 4.8 mSv, 4.9 mSv로 5회 평균은 4.7 mSv의 결과 값을 보였다. 산란광자와 광중성자 모두 3D-CRT 보다 IMRT가 높은 것으로 평가 되었다. 본 연구를 통하여 고에너지를 이용한 방사선치료 시 인접한 정상조직에 상당한 양의 산란선량이 영향을 주는 것으로 평가된 바와 같이 방사선을 이용한 암 치료 종사자는 이러한 내용을 충분히 인지하여 피폭 저감화를 위한 노력이 필요할 것으로 사료된다.

Keywords

References

  1. 대한방사선치료학회 장비통계, 2011.
  2. 통계청, 암 통계, 2008.
  3. 장남준, 석진용, 원희수 등, "두경부암 세기변조 방사선치료 계획 시 부분적 고에너지 광자선 사용에 따른 치료계획 평가", 대한방사선치료학회지, 제21권, 제1호, pp.1-8, 2013.
  4. H. Ing, W. R. Nelson, and R. A. Shore, "Unwanted photon and neutron radiation resulting from collimated photon beams interacting with the body of rad iotherapy patients," Med. Phys. Vol.9, pp.27-33, 1982. https://doi.org/10.1118/1.595137
  5. 정제호, 김홍석, 박윤환, "의료용 선형가속기에서 생성되는 광중성자에 의한 선량 평가", 한국방사성폐기물학회지, 제2004권, 제1호, pp.296-296, 2004.
  6. R. Nath, E. R. Epp, J. S. Langhlin, W. P. Swanson, and V. P. Bond, "Neutrons from high-energy X-ray medical accelerators. Anestimate of risk to the radiotherapy patient," Med. Phys., Vol.11, pp.231-241, 1984. https://doi.org/10.1118/1.595497
  7. 박병석, 안종호, 권동열, "10 MV 이상 고에너지 치료 시 발생되는 광중성자의 영향", 대한방사선치료학회지, 제25권, 제1호, pp.9-14, 2013.
  8. 이화중, 김대영, 김영택, "10MV 이상 고에너지 사용시 wedge filler의 방사화가 작업환경에 미치는 영향평가", 대한방사선치료학회지, 제16권, 제2호, pp.69-79, 2004.
  9. Omar Chibani, and Chang-Ming Charlie Ma., "Photonuclear does calcula ions for high energy photon beams from Simens and Varian lineacs," Medical Physics, Vol.30, No.8, pp.1990-2000, 2003. https://doi.org/10.1118/1.1590436
  10. 김대섭, 김유현, "방사선치료 장치 및 관련시설에서의 산란중성자에 관한 연구", 대한방사선치료학회지, 제17권, 제2호, pp.141-145, 2005.
  11. ICRP(74), "1995 Conversion coefficients for use in radiological protection a gainst external radiation," International Commission on Radiological Protection, ICRP Publication 74.
  12. G. Bauman and R. B. Rumble, "Intensityodulated Radiotherapy in the Treatment of Prostate Cancer," Clinical oncology : a journal of the Royal College of Radiologists, Vol.24, No.7, pp.461-473, 2012. https://doi.org/10.1016/j.clon.2012.05.002
  13. P. M. McCloskey, S. Atallah, and L. Coate, "Comparison of 3D Conformal Radiation Therapy (3D CRT) and Intensity Modulated Radiation Therapy (IMRT) in Stage III Non-small Cell Lung Cancer (NSCLC)," International journal of radiation oncology, biology, physics, Vol.84, No.3, suppl, pp.S597-S598, 2012.
  14. S. Sharma, J. T. Whaley, and W. Zou, "Incidental Nodal Irradiation in Stage III Lung Cancer Treated With Involved-Field Radiation: Comparison Between 3D CRT and IMRT," International journal of radiation oncology, biology, physics, Vol.84, No.3, suppl, p.S564, 2012.
  15. 정제호, 의료용 선형가속기에서 생성되는 광중성자에 의한 선량평가, 한양대학교 대학원, 원자력공학과 학위논문집, pp.1-56, 2005.
  16. M. H. Lee, "The Effect of Scattering Dose on the Thyroid During Mammography," Journal of the Korean institute of electronic material engineers, Vol.23, No.10, pp.826-830, 2010. https://doi.org/10.4313/JKEM.2010.23.10.826
  17. 이혜림, 김현영, 최형욱 이혜미, 임창선, 파노라마촬영 시 조직등가물질을 이용한 갑상선보호대의 갑상선 피폭선량 감소효과", Journal of the Korea Academia-Industrial cooperation Society, Vol.13, No.5, pp.2278-2284, 2012. https://doi.org/10.5762/KAIS.2012.13.5.2278
  18. ICRP(1991), "1990 Recommendations of the Internaitonal Commission on Radiological Protection," Internaional Commission on Radiological Protection, ICRP Publicaton 60.
  19. ICRP(1996), "1995 Conversion coefficients for use in radiological protect ion a gainst external radiation," International Commission on Radiological Protection, ICRP Publication 74.