DOI QR코드

DOI QR Code

Testing Evaluation for Stand-Alone Cervical Cage

경추용 자립형 케이지의 실험적 평가

  • Baek, Myong-Hyun (Medical Device Development Center, Osong Medical Innovation Foundation) ;
  • Lee, Mun-Hwan (Medical Device Development Center, Osong Medical Innovation Foundation)
  • 백명현 (오송첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 이문환 (오송첨단의료산업진흥재단 첨단의료기기개발지원센터)
  • Received : 2015.12.31
  • Accepted : 2016.07.26
  • Published : 2016.11.01

Abstract

Stand-alone cervical cage consists of a PEEK body, Ti plate, and screw, which are configured as a single-piece. Through a single operation, this implantable medical device is capable of completely fixing the cervical vertebral body. For example, instead of a plate, which is normally used, the intervertebral disc is removed and replaced with a cervical cage. It should be noted that in Korea, KFDA guidelines for a stand-alone cervical cage have not yet been suggested. Therefore, the aim of this study is to present the systematic study of the static compression test, static torsion test, dynamic compression test, and expulsion test. Further, the test method is designed to refer to the ASTM standard and relative literature.

Keywords

References

  1. Health Insurance Review & Assessment Service. "In 2014, One Quarter of the Population of South Korea got Treatment for Spinal Disease," 2014.
  2. Jeon, T.-S., Chang, H., and Choi, B.-W., "Current Concept on the Operative Treatment for Degenerative Cervical Disc Disease," Journal of the Korean Medical Association, Vol. 54, No. 9, 2011.
  3. ASTM F2077-11, "Test Methods for Intervertebral Body Fusion Devices," 2011.
  4. Goel, V. K., Dick, D., Jackson, M., Kuroki, H., and Ebraheim, N., "Expulsion Testing of an Intervertebral Fusion Cage: Effect of Tooth Design on Pull out Resistance," Proc. of the Summer Bioengineering Conference, pp. 111-112, 2003.