A Study on the Using of Automatic Exposure Control in the Chest Radiography

흉부 방사선검사에서 자동노출제어 사용에 관한 연구

  • 최성식 (한서대학교 방사선학과) ;
  • 임청환 (한서대학교 방사선학과) ;
  • 정성훈 (한서대학교 방사선학과)
  • Received : 2019.01.30
  • Accepted : 2019.02.26
  • Published : 2019.02.28

Abstract

For general radiological examinations, even in the same area and the same test, the test conditions must be set differently according to the patient. However, since it is impossible to consider the body shape and conditions of patients every time in medical institutions where various patients visit, the tests are conducted by setting the AEC which automatically sets the test conditions. AEC is most commonly used in chest radiography. Therefore, the purpose of this study is to propose the improvement plans for using AEC by measuring the exposure dose and evaluating the image quality according to whether the AEC is used or not, and to provide basic data for AEC research. In the present study, images were acquired while varying tube voltage and test distance according to the use of AEC in chest radiography. The radiation dose was measured by placing the dosimeter in front of the chest phantom, and the CNR and SNR of acquired images were analyzed using Image J. The t-test was conducted for the statistical analysis and the significance was determined at the level of 95%(p<.05). As a result of this study, in the inspection distance (100cm, 140cm, 180cm) according to the use of AEC, high doses were observed when the AEC was used and there was statistically significant difference(p<.05). In the t-test to determine the difference between CNR and SNR depending on whether AEC was used or not, there was no significant difference according to the use of AEC(p>.05). Therefore, when performing chest radiography, if the radiologist establishes the appropriate examination conditions and conducts the examination by not just relying solely on AEC, it may be possible to reduce unnecessary radiation exposure to the patient.

Keywords

References

  1. Kim CS. Image Quality Evaluation of Medical Image Enhancement Parameters in the Digital Radiography System. The Korea Contents Association. 2010;10(6):330-32. https://doi.org/10.5392/JKCA.2010.10.2.330
  2. Ministry of Food and Drug Safety. Guidelines for Standard Techniques in Imaging(general radiography); 2017.
  3. Sund P, Bath, M, Kheddache, S, Mansson, LG. Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiography. European Radiology. 2004;14(1):48-58. https://doi.org/10.1007/s00330-003-1971-z
  4. Jung JS. Effect of Automatic Exposure Control Marker with Chest Radiography in Radiation Reduction. Korea Society of Radiological Science. 2014;37(3):177-85.
  5. Veldkamp WJH, Kroft LJM, Geleijns J. Dose and Perceived image quality in chest radiography. European Journal of Radiology. 2009;72(2):209-17. https://doi.org/10.1016/j.ejrad.2009.05.039
  6. Sin SK. Reducing Methods of Patient's Exposed Dose Using Auto Exposure Control System in Digital Radiography. Korea Society of Radiological Science. 2013;36(2):111-22.
  7. Lee JS. Quantitative Evaluation of Image Quality using Automatic Exposure Control & Sensitivity in the Digital Chest Image. The Korea Contents Association. 2013;13(8):275-83.
  8. Kim KH. A Study on Effects of Automatic Exposure Control for the Radiation Dose and Image Quality [Masters dissertation]. Seoul National University of Technology, Seoul, Korea; 2003.
  9. Kim EY. A Study on the State of Setting Irradiation Conditions and Awareness of Dose and Quality Control of Diagnostic Digital Radiography [Masters dissertation]. Korea University, Seoul, Korea; 2010.
  10. Kim HI. X-ray Absorptiometry Image Enhancement using Sparse Representation. Journal of Korea Multimedia Society. 2012;15(10):1205-11. https://doi.org/10.9717/kmms.2012.15.10.1205
  11. Ko SJ, Lim CH, Koo HK, et al. Introduction to Radiologic Science. Chungkoomoonwasa, Seoul, Korea; 2013.