Analysis of implant strain value exerted using different screw tightening protocols in screw-retained 3-unit prostheses

3본 나사 유지형 임플란트 보철물의 고정 방식에 따른 임플란트 고정체 치경부에 발생하는 변형율 비교분석

  • Kim, Sang-Beom (Department of Prosthodontics, School of Dentistry, Kyungpook National University) ;
  • Lee, Du-Hyeong (Department of Prosthodontics, School of Dentistry, Kyungpook National University) ;
  • Lee, Cheong-Hee (Department of Prosthodontics, School of Dentistry, Kyungpook National University)
  • 김상범 (경북대학교 치과대학 치과보철학교실) ;
  • 이두형 (경북대학교 치과대학 치과보철학교실) ;
  • 이청희 (경북대학교 치과대학 치과보철학교실)
  • Received : 2020.08.19
  • Accepted : 2020.10.06
  • Published : 2020.10.30

Abstract

Purpose: The purpose of this study was to measure and compare the strain value exerted on the cervical area using different screw tightening protocols in implant-supported, screw-retained 3-unit prostheses. Materials and methods: Strain gauges were attached to four implants: two external and two internal. Thereafter, two study model were designed each type using acrylic resin. CAD-CAM was used to design hex and nonhex abutments for each group (EH, ENH, IH, and INH group) and Screw-cement-retained prostheses were also designed using a nonprecious base metal. Abutment was fixed with 10 Ncm torque, and the prosthesis was cemented. Screws were fixed with 30 Ncm torque using different three protocols. After 5 min, the strain gauge level was measured, and group analysis was performed (α=.05). Results: External group showed significantly lower strain values than internal group and the EH group showed significantly lower strain values than the ENH group (P<.05). There was no difference in strain value based on the types of screw tightening protocols in same group (P>.05). The IH group exhibited significantly higher strain values than the INH group and the IH group showed a significant difference in strain values based on the types of screw tightening protocols used (P<.05). Conclusion: There was no significant effect on the external type in the implant-supported, screw-retained prostheses. However, strain values were high in the internal type, and the types of screw tightening protocol significantly affected these implants.

목적: 본 연구의 목적은 3본 나사 유지형 임플란트 보철물의 나사 고정 방식에 따른 치경부에 발생하는 변형율을 측정하고 비교하기 위함이다. 재료 및 방법: 2가지 종류의 임플란트(외부연결방식, 내부연결방식)를 각각 2개씩 4개 임플란트 고정체 경부에 스트레인 게이지를 부착한 후, 각 종류의 아크릴릭 레진 모형 2개를 제작하였다. CAD-CAM을 이용하여 hex 및 nonhex 지대주를 제작하였고,(EH, ENH, IH& INH 군) 나사-시멘트 유지보철 개념으로 보철물을 비귀금속으로 제작하였다. 각 지대주를 10 Ncm 토크로 고정한 후, 보철물을 합착하였다. 나사를 3가지의 다른 방법으로 30 Ncm까지 고정한 다음, 5분 후 변형율을 측정한 후 평균비교 및 군간 상호작용을 평가하였다 (α = .05). 결과: 외부연결방식 군은 내부연결방식 군에 비하여 유의하게 낮은 변형율을 나타냈으며, EH군이 ENH군보다 유의하게 낮은 변형율을 나타내었으나 (P < .05), 나사 고정 방법에 따른 차이는 없었다 (P > .05). IH군이 INH군에 비하여 유의하게 높은 변형율을 나타내었으며 IH군에서만 나사 고정 방법에 따라서도 차이가 있었다 (P < .05). 결론: 나사 유지형 임플란트 보철물에서는 외부연결방식인 경우에는 큰 영향이 없으나, 내부연결방식에서는 보다 큰 변형율이 발생하고 나사 고정 방법도 영향을 주는 것으로 나타났다.

Keywords

References

  1. Li J, Jansen JA, Walboomers XF, van den Beucken JJ. Mechanical aspects of dental implants and osseointegration: A narrative review. J Mech Behav Biomed Mater 2020;103:103574. https://doi.org/10.1016/j.jmbbm.2019.103574
  2. Nishioka RS, Nishioka LN, Abreu CW, de Vasconcellos LG, Balducci I. Machined and plastic copings in three-element prostheses with different types of implant-abutment joints: a strain gauge comparative analysis. J Appl Oral Sci 2010;18:225-30. https://doi.org/10.1590/S1678-77572010000300005
  3. de Vasconcellos LG, Kojima AN, Nishioka RS, de Vasconcellos LM, Balducci I. Axial loads on implant-supported partial fixed prostheses for external and internal hex connections and machined and plastic copings: strain gauge analysis. J Oral Implantol 2015;41:149-54. https://doi.org/10.1563/AAID-JOI-D-10-00174
  4. Asvanund P. A strain gauge analysis comparing external and internal implant-abutment connections. Implant Dent 2014;23:206-11. https://doi.org/10.1097/ID.0000000000000063
  5. Albrektsson T, Dahlin C, Jemt T, Sennerby L, Turri A, Wennerberg A. Is marginal bone loss around oral implants the result of a provoked foreign body reaction? Clin Implant Dent Relat Res 2014;16:155-65. https://doi.org/10.1111/cid.12142
  6. Nishioka RS, de Vasconcellos LG, de Melo Nishioka LN. External hexagon and internal hexagon in straight and offset implant placement: strain gauge analysis. Implant Dent 2009;18:512-20. https://doi.org/10.1097/ID.0b013e3181bcc621
  7. Nissan J, Gross M, Shifman A, Assif D. Stress levels for wellfitting implant superstructures as a function of tightening force levels, tightening sequence, and different operators. J Prosth Dent 2001;86:20-3. https://doi.org/10.1067/mpr.2001.115182
  8. Watanabe F, Uno I, Hata Y, Neuendorff G, Kirsch A. Analysis of stress distribution in a screw-retained implant prosthesis. Int J Oral Maxillofac Implants 2000;15:209-18.
  9. Al Quran FA, Rashdan BA, Zomar AA, Weiner S. Passive fit and accuracy of three dental implant impression techniques. Quintessence Int 2012;43:119-25.
  10. Luk HW, Pow EH, McMillan AS, Hui CF. A double casting technique to minimize distortion when constructing fixed partial dentures on implants. J Prosth Dent 2004;91:93-6. https://doi.org/10.1016/j.prosdent.2003.09.013
  11. Karl M, Holst S. Strain development of screw-retained implant-supported fixed restorations: procera implant bridge versus conventionally cast restorations. Int J Prosthodont 2012;25:166-9.
  12. Misch CE. Dental implant prosthetics. 2nd ed. St. Louis; CV Mosby; 2014. p. 747.
  13. Maeda Y, Satoh T, Sogo M. In vitro differences of stress concentrations for internal and external hex implant-abutment connections: a short communication. J Oral Rehabil 2006;33:75-8. https://doi.org/10.1111/j.1365-2842.2006.01545.x
  14. Heckmann SM, Karl M, Wichmann MG, Winter W, Graef F, Taylor TD. Cement fixation and screw retention: parameters of passive fit. An in vitro study of three-unit implant-supported fixed partial dentures. Clin Oral Implants Res 2004;15:466-73. https://doi.org/10.1111/j.1600-0501.2004.01027.x
  15. Karl M, Taylor TD. Bone adaptation induced by non-passively fitting implant superstructures: A randomized clinical trial. Int J Oral Maxillofac Implants 2016;31:369-75. https://doi.org/10.11607/jomi.4331
  16. Im SM, Kim DG, Park CJ, Cha MS, Choi LR. Biomechanical considerations for the screw of implant prosthesis: A literature review. J Korean Acad Prosthodont 2010;48:61-8. https://doi.org/10.4047/jkap.2010.48.1.61
  17. Huang Y, Wang J. Mechanism of and factors associated with the loosening of the implant abutment screw: A review. J Esthet Restor Dent 2019;31:338-345. https://doi.org/10.1111/jerd.12494
  18. Matsuzaki M, Ayukawa Y, Sakai N, Matsuzaki T, Matsushita Y, Koyano K. A comparison of the peri-implant bone stress generated by the preload with screw tightening between 'bonded' and 'contact' model. Comput Methods Biomech Biomed Engin 2017;20:393-402. https://doi.org/10.1080/10255842.2016.1236370
  19. Al-Otaibi HN, Akeel RF. The effects of two torque values on the screw preload of implant-supported prostheses with passive fit or misfit. Int J Oral Maxillofac Implants 2014;29:1058-63. https://doi.org/10.11607/jomi.3419
  20. Wentaschek S, Tomalla S, Schmidtmann I, Lehmann KM. Preload, coefficient of friction, and thread friction in an implant-abutment-screw complex. Int J Prosthodont 2017;30:542-4. https://doi.org/10.11607/ijp.5286
  21. Rutkunas V, Larsson C, Vult von Steyern P, Mangano F, Gedrimiene A. Clinical and laboratory passive fit assessment of implant-supported zirconia restorations fabricated using conventional and digital workflow. Clin Implant Dent Relat Res 2020;22:237-45. https://doi.org/10.1111/cid.12885
  22. Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol 2002;28:67-73. https://doi.org/10.1563/1548-1336(2002)028<0067:TRBIPA>2.3.CO;2
  23. Zipprich H, Rathe F, Pinz S, Schlotmann L, Lauer HC, Ratka C. Effects of screw configuration on the preload force of implant-abutment screws. Int J Oral Maxillofac Implants 2018;33:e25-32. https://doi.org/10.11607/jomi.5837
  24. Lee JH. Axial displacement of internal conical connection abutments under static load. MS Thesis, Kyungpook National University, 2019.