DOI QR코드

DOI QR Code

Commissioning and Validation of a Dedicated Scanning Nozzle at Samsung Proton Therapy Center

  • Chung, Kwangzoo (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Han, Younyih (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Ahn, Sung Hwan (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
  • Kim, Jin Sung (Department of Radiation Oncology, Yonsei Cancer Center, Yonsei University College of Medicine) ;
  • Nonaka, Hideki (Division of Industrial Equipment, Sumitomo Heavy Industries Ltd.)
  • Received : 2016.12.16
  • Accepted : 2016.12.24
  • Published : 2016.12.31

Abstract

In this study, we present the commissioning and validation results of a dedicated scanning nozzle. The dedicated scanning nozzle is installed in one of the two gantry treatment rooms at Samsung Proton Therapy Center. Following a successful completion of the acceptance test, the commissioning process including the beam data measurement for treatment planning system has been conducted. Extended measurements have been conducted as a validation of the clinical performance of the nozzle and various quality assurance protocols have been prepared.

Keywords

References

  1. R. R. Wilson: Radiology 47, 487 (1946) https://doi.org/10.1148/47.5.487
  2. M. Jermann: Int J Particle Ther. 2, 50 (2015) https://doi.org/10.14338/IJPT-15-00013
  3. K. Chung, Y. Han, J. Kim et al: Radiat Oncol J. 33, 337 (2015) https://doi.org/10.3857/roj.2015.33.4.337
  4. K. Chung, J. Kim, D.-H. Kim, S. Ahn, and Y. Han: J. Korean Phys Soc. 67, 170 (2015) https://doi.org/10.3938/jkps.67.170

Cited by

  1. Development of an Analytic Software Using Pencil Beam Scanning Proton Beam vol.28, pp.1, 2016, https://doi.org/10.14316/pmp.2017.28.1.22
  2. A Pilot Study of the Scanning Beam Quality Assurance Using Machine Log Files in Proton Beam Therapy vol.28, pp.3, 2016, https://doi.org/10.14316/pmp.2017.28.3.129
  3. Increased Efficiency of Range Verification in Routine QA for Pencil-Beam Scanning Proton Therapy vol.73, pp.7, 2018, https://doi.org/10.3938/jkps.73.983
  4. Feasibility study of Fe3O4/TaO x nanoparticles as a radiosensitizer for proton therapy vol.63, pp.11, 2016, https://doi.org/10.1088/1361-6560/aac27b