Effects of Fully Filling Deep Electron/Hole Traps in Optically Stimulated Luminescence Dosimeters in the Kilovoltage Energy Range

  • Chun, Minsoo (Department of Radiation Oncology, Chung-Ang University Gwang Myeong Hospital) ;
  • Jin, Hyeongmin (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Lee, Sung Young (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Kwon, Ohyun (Department of Medical Physics, University of Wisconsin) ;
  • Choi, Chang Heon (Institute of Radiation Medicine, Seoul National University Medical Research Center) ;
  • Park, Jong Min (Institute of Radiation Medicine, Seoul National University Medical Research Center) ;
  • Kim, Jung-in (Institute of Radiation Medicine, Seoul National University Medical Research Center)
  • 투고 : 2021.09.24
  • 심사 : 2022.03.06
  • 발행 : 2022.09.30

초록

Background: This study investigated the characteristics of optically stimulated luminescence dosimeters (OSLDs) with fully filled deep electron/hole traps in the kV energy ranges. Materials and Methods: The experimental group consisted of InLight nanoDots, whose deep electron/hole traps were fully filled with 5 kGy pre-irradiation (OSLDexp), whereas the non-pre-irradiated OSLDs were arranged as a control group (OSLDcont). Absorbed doses for 75, 80, 85, 90, 95, 100, and 105 kVp with 200 mA and 40 ms were measured and defined as the unit doses for each energy value. A bleaching device equipped with a 520-nm long-pass filter was used, and the strong beam mode was used to read out signal counts. The characteristics were investigated in terms of fading, dose sensitivities according to the accumulated doses, and dose linearity. Results and Discussion: In OSLDexp, the average normalized counts (sensitivities) were 12.7%, 14.0%, 15.0%, 10.2%, 18.0%, 17.9%, and 17.3% higher compared with those in OSLDcont for 75, 80, 90, 95, 100, and 105 kVp, respectively. The dose accumulation and bleaching time did not significantly alter the sensitivity, regardless of the filling of deep traps for all radiation qualities. Both OSLDexp and OSLDcont exhibited good linearity, by showing coefficients determination (R2) > 0.99. The OSL sensitivities can be increased by filling of deep electron/hole traps in the energy ranges between 75 and 105 kVp, and they exhibited no significant variations according to the bleaching time.

키워드

과제정보

This work was supported by the National Research Foundation (NRF) of Korea grant funded by the Korean government (NRF-2019R1F1A1041944).

참고문헌

  1. Mrcela I, Bokulic T, Izewska J, Budanec M, Frobe A, Kusic Z. Optically stimulated luminescence in vivo dosimetry for radiotherapy: physical characterization and clinical measurements in 60Co beams. Phys Med Biol. 2011;56(18):6065-6082. https://doi.org/10.1088/0031-9155/56/18/018
  2. Yukihara EG, McKeever SW. Optically stimulated luminescence (OSL) dosimetry in medicine. Phys Med Biol. 2008;53(20):R351-R379. https://doi.org/10.1088/0031-9155/53/20/R01
  3. Jursinic PA. Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements. Med Phys. 2007;34(12):4594-4604. https://doi.org/10.1118/1.2804555
  4. Jursinic PA. Angular dependence of dose sensitivity of nanoDot optically stimulated luminescent dosimeters in different radiation geometries. Med Phys. 2015;42(10):5633-5641. https://doi.org/10.1118/1.4929558
  5. Han SC, Kim KB, Choi SH, Park S, Jung H, Ji YH. Changes of optically stimulated luminescence dosimeter sensitivity with high dose. Prog Med Phys. 2016;27(2):98-104. https://doi.org/10.14316/pmp.2016.27.2.98
  6. Kerns JR, Kry SF, Sahoo N, Followill DS, Ibbott GS. Angular dependence of the nanoDot OSL dosimeter. Med Phys. 2011;38(7):3955-3962. https://doi.org/10.1118/1.3596533
  7. Leunens G, Van Dam J, Dutreix A, Van der Schueren E. Importance of in vivo dosimetry as part of a quality assurance program in tangential breast treatments. Int J Radiat Oncol Biol Phys. 1994;28(1):285-296. https://doi.org/10.1016/0360-3016(94)90169-4
  8. Al-Senan RM, Hatab MR. Characteristics of an OSLD in the diagnostic energy range. Med Phys. 2011;38(7):4396-4405. https://doi.org/10.1118/1.3602456
  9. Mobit P, Agyingi E, Sandison G. Comparison of the energy-response factor of LiF and Al2O3 in radiotherapy beams. Radiat Prot Dosimetry. 2006;119(1-4):497-499. https://doi.org/10.1093/rpd/nci676
  10. Okazaki T, Hayashi H, Takegami K, Okino H, Kimoto N, Maehata I, et al. Fundamental study of nanoDot OSL dosimeters for entrance skin dose measurement in diagnostic X-ray examinations. J Radiat Prot Res. 2016;41(3):229-236. https://doi.org/10.14407/jrpr.2016.41.3.229
  11. Aznar MC, Hemdal B, Medin J, Marckmann CJ, Andersen CE, Botter-Jensen L, et al. In vivo absorbed dose measurements in mammography using a new real-time luminescence technique. Br J Radiol. 2005;78(928):328-334. https://doi.org/10.1259/bjr/22554286
  12. Yukihara EG, Ruan C, Gasparian PB, Clouse WJ, Kalavagunta C, Ahmad S. An optically stimulated luminescence system to measure dose profiles in X-ray computed tomography. Phys Med Biol. 2009;54(20):6337-6352. https://doi.org/10.1088/0031-9155/54/20/020
  13. Ruan C, Yukihara EG, Clouse WJ, Gasparian PB, Ahmad S. Determination of multislice computed tomography dose index (CTDI) using optically stimulated luminescence technology. Med Phys. 2010;37(7):3560-3568. https://doi.org/10.1118/1.3455285
  14. Jaffray DA, Siewerdsen JH, Wong JW, Martinez AA. Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys. 2002;53(5):1337-1349. https://doi.org/10.1016/S0360-3016(02)02884-5
  15. Vijayan S, Xiong Z, Rudin S, Bednarek DR. A system to track skin dose for neuro-interventional cone-beam computed tomography (CBCT). Proc SPIE Int Soc Opt Eng. 2016;9783:97832X.
  16. Scarboro SB, Cody D, Alvarez P, Followill D, Court L, Stingo FC, et al. Characterization of the nanoDot OSLD dosimeter in CT. Med Phys. 2015;42(4):1797-1807. https://doi.org/10.1118/1.4914398
  17. Chruscinska A, Palczewski P. OSL characteristics: theory and experiments. Radiat Prot Dosimetry. 2020;192(2):266-293. https://doi.org/10.1093/rpd/ncaa205
  18. Jursinic PA. Changes in optically stimulated luminescent dosimeter (OSLD) dosimetric characteristics with accumulated dose. Med Phys. 2010;37(1):132-140. https://doi.org/10.1118/1.3267489
  19. Park SY, Choi CH, Park JM, Chun M, Han JH, Kim JI. Sensitivity and stability of optically stimulated luminescence dosimeters with filled deep electron/hole traps under pre-irradiation and bleaching conditions. Phys Med. 2017;38:81-87. https://doi.org/10.1016/j.ejmp.2017.05.057
  20. Yukihara EG, Whitley VH, McKeever SW, Akselrod AE, Akselrod MS. Effect of high-dose irradiation on the optically stimulated luminescence of Al2O3:C. Radiat Meas. 2004;38(3):317-330. https://doi.org/10.1016/j.radmeas.2004.01.033
  21. Schembri V, Heijmen BJ. Optically stimulated luminescence (OSL) of carbon-doped aluminum oxide (Al2O3:C) for film dosimetry in radiotherapy. Med Phys. 2007;34(6):2113-2118. https://doi.org/10.1118/1.2737160
  22. Yukihara EG, Whitley VH, Polf JC, Klein DM, McKeever SW, Akselrod AE, et al. The effects of deep trap population on the thermoluminescence of Al2O3:C. Radiat Meas. 2003;37(6):627-638. https://doi.org/10.1016/S1350-4487(03)00077-5
  23. Omotayo AA, Cygler JE, Sawakuchi GO. The effect of different bleaching wavelengths on the sensitivity of Al2O3:C optically stimulated luminescence detectors (OSLDs) exposed to 6 MV photon beams. Med Phys. 2012;39(9):5457-5468. https://doi.org/10.1118/1.4742865
  24. Ma CM, Coffey CW, DeWerd LA, Liu C, Nath R, Seltzer SM, et al. AAPM protocol for 40-300 kV X-ray beam dosimetry in radiotherapy and radiobiology. Med Phys. 2001;28(6):868-893. https://doi.org/10.1118/1.1374247
  25. Bhat M, Pattison J, Bibbo G, Caon M. Diagnostic X-ray spectra: a comparison of spectra generated by different computational methods with a measured spectrum. Med Phys. 1998;25(1):114-120. https://doi.org/10.1118/1.598170
  26. Seibert JA. The AAPM/RSNA physics tutorial for residents: X-ray generators. Radiographics. 1997;17(6):1533-1557. https://doi.org/10.1148/radiographics.17.6.9397462
  27. Njiki CD, Ndah TN, Ben Bolie GH, Beyala Ateba JF, Simo A, Yigbedeck YH. Comparison of the performance of two OSL readers for the personal dose equivalent Hp(10) measurement. Radiat Prot Dosimetry. 2021;193(1):37-42. https://doi.org/10.1093/rpd/ncab022
  28. Klevenhagen SC. Experimentally determined backscatter factors for X-rays generated at voltages between 16 and 140 kV. Phys Med Biol. 1989;34(12):1871. https://doi.org/10.1088/0031-9155/34/12/010
  29. Reft CS. The energy dependence and dose response of a commercial optically stimulated luminescent detector for kilovoltage photon, megavoltage photon, and electron, proton, and carbon beams. Med Phys. 2009;36(5):1690-1699. https://doi.org/10.1118/1.3097283