DOI QR코드

DOI QR Code

Effective of Body Temperature Increasing during Brain MRI scan

MRI 검사 시 체온상승 효과: 1.5 T vs 3.0 T

  • Received : 2017.02.10
  • Accepted : 2017.02.28
  • Published : 2017.02.28

Abstract

As the Radiofrequency(RF) increases with the magnetic field strength, the wavelength of the RF excitation field becomes smaller, which leads to more the thermal effect in the human-body placed in the electric field. MRI scanner used was GE signa 1.5T, HDx 3.0T and Philips 3.0T with same routine clinical sequence protocol. Therefore temperature was measured before and after each scan. Taken the temperatures in the ear with ear infra-red type thermometer(Braun co). 3.0T were temperature increases more than $0.15^{\circ}C$ and GE 3.0T MRI equipment about $0.14^{\circ}C$ higher than the Philips 3.0T MRI(p<0.012). Psychogenic status was investigated by the survey respondents about their status can not just answer therefore, a little different from the expected. In our study of Thermal effect of clinical MRI with clinical protocol sequence, we found that the 3.0T in the body-temperature rise was greater than the 1.5T. Therefore, in clinical 3.0T examine the dangerous situation caused by the temperature rise occurred (burns, impaired thermoregulatory mechanism in patients with high-temperature damage, exhaustion occurs due to excessive sweating), not to appear the more watched the patient's condition with procedure.

MRI장비에서 자장의 세기가 증가 하게 되면 사용하는 RF(Radiofrequency) pulse 또한 증가 하게 되고 이는 MRI 장비 안에 놓인 인체의 체온 상승을 증가 시키게 하는 역할을 하게 된다. 이에 국소부위에서의 열 발생을 알아보고자 젤라틴과 pork sample 이용하여 측정하였다. 본 연구는 2014년 12월 21일부터 2015년 8월 14일까지 153명의 환자를 대상으로 실시하였다. 3대의 MRI(1.5T- 1대, 3.0T- 2대)장비를 사용하여 뇌 또는 간 부위에서 일상적으로 쓰이는 sequence 프로토콜을 동일하게 적용하여 검사하였다. 검사 전 후 체온 측정은 적외선 타입의 귀 체온계(Braun社)를 사용하였으며 대상자의 심리적 상태는 직접 설문을 통하여 파악했다. 임상 환자를 대상으로 한 체온 상승 결과를 보면 3.0T가 1.5T에서보다 평균 $0.15^{\circ}C$정도 높았고(p<0.012) 3.0T내 에서도 Philips제조회사에서 보다 GE社 MRI장비에서 $0.14^{\circ}C$정도 더 높았다. 심인성 상태에 따른 결과를 살펴보면 MRI검사 진행 중 나는 소리에 대한 민감성 정도와 체온상승과의 관계는 무관하였고, 폐쇄성에 대한 응답이 긴장감으로 느꼈다고 대답하는 사람일수록 체온이 더 상승하는 경향을 보였다. 자장의 세기가 높은 MRI장비일수록 RF 반응물질(물, 금속물질)에 의한 화상이나 체온상승으로 인한 위험한 상황발생(체온조절 장해 환자의 경우 고온 손상, 과다 땀 발생으로 인한 탈진)이 나타나지 않도록 환자의 상태를 좀 더 예의 주시하며 MRI검사를 진행할 필요가 있겠다. 높은 자기장을 기반으로 한 MRI장비는 인체에 흡수되는 전자파 흡수율인 SAR를 비례적으로 증가시키므로 앞으로는 RF 코일 성능을 향상하거나 영상의 질을 향상시키기 위한 이미지 프로그램을 개발하는 등 자기장 이외의 방법을 강구하는 것이 필요하다.

Keywords

References

  1. MRI AND THE PHYSICAL AGENTS(EMF) DIRECTIVE: AN INSTITUTE OF PHYSICS REPORT. 2008. https://www.iop.org/publications/iop/2008/file_3821 5.pdf
  2. M. G. Restivo, C. A. van den Berg, A. L. van Lier, D. L. Polders. A. J. Raaiijmakers, P. R. Luijten, H. Hooqduin. "Local specific absortion rate in brain tumors at 7 tesla". Magnetic Resonance Medicine. Vol. 7., No. 1. pp. 381-389. 2016.
  3. Z. Wang, J. C. Lin. "Partial-Body SAR Calculations in Magnetic-Resonance Image (MRI) Scanning Systems [Telecommunications Health and Safety]". IEEE Antennas and Propagation Magazine. Vol. 54. No. 2. pp. 230-237. 2012. https://doi.org/10.1109/MAP.2012.6230763
  4. N. wertheimer and E. Leeper. "Electrical wiring configurations and childhood cancer". American Journal of Epidemiolgy". Vol. 109, No. 3. pp. 273-284. 1979. https://doi.org/10.1093/oxfordjournals.aje.a112681
  5. S. X. Xin, Q. Huang, Y. Gao, B. Li, Y. Xu, W. Chen. "Fetus MRI at 7 T: B1 Shimming strategy and SAR safety implications". IEEE Transactions on microwave theory and techniques. Vol. 61, No. 5. pp. 2146-2152. 2013. https://doi.org/10.1109/TMTT.2013.2247053
  6. S. Romeo, A. Sannino, O. Zeni, M. R. Scarfi, R. Massa, V. Cerciello, R d'Angelo. "Induced electric fields and currents in the body by movements in a MRI facility: A numerical analysis". IEEE. Vol. 2015.
  7. B. A. Chronik and M. Ramachandran. "Simple anatomical measurements do not correlate significantly to individual peripheral nerve stimulation thresholds as measured in MRI gradient coils". Journal of Magnetic Resonance Imaing. Vol. 17. No. 6, pp. 716-721. 2003. https://doi.org/10.1002/jmri.10300
  8. M. S. Kim. "Investigation of Factors Affecting Body Temperature Changes During Routine Clinical Head Magnetic Resonance Imaging". Iran Journal of Radiology. Vol. 13, No. 4. e34016. 2016.
  9. H. A. Sharma. "MRI physics-basic principles". Acta Neuropsychiatrica. Vol. 21. No. 4. pp. 200-201. 2009. https://doi.org/10.1111/j.1601-5215.2009.00404.x
  10. A. Hirata and O Fujiwara. "The correlation between mass-averaged SAR and temperature elevation in the human head model exposed to RF near-fields from 1 to 6 GHz". Physics in medicine and biology, Vol. 54. No. 23. 2009.
  11. J. Jerrolds and Shane. Keene. "MRI safety at 3T versus 1.5T". The internet journal of world health and societal politics. Vol. 6. No. 1. 2009.
  12. K. B. Baker, J. A. Tkach, J. A. Nyenhuis, M. Phillips, F. G. Shellock, J. Gonzalez-Martinez, A. R. Rezai. "Evaulation of specific absorption rate as a dosimeter of MRI-related implant heating". Journal of Magnetic Resonance of Imaging. Vol. 20, No. 2. pp. 315- 320. 2004. https://doi.org/10.1002/jmri.20103
  13. H. Muranaka, T. Horiguchi, S. Usui, Y. Ueda, O. N akamura, F. Ikeda. "Dependence of RF heating on SAR and implant position in a 1.5T MR system". Magnetic Resonance in medical sciences. Vol. 6. No. 4. pp. 199-209. 2007. https://doi.org/10.2463/mrms.6.199
  14. A. Kangarlu, F. G. Shellock, D. W. Chakeres. "8.0-Tesla human MR system: Temperature changes associated with radiofrequency-induced heating of a head phantom". Journal of Magnetic Resonance Imaging. Vol. 17. No. 2. pp. 220-226. 2003. https://doi.org/10.1002/jmri.10236
  15. A. Ahlbom, A. Green, L. Kheifets, D. Savitz, A. Swerdlow. "Epidemiology of health effects of radiofrequency exposure". Environmental health perspectives. Vol. 112. No. 17. pp. 1741-175. 2004. https://doi.org/10.1289/ehp.7306