Cone beam Computed Tomography(CBCT) is an increasing trend in clinical applications due to its ability to increase the accuracy of radiation therapy. However, this leaded to an increase in exposure dose. In this study, the simulation using Monte Carlo method is performed and the absorbed dose of CBCT is analyzed and standardized data is presented. First, after simulating the CBCT, the photon spectrum was analyzed to secure the reliability and the absorbed dose of the tissue in the human body was evaluated using the MIRD phantom. Compared with SRS-78, the photon spectrum of CBCT showed similar tendency, and the average absorbed dose of MIRD phantom was 8.12 ~ 25.88 mGy depending on the body site. This is about 1% of prescription dose, but dose management will be needed to minimize patient side effects and normal tissue damage.

Keywords

References

  1. Park. CJ, Park. SH, Kim. JS, Han. YY, Ju. SG, Shun. EH, et al. Digital Tomosynthesis for Patient Alignment System Using Half-fan Mode CBCT Projection Images. Korean Journal of Medical Physics. 2010;21(4):360-67.
  2. Do. KM, Jung. DY, Kim, YB. Patient Position Verification and Corrective Evaluation Using Cone Beam Computed Tomography (CBCT) in Intensity. modulated Radiation Therapy. The Journal of Korean Society for Radiation Therapy. 2009;21(2):83-8.
  3. Bang. SJ, Kim. YY, Jeong. IS, Kim. JS, Kim. YG. A Study of Image Quality and Exposed Dose by Field Size Changing on CBCT. The Journal of Korean Society for Radiation Therapy. 2013;25(2):175-180.
  4. Oh. SJ, Kim. SY, Suh. TS. How Image Quality Affects Determination of Target Displacement When Using kV Cone-beam Computed Tomography (CBCT). Korean Journal of Medical Physics. 2006;17(4):207-211.
  5. Yoo. YS, Lee. HJ, Kim. DY, Yu. R. Implementation of KV Cone Beam CT for Image Guided Radiation Therapy. The Journal of Korean Society for Radiation Therapy. 2007;19(1):43-9.
  6. Lee. IS, Yoo. EJ, Kim. SK, Choi. KS, Lee. JW, Suh. TS, et al. A Study on Mechanical Errors in Cone Beam Computed Tomography(CBCT) System. Journal of Radiological Science and Technology. 2013;36(2):123-129.
  7. Park. BS, Ahn. JH, Kim. JS, Song. KW. Evaluation of Every Other Day - Cone Beam Computed Tomography in Image Guided Radiation Therapy for Prostate Cancer. 2014;26(2):289-295.
  8. Semelka RC, Armao DM, Elia J, Huda W. Imaging strategies to reduce the risk of radiation in CT studies, including selective substitution with MRI. Journal of Magnetic Resonance Imaging. 2007;25(5):900-9. https://doi.org/10.1002/jmri.20895
  9. Song. YG. Diagnostic Radiation Imaging Detector Implementation and Image Evaluation using Monte Carlo Simulation. [master's thesis]. Kyeong Sang Nam Do:Inje University. 2013.
  10. Lee. JK. The study of radiation dose estimates for computed tomography using Monte Carlo method. [master's thesis]. Kyeong-Ki Do:Kyung Hee University. 2011.
  11. Lee. SH. A comparative study of aluminum, copper, and lead as shielding materials in 6 MeV electron therapy. [master's thesis]. Jeon-La Buk Do:Chonbuk National University. 2013.
  12. Lee. JO, Jeong. DH. Calculation of Energy Spectra for 6 MeV Electron Beam of LINAC Using MCNPX. Korean Journal of Medical Physics. 2006;17(4):224-231.
  13. Lim. HJ. Status of Domestic Linear Accelerator. Kang-Won. Health Insurance Review & Assessment Service. 2017.
  14. Fouad AA. On Board Cone Beam CT for Treatment Planning in Image Guided Radiotherapy. [dissertation]. Saudi Arabia:King Abdulaziz University. 2015.
  15. Downes P, Jarvis R, Radu E, Kawrakow I, Spezi E, "Monte Carlo simulation and patient dosimetry for a kilovoltage cone-beam. Medical Physics. 2009;36(9):4156-4167. https://doi.org/10.1118/1.3196182
  16. Han EY, Wesley EB, Eckerman KF. Revisions to the ORNL serise of adult and pediatric computational phantoms for use with the MIRD schema. University of Florida Health Phys. 2006;90(4)337-356.
  17. Valentin J. ICRP Publication 89:Basic Anatomical and Physiological Data for Use in Radiological Protection:Reference Values. Elsevier Publishing Company. 2003.
  18. White. DR, Griffith. RV, Wilson. IJ. ICRU REPORT 46: Photon, Electron, Proton and Neutron Interaction Data for Body Tissues. Oxford Univ. Pr. UK. 1992.
  19. Park DO. Development of Computational Procedure for Assessment of Patient Dose in Multi-Detector Computed Tomography. [master's thesis]. Seoul: Hanyang University. 2007.
  20. Park. I, Kim. KH, Oh. SC, Song. JY, Kwon. OH, Kim. KP. Evaluation of Radiation Dose to Patients in Intraoral Dental Radiography Using Monte Carlo Method. Journal of Radiation Industry. 2016;10(3):139-144.
  21. Ding. GX, Duggan. DM, Coffey. CW. Characteristics of kilovoltage x-ray beams used for cone-beam computed tomography in radiation therapy. Physics in Medicine and Biology. 2007;52(6):1595-1615 https://doi.org/10.1088/0031-9155/52/6/004
  22. Palm A, Nilsson E, Herrnsdorf L. Absorbed dose and dose rate using the Varian OBI 1.3 and 1.4 CBCT system. Journal of applied clinical medical physics. 2009;11(1):S93.
  23. Moon. YM. A study on the effective dose measurement for cone beam computed tomography using glass dosimeter. Pusan:Dong-A University. 2012.