Abstract

Objectives : Recently, people who have normal brachial blood pressure(BP) are being threatened by high-risk disease such as stroke. The aim of this study is to suggest that new method to assess systemic circulation. It can be performed by analyzing optimal blood pulse wave on 3 sites belonging to subjects that have normal BP. Methods : We respectively extracted main peaks(h1) of optimal blood pulse wave on left/right temporal artery(LR1=h1), radial artery(LR2=h1) and dorsalis pedis artery(LR3=h1). We obtained h1 from 30 subjects who are discreetly chosen and have normal BP. Main peak(h1) can be extracted by using 3D pulse imaging analyser(DMP-1000+, DAEYOMEDI Co., Korea) that has 5-level pressure method. We analyzed the ratio of [LR1/LR2] and [LR3/LR2]. Results : In the case of male group, the results are [LR1/LR2=0.7100.177] and [LR3/LR2=0.9290.317]. In the case of female group, the results are [LR1/LR2=0.6680.121] and [LR3/LR2=0.7050.195]. Especially, it is statistically verified that the result of ratio [LR3/LR2] is much higher in male group than in female group(p<0.05). Conclusions : We suggested the standard ratio of [LR1/LR2] and [LR3/LR2] for normal subjects, respectively. It can be adopted as a new method to evaluate the systemic circulation.

Keywords

References

  1. Ji NK, Lee KS, Mun SK, Ko CN, Cho KH, Kim YS, et al. Differenting Effects of Cerebral Hemispheric Lateralization and Gender by Power Spectral Analysis of Heart Rate Variability in Cerebrovascular disease. Korean Journal of Oriental Internal Medicine. 1998 ; 19(2) : 7-16.
  2. Jeong JH, Sun JJ, Hong JW, Park SW, Jeng WS, Mun SK, et al. Sex Differences in Patients with First-ever Stroke(in terms of stroke types, risk factors, warning signs and stroke complications). Journal of Korean Medicine. 2007 ; 28(3) : 207-15.
  3. Kim HS, Kwen WH, Mun KC, Lee IW. Diagnosis of Primary Aldosteronism and Usefulness of Aldosterone/Renin Ratio in Secondary Hypertension. Korea Society of Nuclear Medicine Technology. 2008 ; 12(3) : 241-6.
  4. Lee DT, Lee YS. The Age-Related Trend in Blood Pressure and the Prevalence of Hypertension in Korean Adults. Journal of Life Science. 2012 : 22(2) : 148-55. https://doi.org/10.5352/JLS.2012.22.2.148
  5. Kim KY, Bae MH, Pakr SM, Kim HN, Yang YY, Jang SY, et al. Case Reports : Cardiology: Acute Myocardial Infarction with Cardiogenic Shock in a Young Adult with Minimal Change Disease. Korean journal of internal medicine. 2006 ; 90(4) : 326-30.
  6. Lee HD, Seong DH. Paradoxical Cerebral Embolism in a Young Patient - A case report -. Journal of Korean Academy of Rehabilitation Medicine. 2005 ; 29(5) : 527-30.
  7. Park JK, Lee SH, Choi SH, Jeong JH, Lee NH. Two patients with acute myocardial infarction presenting with simultaneous acute ischemic stroke. Korean journal of internal medicine. 2008 ; 74(6) : 672-5.
  8. Sun SH, Ko HY, Park JH, Jen CY. Delphi Study For Treatment Standardization of Coldness in hands and Feet. The Society of Korean Preventive Medicine. 2015 ; 19(3) : 131-40.
  9. Cho JY, Kim JW, Park KS, Lee CH, Cho JH, Jang JB, et al. A Case Report of Cold Sensitivity of Hands and Feet Patient. The Journal of Oriental Obstetrics and Gynecology. 2011 ; 24(3) : 195-202.
  10. Mun IS. Clinical Characteristics of Lower Limb Ischemia due to Multi-Level Arterial Occlusive Disease. Korean Journal of Vascular and Endovascular Surgery. 2006 ; 22(2) : 141-3.
  11. Cho JY, Kim JW, Park KS, Lee CH, Jang JB, Lee KS, et al. A Case Report of Cold Sensitivity of Hands and Feet Patient. The Journal of Oriental Obstetrics and Gynecology. 2011 ; 24(3) : 195-202.
  12. Kim PK, Cho SH, Kim BS, Kang MH, An CB. Blood Vessel Aging Diagnosis System using Blood Pressure and Photoplethysmography. The Transactions of the Korean Institute of Electrical Engineers. 2008.
  13. Lee NB, Im JJ, Park YB, Chen YJ. Development of a Pulse Wave Velocity Measurement System and Assessment of the System Reproducibility for the Diagnosis of Arteriosclerosis. The Korea Institute of Oriental Medical Diagnostics. 2005 ; 9(1) : 112-24.
  14. Jang JH, Park JM, Kim SS. Usefulness of Brachial-Ankle Pulse Wave Velocity and Ankle-Brachial Index as Predictors of Early Age-Related Maculopathy. Korean Journal of Opthalmology. 2011 ; 52(6) : 696-701. https://doi.org/10.3341/jkos.2011.52.6.696
  15. Kwen SM, Kang HJ, Lim YK, Lee YH. Analysis of Arterial Stiffness by Age Using Pulse Waveform Measurement of 5-levels Graded Pressure. Korean Journal of Acupuncture. 2010 ; 27(2) : 107-20.
  16. Lee YH. -Basic research for diagnosis of hypertension- Correlation of Pulse Wave Variation and change of Blood Pressure(BP) and Analysis of BP. Research Report. Yonsei University. 2012.
  17. Kang HJ, Kwen YS, Kim DR, Kim KC, Lim YK. A study on wiry pulse in hypertensive patients analyzed at 5 levels of applied pressure using 3 dimensional pulse imaging analyzer. Korean Journal of Acupuncture. 2010 ; 27(1) : 1-12.
  18. Kwen SM, Kang HJ, Lim YK, Lee YH. A study on floating and sinking pulse by classification of pulse pattern through analysis of P-H volume-curve at 5 applied pressure levels. Korean Journal of Acupuncture. 2010 ; 27(1) : 13-22.
  19. Bae JH, CHen YJ, Kim JY, Kim JW. Novel Detection Algorithm of The Upstroke of Pulse Waveform for Continuously Varying Contact Pressure Method. Journal of semiconductor technology and science. 2012 ; 49(2) : 46-54.
  20. Heo JE, Kim YK, Kwen JN, Kim KM, Kim BH, Kim MK, et al. Case-control Study : Cerebral Blood Flow as Measured by Transcranial Doppler Ultrasonography(TCD) in Hypertensives. Korean Journal of. Oriental Internal Medicine. 2008 ; 29(4) : 950-61.
  21. Lee YH, Ko SB, Jeong DM. Diagnosis parameters extraction by correlativity analysis of blood pressure(BP) and head blood pressure(HBP) and Development of multi-function automatic blood pressure monitor. Journal of semiconductor technology and science. 2003 ; 40(6) : 58-67.