DOI QR코드

DOI QR Code

Monte Carlo Simulation-Based Mammographic Anti-Scatter Grids to Evaluate Performance of Digital Mammography Detector

디지털 맘모 디텍터 성능평가를 위한 몬테카를로용 산란선 제거 그리드 작성에 관한 연구

  • Yeji Kim (Department of Multidisciplinary Radiological Science, The Graduate School of Dongseo University) ;
  • Hyejin Jo (Department of Multidisciplinary Radiological Science, The Graduate School of Dongseo University) ;
  • Yongsu Yoon (Center for Radiological Environment & Health Science, Dongseo University)
  • 김예지 (동서대학교 일반대학원 융합방사선학과) ;
  • 조혜진 (동서대학교 일반대학원 융합방사선학과) ;
  • 윤용수 (동서대학교 방사선보건환경연구센터)
  • Received : 2023.12.26
  • Accepted : 2024.01.08
  • Published : 2024.02.28

Abstract

In Recent years, there has been a noticeable increase in the global incidence of breast cancer, with approximately 2.3 million cases of female breast cancer reported worldwide in 2020. Numerous studies are currently underway to enhance the accuracy of breast cancer diagnosis through the development of digital mammography detectors. This study aims to create Monte Carlo simulation-based mammographic anti-scatter grids and investigate their utility in evaluating the performance of digital mammography detector. Two types of mammographic anti-scatter grids, MAM-CP and Senographe 600T HF, were created using Monte Carlo simulation software (MCNPX 2.7.0), with grid ratios of 3.7 : 1 and 5 : 1, respectively. The grid physical characteristics (sensitivity, exposure factor, contrast improvement ratio) were calculated based on the KS C IEC60627 in the simulations using two X-ray qualities, RQA-M2 (28 kVp) and MW4 (35 kVp). As the X-ray tube voltage increased from 28 kVp to 35 kVp, sensitivity and exposure factor exhibited a decreasing trend, while contrast improvement ratio demonstrated an increasing trend. With an increase in grid ratio from 3.7 : 1 to 5 : 1, all physical characteristics showed an upward trend. Our results were consistent with a previous study that conducted measurements of physical properties using a real phantom. However, the pattern of change in the contrast improvement ratio with X-ray tube voltage differed from the previous study.

Keywords

Acknowledgement

This work was supported by the National Research Foundation of Korea(NRF) grant funded by the Korea government(MSIT) (No.2022R1G1A1009477)

References

  1. Sung H, Ferlay J, Siegel RL, Laversanne M, SoerjomataramI, Jemal A, et al. Global cancer statistics 2020: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2021;71:209-49. DOI: https://doi.org/10.3322/caac.21660
  2. Hong S, Won YJ, Park YR, Jung KW, Kong HJ, Lee ES, et al. Cancer statistics in Korea: Incidence, mortality, survival, and prevalence in 2017. Cancer Res Treat. 2020;52(2):335-50. DOI: https://doi.org/10.4143/crt.2020.206
  3. Kang MJ, Won YJ, Lee JJ, Jung KW, Kim HJ, Kong HJ, et al. Cancer statistics in Korea: Incidence, mortality, survival, and prevalence in 2019. Cancer Res Treat. 2022;54(2):330-44. DOI: https://doi.org/10.4143/crt.2022.128
  4. Berry D, Cronin K, Plevritis S, Fryback D, Clarke L, Zelen M, et al. Effect of screening and adjuvant therapy on mortality from breast cancer. N Engl J Med. 2005;353:1784-92. DOI: http://doi.org/10.1056/NEJMoa050518
  5. Song SY, Hong S, Jun JK. Digital mammography as a screening tool in Korea. J Korean Soc Radiol. 2021;82(1):2-11. DOI: https://doi.org/10.3348/jksr.2021.0004
  6. Doi K. Diagnostic imaging over the last 50 years: Research and development in medical imaging science and technology. Phys Med Biol. 2006;51(13):R5-27. DOI: http://doi.org/10.1088/0031-9155/51/13/R02
  7. Yoon Y, Morishita J, Park M, Kim H, Kim K, Kim J. Monte Carlo simulation-based feasibility study of novel indirect flat panel detector system for removing scatter radiation. Physica Medica. 2016;32(1):182-7. DOI: https://doi.org/10.1016/j.ejmp.2015.11.008
  8. Roh Y, Yoon Y, Kim K, Kim J, Kim J, Morishita J. A novel radiation detector for removing scattered radiation in chest radiography: Monte Carlo simulation-based performance evaluation. Journal of Instrumentation. 2016;11(10):T10008. DOI: http://doi.org/10.1088/1748-0221/11/10/T10008
  9. Yoon Y. An optimal structure of a novel flat panel detector to reduce scatter radiation for clinical usage: Performance evaluation with various angle of incident X-ray. Journal of Radiological Science and Technology. 2017;40(4):533-42. DOI: https://doi.org/10.17946/JRST.2017.40.4.01
  10. Jeong H, Min J, Yoon Y, Kim J. Investigation of physical imaging properties in various digital radiography systems. Journal of Radiological Science and Technology. 2017;40(3):363-70. DOI: https://doi. org/10.17946/JRST.2017.40.3.02
  11. Saunders Jr. R, Samei E, Jesneck J, Lo J. Physical characterization of a prototype selenium-based full field digital mammography detector. Med Phys. 2005;32(2):588-99. DOI: https://doi.org/10.1118/1.1855033
  12. Rezentes PS, De Almeida A, Barnes GT. Mammography grid performance. Radiology. 1999;210(1):227-32. DOI: https://doi.org/10.1148/radiology.210.1.r99dc35227
  13. IEC Standard 61267: 2.0. Medical diagnostic x-ray equipment-radiation conditions for use in the determination of characteristics. 2005. Retrieved from https://webstore.iec.ch/publication/5079
  14. KS C IEC60627: 3.0. Diagnostic X-ray imaging equipment-Characteristics of general purpose and mammographic anti-scatter grids. 2019. Retrieved from https://standard.go.kr/KSCI/standardIntro/getStandardSearchView.do?menuId=919&topMenuId=502&upperMenuId=503&ksNo=KSCIEC60627&tmprKsNo=KSCIEC60627&reformNo=03
  15. Mizuta M, Sanada S, Akazawa H, Kasai T, Abe S, Ikeno Y, et al. Comparison of anti-scatter grids for digital imaging with use of a direct-conversion flat-panel detector. Radiological Physics and Technology. 2011;5(1):46-52. DOI: https://doi.org/10.1007/s12194-011-0134-1