DOI QR코드

DOI QR Code

Interference Issuses of Radio Frequency Identification Devices in Magnetic Resonance Imaging Systems and Computed Tomography Scan

  • Periyasamy, M. (Syed Ammal Engineering College) ;
  • Dhanasekaran, R. (Syed Ammal Engineering College)
  • Received : 2015.05.07
  • Accepted : 2015.08.11
  • Published : 2015.09.30

Abstract

We evaluated certain issues related to magnetic resonance imaging (MRI) coupled with the use of active 2.5 GHz radio frequency identification (RFID) tags for patient identification using low field (0.3 T) MRI and computed tomography (CT) scans. We also investigated the performance of the RFID reader located outside the MRI room by considering several factors. A total of ten active RFID tags were exposed to several MRI sequences and X-rays of CT scan. We found that only card type active RFID tags are suitable for patient identification purpose in MRI environment and both wristbands as well as card tags were suitable for the same in CT environment. Severe artifacts were found in the captured MRI and CT images when the area of the imaging was in proximity to the tags. No external factors affected the performance of active RFID reader stationed outside the MRI scan room.

Keywords

References

  1. F. G. Shellock, J. Magn. Reson. Imaging 16, 721 (2002). https://doi.org/10.1002/jmri.10207
  2. F. G. Shellock and J. V. Crues, Radiol. 232, 635 (2004). https://doi.org/10.1148/radiol.2323030830
  3. T. O. Woods, J. Magn. Reson. Imaging 26, 1186 (2007). https://doi.org/10.1002/jmri.21140
  4. F. G. Shellock and A. Spinazzi, Am. J. Roentgenol. 191, 1 (2008). https://doi.org/10.2214/AJR.08.1098
  5. F. G. Shellock, T. O. Woods, and J. V. Crues, Radiol. 253, 26 (2009). https://doi.org/10.1148/radiol.2531091030
  6. K. Ohashi, S. Ota, L. O. Machado, and H. Tanaka, Comput. Biol. Med. 40, 545 (2010). https://doi.org/10.1016/j.compbiomed.2010.03.007
  7. M. Cavalleri, R. Morstabilini, and G. Reni, Conf. Proc. IEEE Eng. Med. Biol. Soc. 5, 3278 (2004).
  8. C. C. Liu, C. H. Chang, M. C. Su, H. T. Chu, S. H. Hung, J. M. Wong, and P. C. Wang, Comput. Meth. Prog. Bio. 104, 435 (2011). https://doi.org/10.1016/j.cmpb.2010.08.017
  9. P. Peris-Lopez, A. Orfila, A. Mitrokotsa, and J. C. Van der Lubbe, Int. J. Med. Informs. 80, 13 (2011). https://doi.org/10.1016/j.ijmedinf.2010.10.008
  10. E. Etchells, C. O. Neill, and M. Bernstein, World J. Surg. 27, 936 (2003). https://doi.org/10.1007/s00268-003-7097-2
  11. G. Calcagnini, F. Censi, M. Maffia, L. Mainetti, E. Mattei, L. Patrono, and E. Urso, IEEE Trans. Inf. Technol. Biomed. 16, 1051 (2012). https://doi.org/10.1109/TITB.2012.2204895
  12. A. Oztekin, F. M. Pajouh, D. Delen, and L. K. Swim, Decis. Support. Syst. 49, 100 (2010). https://doi.org/10.1016/j.dss.2010.01.007
  13. W. Zhou and S. Piramuthu, Decis. Support. Syst. 50, 222 (2010). https://doi.org/10.1016/j.dss.2010.08.003
  14. E. W. T. Ngai, J. K. L. Poon, F. F. C. Suk, and C. C. Ng, Inf. Syst. Front. 11, 405 (2009). https://doi.org/10.1007/s10796-009-9154-3
  15. T. Steffen, R. Luechinger, S. Wildermuth, C. Kern, C. Fretz, J. Lange, and F. H. Hetzer, Patient. Saf. Surg. 4, 1 (2010). https://doi.org/10.1186/1754-9493-4-1
  16. B. Titterington and F. G. Shellock, Magn. Reson. Imaging 31, 1439 (2013). https://doi.org/10.1016/j.mri.2013.04.005
  17. C. Y. Cheng and J. W. Chai, J. Med. Syst. 36, 3423 (2012). https://doi.org/10.1007/s10916-011-9796-9
  18. J. Lamberg, http://www.rfidjournal.com. Accessed July 12, 2012.
  19. M. Periyasamy and R. Dhanasekaran, Scientific World Journal, 1 (2014).
  20. M. Periyasamy and R. Dhanasekaran, Biomed. Eng.-App. Bas. C 26, 1 (2014).
  21. X. Fei, S. Lei, S. Gao, L. Wei, and L. Wang, Biomed. Eng. Online 13, 1 (2014). https://doi.org/10.1186/1475-925X-13-1
  22. Fiocchi Serena, Parazzini Marta, Paglialonga Alessia, and Ravazzani Paolo, Bioelectromagnetics 32, 408 (2011). https://doi.org/10.1002/bem.20653
  23. N. R. Jagannathan, Biophys. 36, 341 (1999).
  24. M. A. Baker and I. Macdonald, Vet. Radiol. Ultrasoun. 52, 161 (2011). https://doi.org/10.1111/j.1740-8261.2010.01762.x

Cited by

  1. Differential Absorption Analysis of Nonmagnetic Material in the Phantom using Dual CT vol.21, pp.2, 2016, https://doi.org/10.4283/JMAG.2016.21.2.286
  2. Measurement of MRI Monitor Luminance and MRI Room Illuminance with a Light Probe vol.26, pp.5, 2016, https://doi.org/10.4283/JKMS.2016.26.5.168
  3. A study on a pedicle-screw-based reduction method for artificially reduced artifacts vol.71, pp.5, 2017, https://doi.org/10.3938/jkps.71.304