Processing and Accuracy Assessment of MICROTOPS II Atmospheric Measurements for Absolute Radiometric Calibration of Optical Satellite Imagery

  • 투고 : 2026.02.24
  • 심사 : 2026.03.30
  • 발행 : 2026.04.30

초록

Field-based absolute radiometric calibration of optical satellites requires atmospheric profiles measured at the time of image acquisition asinputsto radiative transfer models under clear-sky conditions. However, commonly used fixed-site observations (e.g., RadCalNet) or satellite products may introduce spatial and temporal discrepancies. Therefore, continuous atmospheric measurements during satellite overpasses are necessary for accurate calibration. In thisstudy, aerosol optical depth (AOD) and totalwater vapor (TWV) were measured using a MICROTOPS II sunphotometer, and total ozone column (TOC) was measured using a MICROTOPS II ozonometer. A total of nine field campaigns were conducted in 2024 at four sites in Korea, coinciding with satellite image acquisitions. Quality control procedures were applied to ensure data reliability. AOD at 550 nm was estimated using two spectral channels near 550 nm. After applying quality control (QC) procedures, including cloud screening and outlier removal, the processed AOD showed strong agreement with SKYNET and AERONET observations (correlation coefficient[r] > 0.82 and root mean square error[RMSE] = 0.009-0.015),with slightly higher valuesfrom MICROTOPS II. TWV datawere validated based onAOD QCresults and showed a high correlationwith radiosonde data (r = 0.974). TOC measurements, obtained under controlled solar zenith angles(26-63°), showed strong agreement with Aura Ozone Monitoring Instrument (OMI) products(r = 0.952). In terms of uncertainty, the quality-controlled MICROTOPS II measurements showed uncertainties of less than 0.01 for bothAOD and TWV, and approximately 0.002 for TOC,which are comparable to the uncertainty levels of AERONET observations. Quality-controlled atmospheric parameters can be used as inputs to radiative transfer modelsto improve the stability of absolute radiometric calibration. Furthermore, the use of campaign-specific measurements enables more accurate uncertainty estimation compared to assuming a uniform atmospheric uncertainty.

키워드

1. Introduction

When modeling the process in which solar radiation travels through the atmosphere, reflects off the Earth’s surface, and is ultimately detected by an optical satellite sensor, aerosol optical depth (AOD), total water vapor (TWV), and total ozone column (TOC) are key atmospheric input variables required to estimate the attenuation and modification of radiative energy within the atmosphere (Berk et al., 2014; Gao et al., 2009). AOD represents the degree to which aerosol particles in the atmosphere scatter and absorb direct or reflected solar radiation, and is used to calculate direct radiance attenuation and scattering components (Lin et al., 2023; Qiao et al., 2023). It is one of the most sensitive variables when converting satellite top-of-atmosphere (TOA) radiance and surface reflectance in the visible and near-infrared spectral regions (Vermote et al., 2002). Water vapor is the major gaseous absorber in the near-infrared and shortwave infrared regions, and TWV represents the total amount of water vapor integrated over a vertical atmospheric column; it is an essential input variable for estimating absorption coefficients, atmospheric transmittance, and path radiance (Gao et al., 2009). TOC refers to the total vertically integrated ozone amount in the atmosphere and induces strong absorption in the 0.5–0.7 μm spectral range (Brion et al., 1998; Gorshelev et al., 2014). Accordingly, TOC is used to calculate radiative transmittance in the visible range and serves as input to determine absorption coefficients in radiative transfer models such as MODerate Resolution Atmospheric TRANsmission (MODTRAN) and 6S (Berk et al., 2014).

Among these atmospheric variables, AOD is widely recognized as the most influential parameter in radiative transfer simulations and atmospheric correction, as it directly governs atmospheric scattering and path radiance. Previous studies have reported that an uncertainty of ±0.1 in AOD can lead to approximately 5–10% variation in TOA reflectance, particularly in the visible spectral region (Vermote et al., 2002; Gao et al., 2009). In contrast, uncertainties in TWV and TOC generally result in smaller reflectance variations on the order of 1–3% and less than 2%, respectively, depending on wavelength (Ichoku et al., 2002; Giles et al., 2019). These differences indicate that AOD is the dominant source of uncertainty in atmospheric correction and absolute radiometric calibration.

Absolute radiometric calibration converts the digital number (DN) recorded by an optical satellite sensor into actual radiance or reflectance. This process requires modeling the scattering and absorption of solar radiation along its path at the time of image acquisition, which in turn requires accurate measurements of AOD, TWV, and TOC corresponding to the observation time (Tang et al., 2023; Jin et al., 2020). Maxar, for example, conducts absolute radiometric calibration using sunphotometers linked to the AERONET network to measure AOD, TWV, and TOC (Maxar Technologies, 2020). For the radiometric calibration of Korea Multi-Purpose Satellite-3A (KOMPSAT-3A), atmospheric parameters such as pressure, temperature, and relative humidity were collected using an Automatic Weather Station (AWS) in Goheung, Korea, while AOD, TWV, and TOC were measured using an multifilter rotating shadowband radiometer (MFRSR) and UV-MFR (Jin et al., 2020). Although collected at different times, AWS and MICROTOPS II instruments were used in the same region to measure AOD, TWV, and TOC for the absolute calibration of the same satellite (Yeom et al., 2018).

At the Baotou calibration site, one of the RadCalNet locations, AOD and TWV measurements obtained from a sun-tracking photometer installed at a nearby AERONET site have been utilized for the absolute radiometric calibration of satellites such as SuperView-1 and ZY3-02 (Liu et al., 2020; Tang et al., 2022). For the Sentinel-2 mission, AOD and TWV measured using CIMEL instruments at RadCalNet sites such as La Crau in France have similarly been employed (Revel et al., 2019). In RadCalNet, atmospheric parameters are measured in situ at each calibration site using dedicated instruments, rather than relying on nearby AERONET stations. This approach ensures strict spatial and temporal collocation with satellite observations, which is essential for accurate radiative transfer simulations and calibration.

Because atmospheric conditions such as aerosol loading and water vapor can vary significantly over short distances and time intervals, the use of co-located measurements enables more reliable estimation of TOA radiance and reduces uncertainties associated with spatial and temporal mismatches (Bouvet et al., 2019). In operational calibration frameworkssuch as RadCalNet, quality control is primarily driven by AOD, which is retrieved from AERONET instruments with well-established cloud-screening and calibration procedures (Giles et al., 2019). Other atmospheric parameters, such as TWV and TOC, are typically obtained either from the same measurement system or from external datasets (e.g., satellite products or reanalysis data), without applying independent quality control (QC) procedures of comparable strictness. This reflects the relatively lower sensitivity of TWV and TOC in radiative transfer simulations compared to AOD.

Atmospheric data used for absolute radiometric calibration can be broadly categorized into fixed-site observations, such as AERONET, and portable field observations. The CIMEL sunphotometer, the official AERONET standard instrument, performs automatic solar tracking and produces datasets processed by the AERONET Level 2 algorithm, which includes Langley calibration and automated quality control (Giles et al.,2019). CIMEL instruments provide high-accuracy measurements with uncertainties of less than ±0.02 for AOD, ±0.1 g cm–² for water vapor, and ±3 DU for TOC (Holben et al., 1998; Giles et al., 2019). SKYNET also operates approximately 450 global and 60 regional sites (Choi et al., 2011), using skyradiometers to measure solar irradiance for applications such as radiative transfer modeling and validation of satellite products (Ichoku et al., 2002). SKYNET measurements exhibit an uncertainty of approximately ±0.02 and undergo cloud screening, Langley calibration, high-altitude outdoor calibration, and calibration sphere procedures.

However, instruments such as AERONET and SKYNET cannot be deployed at all required times and locations. For outdoor field campaigns supporting domestic absolute radiometric calibration, itistherefore necessary to conduct measurements at specific times and locations that coincide with satellite overpass conditions. This requirement for temporally and spatially coincident atmospheric observations further highlights the need for portable measurement systems that can be deployed directly at field campaign sites. In such cases, MICROTOPS II (Morys et al., 2001)—a low-cost, portable, and easy-to-use handheld instrument—can serve as an effective alternative. MICROTOPS II is a handheld multiwavelength sunphotometer capable of rapidly measuring key atmospheric variables such as AOD, TWV, and TOC, making it widely used in field campaigns for radiometric calibration of remote sensing imagery. The instrument is manufactured by Solar Light Company in Philadelphia, USA, and is available in two versions: a filter-based sunphotometer and an ozonometer exclusively for ozone measurements (Solar Light Company, 2023). In terms of quantitative accuracy, MICROTOPS II exhibits lower performance compared to fixed-site instruments, with uncertainties of ±0.03 or less for AOD, ±0.2–0.3 g cm–² for TWV, and ±5–10 DU for TOC (Ichoku et al., 2002). Due to its portability, measurement accuracy may be affected by observer skill, environmental conditions, and calibration status, leading to noise and bias in the raw data. Thus, systematic quality control and processing procedures are required for quantitative use in radiometric calibration.

Existing studies involving MICROTOPS II have mostly focused on observational considerations rather than detailed data processing or quality assurance (Lee & Lee, 2016; Gómez-Amo et al., 2012). Data processing studies have predominantly addressed AOD, with quality verification typically performed through comparisons with AERONET (Gong, 2018; Purnama et al., 2023; Ichoku et al., 2002). Although early studies emphasized the need for Langley calibration (Ichoku et al.,2002), current MICROTOPSII instruments are shipped with pre-stored extraterrestrial constants for each spectral band, calibrated against National Institute of Standards and Technology (NIST)-traceable standard spectroradiometers (Solar Light Company, 2016). AOD processing techniques have been proposed primarily for ship-based sensors, addressing cloud screening, air mass restrictions (≤3), and other procedures to improve accuracy, resulting in root mean square error (RMSE) values below 0.1 relative to AERONET (Gong, 2018; Purnama et al., 2023). In regions without fixed observation sites or in field campaigns for absolute radiometric calibration, atmospheric measurements using portable instruments like MICROTOPS II are essential, necessitating reliable measurement practices along with quality control procedures similar to those used for AERONET.

Therefore, this study aims to develop processing and quality-control methods for MICROTOPS II measurements of AOD, TWV, and TOC obtained during field campaigns conducted in 2024 to acquire accurate atmospheric information for absolute radiometric calibration. Considering the dominant influence of AOD on radiative transfer and calibration uncertainty, the proposed quality control procedures are primarily focused on AOD, while TWV and TOC are processed using consistency-based or physically constrained approaches. This strategy is consistent with operational calibration frameworks and is intended to ensure stable and reliable atmospheric inputs for absolute radiometric calibration. In addition, the uncertainty of the QC-processed atmospheric parameters is quantitatively evaluated using standard and relative uncertainty metrics, following internationally recognized guidelines, to assess whether the derived datasets provide consistent uncertainty levels comparable to reference networks such as AERONET. The accuracy of the proposed methods is evaluated using AERONET, SKYNET, radiosonde, and Ozone Monitoring Instrument (OMI) datasets as reference.

2. Materials and Methods

2.1. Field Campaigns and Atmospheric Measurements

A total of nine absolute radiometric calibration field campaigns were conducted throughout 2024 in Daejeon, Jeonju, Cheonan, and Incheon to support the absolute radiometric calibration of very-high-resolution (VHR) satellite imagery (Fig. 1, Table 1). During each campaign, four tarpaulin targets with reflectances of 5%, 20%, 30%, and 50% were deployed. Atmospheric measurements and tarp reflectance observations were performed between approximately 10:00 AM and 12:00 PM to coincide with satellite acquisition times (Fig. 2). The reflectance of the tarpaulins was measured using a FieldSpec spectroradiometer (Analytical Spectral Devices, ASD). Meteorological parameters such as temperature, humidity, wind speed, and pressure were collected using an AWS, while AOD, TWV, and TOC were measured using a MICROTOPS II sunphotometer and ozonometer instruments (Table 2).

OGCSBN_2026_v42n2_181_4_f0001.png 이미지

Fig.1. Spatial distribution of field campaign sites and ground observation stations used for absolute radiometric calibration.

Table 1. List of field campaigns and materials used in this study

OGCSBN_2026_v42n2_181_4_t0001.png 이미지

OGCSBN_2026_v42n2_181_4_f0002.png 이미지

Fig. 2. Field campaign photographs used for absolute radiometric calibration. (a) Tarp layout and ASD measurements. (b) Installation of the MICROTOPS II instrument. (c) Installation of the AWS instrument.

Table 2. Types and specifications of measuring instruments used in field campaigns

OGCSBN_2026_v42n2_181_5_t0001.png 이미지

The MICROTOPS II instruments were installed in an unobstructed, sunlit location to ensure a direct line of sight to the Sun and were connected to external GPS receivers. Atmospheric measurements were conducted only under clear-sky conditions to minimize cloud effects; observations were avoided when clouds overlapped or approached the solar disk. The sun-targeting window of MICROTOPS II is equipped with crosshairs and two concentric circles that share the same center. Since AOD errors increase exponentially when the Sun deviates from the center (Ichoku et al., 2002), measurements were made only when the solar disk was aligned with the intersection of the crosshairs.

The MICROTOPS II sunphotometer measures AOD and TWV using five channels selected from 440, 500, 675, 870, 936, and 1,020 nm. The MICROTOPS II ozonometer measures TOC with five channels at 305.5, 312.5, 320, 936, and 1,020 nm, and TOC is derived using the UV wavelengths (305.5, 312.5, 320 nm). Atmospheric measurements were performed within ±1 hour of the satellite overpass time, and to minimize observational uncertainties, 3–5 measurements were taken per minute, with each measurement completed within 10 seconds.

2.2. Reference Data for Accuracy Assessment

The reference datasets used to evaluate the accuracy of the MICROTOPS II measurements are summarized in Table 3. For AOD validation, AERONET and SKYNET data within a 1-km radius of the field-campaign sites were used. AERONET is a global ground-based aerosol remote sensing network that provides long-term, continuous, and publicly accessible atmospheric datasets (Holben et al., 1998). The CIMEL sunphotometer commonly used in AERONET acquires direct solar irradiance at 90 wavelengths near 340, 380, 440, 500, 675, 870, 936, 1,020, and 1,640 nm with a field of view (FOV) of 1.2°, at 15-minute intervals (Giles et al., 2020). AERONET provides physical aerosol parameters such as columnar AOD, single-scattering albedo, and particle size distribution through calibration and quality control processes, and releases Level 1.0 (raw), Level 1.5 (cloud-screened), and Level 2.0 (cloud-screened and quality-assured) datasets, with an accuracy of approximately ±0.02 (Eck et al., 1999). In this study, Level 1.5AOD data at 440, 500, and 675 nm—comparable to the MICROTOPS II processing level—were used. SKYNET is a global ground-based aerosol and radiation observation network centered in Japan, Asia, and Europe. The Skyradiometer POM-02, manufactured by Prede in Japan, performs solar tracking and measures direct and diffuse irradiance at 11 wavelengths (315, 340, 380, 400, 500, 675, 870, 940, 1,020, 1,627, and 2,200 nm) at 6-minute intervals (Bu et al., 2024). In this study, cloud-screened Level 2.0 AOD data measured at 400, 500, and 675 nm were used. The Korea Institute of Energy Research (KIER), located within 1 km of the Daejeon field campaign site, satisfies the 25 km spatial representativeness requirement for AOD (Park et al., 2023).

Table 3. Types and specifications of reference atmospheric vertical profile data

OGCSBN_2026_v42n2_181_5_t0002.png 이미지

Previous studies have reported differences of less than 0.02 between SKYNET and AERONET AOD measurements, indicating comparable accuracy. For TWV validation, radiosonde data obtained from the Korea Meteorological Administration (KMA) Weather Data Service were used. Radiosondes measure atmospheric variables (temperature, wind, pressure, altitude, and dew-point temperature) up to 30 km altitude at regular time intervals. The KMA conducts observations twice daily at 09 Korea Standard Time (KST) and 21 KST; in this study, 09 KST radiosonde measurements from Osan (No. 47122, 37.1°N, 127.0333°E, 52 m elevation) were used, resulting in nine datasets collected in 2024. Although radiosonde data have traditionally been used for climatological studies of upper-atmosphere humidity and temperature (Elliott & Gaffen, 1991; Luers & Eskridge, 1995; Zhai & Eskridge, 1996), measurement accuracy and representativeness may be affected, particularly in the lower troposphere, due to instrumental issues or environmental factors (Parlange & Brutsaert, 1989; Free et al., 2005). Errors in TWV retrieved from radiosondes can reach up to 6–8% due to latent heat release after icing, time lags between dry- and wet-bulb sensors, or incomplete ascent to the upper atmosphere (Bruegge et al., 1992). For TOC validation, TOC products from the Aura OMI were used. OMI measures global solar backscatter spectra in the ultraviolet and visible regions and has provided near-daily global TOC measurements at a spatial resolution of 13 × 24 km since 2004. In this study, OMI-TOMS (Total Ozone Mapping Spectrometer) Version 004 TOC Level 2 data were obtained from NASA’s GES DISC (Goddard Earth Sciences Data and Information Services Center).

2.3. Processing and Quality Control Methods

To simulate the radiative transfer model MODTRAN for absolute radiometric calibration, AOD at 550 nm at the time of satellite observation is required (Berk et al., 2008). First, AOD for each wavelength measured by the MICROTOPS II sunphotometer is computed using the Beer–Lambert–Bouguer law (Eq. 1)(Holben et al., 1998).

\(\begin{align}A O D(\lambda)=\frac{1}{m} \ln \left(\frac{I_{0}(\lambda)}{I(\lambda)}\right)-\tau_{R}(\lambda)-\tau_{g a s}(\lambda)\end{align}\)       (1)

I(λ): solar irradiance measured at the ground-based instrument,

I0(λ): extraterrestrial solar irradiance, m: air mass,

τR(λ): Rayleigh scattering optical thickness,

τgas(λ): gaseous absorption optical thickness (e.g., O3, NO2, H2O).

Since the MICROTOPS II instrument does not include a 550 nm channel, AOD at 550 nm was estimated using the two nearest wavelength channels (λ1, λ2). First, the Ångström exponent (α) was derived from the AOD values at these two wavelengths (τ(λ1), τ(λ2)) using Eq. (2). Then, the Ångström law (Eq. 3) was applied to estimate AOD550nm(τ(550))(Ångström, 1929; Yeom et al., 2018).

\(\begin{align}\alpha=-\frac{\ln \left(\tau\left(\lambda_{1}\right) / \tau\left(\lambda_{2}\right)\right)}{\ln \left(\lambda_{1} / \lambda_{2}\right)}\end{align}\)       (2)

\(\begin{align}\tau(550)=\tau\left(\lambda_{1}\right)\left(\frac{\lambda_{1}}{550}\right)^{\alpha} \;or \; \tau(550)=\tau\left(\lambda_{2}\right)\left(\frac{\lambda_{2}}{550}\right)^{\alpha}\end{align}\)       (3)

After removing negative AOD values at 550 nm—which were regarded as noise—from the MICROTOPS II dataset, cloud-affected AOD values were filtered out using a triplet stability test. In this procedure, consecutive sets of three measurements with deviations greater than 0.02 were discarded, following the algorithm also employed by AERONET (Smirnov et al., 2000). Considering that MICROTOPS II is a handheld instrument whose accuracy can depend on the observer’s proficiency, a single AOD value at the satellite overpass time was not used. Instead, measurements collected within ±10 minutes of the satellite imaging time were selected, and temporally averaged values were used when their standard deviation was within 0.025 to reduce measurement variability. The standard deviation threshold of 0.025 was determined based on field-measured AOD values identified as cloud-free using the 2024 GEMS Effective Cloud Fraction (ECF) product, and is consistent with the typical short-term variability of AERONET AOD measurements reported in previous studies (Giles et al., 2019). If all measurements within ±10 minutes were removed during the stability QC process, the mean AOD within ±30 minutes of the imaging time was used instead. Finally, to assess whether the resulting AOD values fall within the climatologically valid range for the Korean Peninsula, measurements were checked to ensure that the mean AOD remained below 0.7, based on AOD statistics from 2003 to 2019 (Noh, 2021). AOD values that satisfy all these conditions are regarded as high-quality AOD550nm and were used for absolute radiometric calibration.

The TWV (precipitable water, u) is derived using two MICROTOPS II wavelength channels: the water vapor absorption band at 936 nm and the non-absorbing reference band at 1,020 nm. First, for the 936 nm channel, the Bouguer–Lambert–Beer (BLB) law is modified by adding a water vapor absorption term, as shown in Eq. (4). In this equation, k is the absorption coefficient, representing the strength of water vapor absorption at the 936 nm band, and b is the absorption exponent, indicating the rate at which absorption reaches saturation. Both coefficients are empirical parameters derived from radiative transfer models and Langley calibration and are predefined (Schmid et al., 2001; Smirnov et al., 2000).

\(\begin{align}\frac {I(936)} {I_0(936)} = exp(-\tau (936) m-k(um)^b)\end{align}\)       (4)

OGCSBN_2026_v42n2_181_7_f0001.png 이미지

Fig. 3. Flowchart of the quality control process for AOD measurement data from MICROTOPS II.

For the non-absorbing 1,020 nm band, the relationship can be expressed as Eq. (5). The relationship between aerosol scattering at the 936 nm and 1,020 nm bands can then be formulated as Eq. (6). By substituting Eq. (6) into Eq. (4), the total column water vapor u can be derived, asshown in Eq. (7)(Solar Light Company, 2001).

\(\begin{align}\frac {I(1020)}{I_0(1020)}=exp(-\tau (1020)m)\end{align}\)       (5)

τ(936) = 1.16τ(1020)       (6)

\(\begin{align}u=\left[\frac{\tau(1020) m(1-1.16)-\ln \left(\frac{I(936)}{I_{0}(936)}\right)}{I_{0}(1020)}\right]^{1 / b}\end{align}\)       (7)

Since TWV is measured with the same instrument as AOD, no separate quality-control procedure was applied. Instead, TWV measurements acquired at the same timestamps as the final validated AOD data were defined and used as the valid TWV dataset.

In the ultraviolet spectral region, Rayleigh scattering and ozone absorption dominate, and thus the BLB law can be expressed as shown in Eq. (8).

I(λ) = I0(λ)exp[– m{τR(λ) + σO3(λ)uO3}] (8)

σO3(λ): ozone absorption cross-section at wavelength λ (cm2 molecule–1),

uO3(λ): total column ozone amount (cm · atm or DU [Dobson unit]).

Multiple wavelength pair combinations were generated using the ultraviolet channels of the MICROTOPS II (305.5 nm, 312.5 nm, and 320.0 nm). These combinations were inserted into Eq. (8), and the differenced results were rearranged in terms of the ozone amount (uO3), yielding Eqs. (9) and (10) (Komhyr et al., 1989). MICROTOPS II provides three ozone estimates: u(12)O3, derived from the 305.5–312.5 nm pair; u(23)O3, derived from the 312.5–320.0 nm pair; and a weighted average of the two, u(123)O3, as defined in Eq. (11). The weighted combination u(123)O3 is generally used as the primary TOC estimate because it simultaneously incorporates the high sensitivity of the shorter wavelengths and the stability of the longer wavelengths, thereby minimizing the effects of temperature dependence, Rayleigh correction errors, and extraterrestrial constant uncertainties (Solar Light Company Inc., 2001). In this study, the mean value of u(123)O3 was used as the TOC input for absolute radiometric calibration. Among the factors determining the quality of ozone retrievals, air mass (m) is critical because it represents the relative optical path length of solar radiation through the atmosphere and is defined by the solar zenith angle. As the solar zenith angle increases—that is, when the Sun is lower in the sky—the atmospheric path length becomes longer, leading to increased absorption and scattering, reduced detector signal, and consequently a greater influence on TOC estimation. Therefore, the MICROTOPS II manual recommends using ozone measurements only when the air mass is below 3, corresponding to solar zenith angles less than 70° (Solar Light Company, 2001). Since field campaigns for absolute radiometric calibration were conducted between 10:00 AM and noon, all measurements were collected within solar zenith angles of 26–63°, corresponding to air mass values below 3 (Gómez-Amo et al., 2012). Thus, all TOC measurements were considered valid, and no additional QC procedures were applied.

\(\begin{align}u_{O 3}^{(12)}=\frac{\ln \left(I(312.5) / I(305.5)-m\left(\tau_{R}(305.5)-\tau_{R}(312.5)\right)\right.}{m\left[\sigma_{O 3}(305.5)-\sigma_{O 3}(312.5)\right]}\end{align}\)       (9)

\(\begin{align}u_{O 3}^{(23)}=\frac{\ln \left(I(320.0) / I(312.5))-m\left(\tau_{R}(312.5)-\tau_{R}(320.0)\right)\right.}{m\left[\sigma_{O 3}(312.5)-\sigma_{O 3}(320.0)\right]}\end{align}\)       (10)

\(\begin{align}u_{O 3}^{(123)}=\frac{\ln \left(I(320.0) / I(305.5))-m\left(\tau_{R}(305.5)-\tau_{R}(320.0)\right)\right.}{m\left[\sigma_{O 3}(305.5)-\alpha_{123} \sigma_{O 3}(320.0)\right]}\end{align}\)       (11)

α123: absorption ratio for the three wavelengths, typically ranging from 1.20 to 1.25, with instrument-specific values preset internally (Solar Light Company, 2001).

2.4.Evaluation Methods for MICROTOPS II Data Processing

To evaluate the accuracy of the atmospheric parameter processing methods developed for the MICROTOPS II instrument used in absolute radiometric calibration, correlation coefficients and RMSE values were calculated using fixed-site observations and satellite datasets as references. Valid AOD values measured and processed at each field campaign site were compared with AERONET and SKYNET observations located within 1 km of the field measurement points, and correlation andRMSE statistics were computed. For spatial and temporal collocation, AOD and TWV observations from fixed-site stations located within 1 km of the field measurement sites were used, and only data acquired within ±1 minute of the field measurements were selected to ensure temporal consistency. Since neither AERONET nor SKYNET provides AOD at 550 nm, AOD at 400 nm and 675 nm from each network were used to generate AOD550nm through the same processing procedure applied to MICROTOPS II data (Eqs. 2 and 3), and the resulting AOD550nm values were compared.

For TWV, radiosonde data were used asreference. Water vapor pressure (e) was first computed from dew-point temperature (Td) using Eq. (12). Specific humidity (q) was then derived using water vapor pressure and observed atmospheric pressure (p) according to Eq. (13). TWV was subsequently calculated using Eq. (14), which incorporates specific humidity, pressure differences between atmospheric layers (Δp), and gravitational acceleration (g=9.806 ms–2) (Wallace & Hobbs, 2006; Campmany et al., 2010). TWV values derived from radiosonde observations were then compared with the valid TWV values obtained from the field campaigns to compute correlation and RMSE. For TOC, OMI satellite products were used as reference data. Because OMI provides global coverage once per day at approximately 13:30 local time, a temporal difference of approximately 2–3 hours exists between the satellite observations and field campaign measurements. Considering the spatial resolution of OMI (13 × 24 km), the pixel containing each field campaign site was extracted and used for comparison. The extracted TOC values were then compared with the corresponding MICROTOPS II measurements.

\(\begin{align}e=6.112 \times exp (\frac {17.67 \;T_d}{T_d + 243.5})\end{align}\)       (12)

\(\begin{align}q = \frac {0.622e}{p-0.378e}\end{align}\)       (13)

\(\begin{align}\mathrm{TWV} \approx \frac{1}{g} \sum_{i} q_{i} \Delta p_{i}\end{align}\)       (14)

To evaluate the reliability of atmospheric parameters derived from MICROTOPS II measurements, uncertainty was estimated using standard uncertainty (u) and relative uncertainty (RU, %). Following international guidelines (JCGM 100:2008;KOLAS-G-001), Type A uncertainty was quantitatively evaluated using a statistical approach based on repeated measurements. The mean and standard deviation of daily observations for AOD, TWV, and TOC were calculated, from which absolute and relative uncertainties were derived. The purpose of this analysis is to assess whether the QC-processed atmospheric data provide consistent uncertainty levels comparable to internationally accepted standards, such as those reported for AERONET.

\(\begin{align}u=\sqrt{\frac{\sum_{i=1}^{N}\left(x_{i}-\mu\right)^{2}}{N}}\end{align}\)       (15)

\(\begin{align}RU(\%) = \frac {U(x_i)}{\mu}\end{align}\)       (16)

N: number of observations, μ: mean value,

xi : individual observation.

3. Results

3.1. Analysis and Evaluation Results of Valid AOD Data

The temporal distribution of MICROTOPS II AOD measurements acquired during the 2024 field campaigns for absolute radiometric calibration was examined both before and after quality-control processing. The processed Level 2 AOD data from SKYNET, used as a reference dataset, were also compared. On June 3 and November 12,substantial variability inAOD was observed before 11:00 AM, whereas on November 18, AOD exhibited relatively little temporal fluctuation. SKYNET AOD values, after QC, were provided only at sparse time intervals due to the removal of a large number of cloud-affected or low-quality observations. On June 3, the SKYNET AOD values showed a range similar to that of MICROTOPS II, whereas on the two November dates, the SKYNET AOD values were slightly lower than those measured by MICROTOPS II. The QC processed MICROTOPS II dataset showed effective removal of abnormal outliers with large deviations while retaining natural temporal variability in AOD. After QC, the MICROTOPS II AOD values exhibited a correlation coefficient of 0.817 and an RMSE of 0.015when compared with SKYNET data, indicating that MICROTOPS II tends to produce slightly higher AOD values relative to SKYNET.

AERONET AOD data from the Incheon site were also used as an additional reference dataset, as shown in the figure below. Because AERONET provides both pre- and post-QC datasets, the temporal variability ofAOD can be examined before and after quality control, compared with the MICROTOPS II measurements. As shown in Fig. 5(a), MICROTOPS II AOD values increased after 12:00 PM due to cloud influence, with a noticeable rise in variability; a similar pattern was observed in the AERONET data after 12:30 PM. These cloud-induced anomalies were removed through the QC process, resulting in significantly reduced temporal variability, as shown in Fig. 5(b). The daily mean AOD from MICROTOPS II decreased from 0.274 to 0.245 after QC, while the standard deviation decreased from 0.097 to 0.021. After QC, the comparison between the fixed-site AERONET observations and the field-measured MICROTOPS II data yielded a correlation coefficient of 0.835 and an RMSE of 0.009,which is smaller than the typical RMSE (~0.01) reported forQC-processed MICROTOPS II AOD (Ichoku et al., 2002). Consistent with the comparison using SKYNET data, the MICROTOPS II AOD values were generally higher than those from the fixed-site observations (Fig. 6).

OGCSBN_2026_v42n2_181_9_f0001.png 이미지

Fig. 4. Time-series data of AOD at 550 nm from MICROTOPS II and SKYNET. (a) June 2, 2024. (b) November 12, 2024. (c) November 18, 2024. (d) Scatter plot between two datasets.

OGCSBN_2026_v42n2_181_10_f0001.png 이미지

Fig. 5. AOD data from MICROTOPS II and AERONET measured in Incheon on May 29, 2024. (a) Before QC processing. (b) After QC processing.

OGCSBN_2026_v42n2_181_10_f0002.png 이미지

Fig. 6. Scatter plot of AERONETAOD at 550 nm versus MICROTOPS II AOD at 550 nm.

3.2. Analysis and Evaluation Results of Valid TWV Data

The MICROTOPS II sunphotometer also provides TWV measurements, which show a wide range of values depending on the meteorological conditions on each observation day. As shown in Fig. 7, the temporal variation pattern of TWV closely resembles that of AOD (Fig. 5). The TWV values selected through QC correspond to the timestamps that passed the AOD QC process. In particular, on June 3, a large portion of the data was filtered out due to measurement-induced variability rather than atmospheric changes (Fig. 7). On November 12, atmospheric variability was more pronounced compared to November 18, and this variability remained visible in the QC-processed dataset.

OGCSBN_2026_v42n2_181_10_f0003.png 이미지

Fig. 7. Time series of TWV observed by MICROTOPS II at Daejeon site from 10:00 to 12:30 KST.

The resulting TWV values were compared with radiosonde TWV data from Osan, the closest observation site to Daejeon. The comparison yielded a high correlation coefficient of 0.974, with the data points closely aligned along the 1:1 line (Fig. 8). MICROTOPS II tended to underestimate TWV relative to radiosonde measurements slightly, a pattern also reported in previous studies where radiosonde, MICROTOPS II, and CIMEL instruments were simultaneously compared (Campmany et al., 2010).

OGCSBN_2026_v42n2_181_10_f0004.png 이미지

Fig. 8. Scatter plot of TWV from MICROTOPS II versus radiosonde measurements.

3.3. Analysis and Evaluation Results of Valid TOC Data

Field campaigns for absolute radiometric calibration are typically conducted between 10:00 KST and 13:00 KST, during which the solar zenith angle is relatively small, and all measurements are obtained under air mass values below3. Under these conditions, the ozone column measurements can be considered valid. Fig. 9 presents the MICROTOPS II TOC measurements collected from May to December 2024, showing a seasonal pattern consistent with mid-latitude ozone climatology—highest in spring (March–May) and lowest in autumn to early winter (September–December) (Gómez-Amo et al., 2012). Following measurement, TOC estimates derived from the wavelength pairs (305.5–312.5 nm), (312.5–320.0 nm), and the weighted combination of all three (u(12)O3, u(23)O3, and u(123)O3) were produced, and their distributions are shown in Fig. 9. Across all dates, the three-band combination u(123)O3 yielded the highest TOC values, whereas the 312.5–320.0 nm combination (u(23)O3) produced the lowest values and exhibited the largest temporal variability. In particular, temporal variability in TOC increased from November to December compared to May (Fig. 10). This can be attributed to larger air mass values during late autumn and early winter, as well as enhanced variability in stratospheric temperature and atmospheric dynamics in midlatitudes during October–December. The increased solar zenith angle also increases air mass, enhancing measurement sensitivity and contributing to larger TOC fluctuations (Kasten & Young, 1989). To evaluate the accuracy of the representative TOC estimate u(123)O3, OMI TOC satellite data were used as the reference. The two datasets showed a strong correlation (≈0.952), with MICROTOPS II TOC values being approximately 60 DU higher but maintaining a linear relationship parallel to the 1:1 line (Fig. 11).

OGCSBN_2026_v42n2_181_11_f0001.png 이미지

Fig. 9. TOC data for three types (u(12)O3, u(23)O3, and u(123)O3) derived from daily MICROTOPS II measurements during the 2024 field campaigns.

OGCSBN_2026_v42n2_181_11_f0002.png 이미지

Fig. 10. Scatter plot of TOC measured by MICROTOPS IIversus airmass (gray: May 16, 2024; blue: May 29, 2024; green: Nov 12, 2024; red: Nov 18, 2024).

OGCSBN_2026_v42n2_181_11_f0003.png 이미지

Fig. 11. Scatter plot of MICROTOPS II TOC versus OMI TOC.

3.4. Uncertainity Analysis of QC-Processed Atmospheric Parameters

Table 4 summarizes the uncertainty of AOD, TWV, and TOC derived from QC-processed MICROTOPS II measurements across all field campaigns. The results show that AOD uncertainty ranged from 0.001 to 0.006, with most values below 0.003, indicating stable and consistent measurement quality. TWV uncertainty ranged from 0.001 to 0.011, with slightly larger variability compared to AOD, but generally remained below 0.01.In contrast, TOC exhibited the smallest uncertainty, ranging from 0.000 to 0.002 across all campaigns. In terms of relative uncertainty, AOD showed values within approximately 0.6–3%, while TWV and TOC exhibited lower relative uncertainties, generally below 1.2% and 0.6%, respectively. These results indicate that the proposed QC procedures effectively reduced variability and provided consistent uncertainty levels across different atmospheric parameters and observation conditions. Overall, the uncertainty levels obtained in this study are comparable to those reported for reference measurements such as AERONET, suggesting that the QC-processed MICROTOPS II data can provide reliable atmospheric inputs for absolute radiometric calibration.

Table 4. Uncertainty (u) and relative uncertainty (RU, %) of AOD, TWV, and TOC derived from QC-processed MICROTOPS II measurements for each field campaign

OGCSBN_2026_v42n2_181_12_t0001.png 이미지

4. Discussion

In previous studies related to MICROTOPS II, accuracy assessments primarily focused on the influence of measurement procedures, considering the instrument’s characteristics as a field-deployable device. In this study, measurement procedures were standardized based on established literature and the instrument manual, and accuracy was analyzed with an emphasis on data processing—including QC—rather than measurement methodology. The developed processing techniques and accuracy evaluation were specifically aimed at generating input data for the absolute radiometric calibration of optical satellite sensors. Since most field campaigns for optical satellite calibration are conducted between 10:30AM and 12:30 PM, the influence ofsolar geometry is minimized, and no additional corrections were required. Through the proposed approach, uncertainty information for atmospheric inputs can now be estimated more accurately at the time of field campaigns, improving the reliability of absolute radiometric calibration.

Unlike previous studies, this work placed the greatest emphasis on the QC procedure for AOD—the atmospheric variable with the highest sensitivity in absolute radiometric calibration—among the three parameters measured by MICROTOPS II (Giles et al., 2019). This prioritization is consistentwith previous studies indicating that AOD has a dominant influence on radiative transfer and calibration uncertainty, whereas TWV and TOC generally have comparatively smaller impacts depending on spectral regions (Vermote et al., 2002; Gao et al., 2009).

For TWV, since the temporal variability due to clouds was not significant, TWV values measured concurrently with QC-validated AOD were selected. This approach is also consistent with AERONET processing strategies, where water vapor retrievals are validated together with aerosol measurements rather than through an independent QC procedure (Giles et al., 2019).

For TOC, all measurements satisfied the air mass constraint and were therefore considered valid, and the representative TOC value was derived using the three-band combination recommended in the instrument manual. Although TOC is often used without strict QC procedures in existing calibration frameworks due to its relatively low sensitivity (Vermote et al., 2006; Gao et al., 2009), this study additionally considered its consistency with other atmospheric parameters.

By establishing QC procedures for AOD, TWV, and TOC in an integrated manner, this study provides atmospheric input datasets with consistent and quantifiable uncertainty characteristics. Such consistency is particularly important for absolute radiometric calibration, asit enables the uncertainty of calibration coefficients to more realistically reflect the actual conditions offield campaigns. Consequently, the proposed approach allows the use of atmospheric inputs with consistent accuracy, improving both the reliability and stability of absolute radiometric calibration.

However, certain studies have proposed an additional QC criterion in which only cases with differences less than 10 DU between the two-band and three-band TOC estimates are retained (Gómez-Amo et al., 2012). Applying this criterion in the present study reduced RMSE, as shown in Fig. 12, although this procedure is not yet used as an official QC standard. Future field campaigns will be used to further assess the applicability of this criterion. The QC procedures proposed in this study were shown to yield reliable MICROTOPS II AOD, TWV, and TOC datasets suitable for use in absolute radiometric calibration. In particular, AOD measurements benefited from cloud-effect removal, resulting in stable and high-quality datasets. Such data screening contributes to reducing the uncertainty associated with field-based absolute radiometric calibration by enhancing the reliability of atmospheric input parameters. While MICROTOPS II offers advantages such as ease of use and flexible temporal and spatial deployment, uncertainties remain due to the instrument’sinternal calibration constants and environmental constraints such as cloud conditions. Therefore, future research should focus on deriving calibration coefficients optimized for domestic regions and climate conditions.

OGCSBN_2026_v42n2_181_13_f0001.png 이미지

Fig. 12. Scatter plot of MICROTOPS II TOC versus OMI TOC before and after QC processing (gray: before QC, red: after QC).

5. Conclusions

In thisstudy, QC procedures were applied to obtain more accurate atmospheric parameters—AOD at 550 nm, TWV, and TOC—which are essential inputs for the absolute radiometric calibration of optical satellite imagery. Errors caused by cloud contamination and measurement variability were minimized, and the accuracy of the processed datasets was evaluated using multiple reference sources. MICROTOPS II AOD data processed through triplet stability and time-series consistency tests exhibited reduced temporal variability and showed strong agreement with QC-processed AERONET Level 1.5 data, with correlation coefficients exceeding 0.82. Although MICROTOPS II AOD tended to be slightly higher than fixed-site observations, RMSE values remained low (0.009–0.015). TWV values extracted at the same timestamps as validated AOD measurements showed very high correlation with radiosonde observations (r = 0.974). TOC, which exhibited notable seasonal variability, was estimated using the 305.5 nm, 312.5 nm, and 320.0 nm channels. TOC measurements obtained under air mass values below 3 were considered valid, and their comparison with OMI TOC satellite products showed a high correlation (≈0.952). In terms of uncertainty, the QC-processed MICROTOPS II measurements exhibited uncertainties of less than 0.01 for both AOD and TWV, and approximately 0.002 for TOC, which are comparable to the uncertainty levels reported for AERONET observations. These results demonstrate that atmospheric measurements obtained from MICROTOPS II can be transformed into high-quality datasets through appropriate processing and QC procedures. The validated datasets provide realistic, site-specific atmospheric information without compromising accuracy relative to fixed-site networks or satellite products, making them highly suitable for absolute radiometric calibration.

Furthermore, the quantified accuracy and variability of these atmospheric parameters can be directly incorporated into radiative transfer models, enabling more rigorous and reliable uncertainty estimation for absolute radiometric calibration. The validated datasets and QC framework presented in this study may also serve as useful benchmarks for evaluating future QC enhancements or new atmospheric measurement instruments. The atmospheric measurements acquired using the MICROTOPS II instrument represent one of the few practical sources of in situ atmospheric information obtainable during field campaigns for absolute radiometric calibration. Through the application of the proposed QC procedures, datasets with quantified uncertainties were generated, and consistent levels of uncertainty were achieved across individual measurements. These quality-controlled atmospheric inputs enable the uncertainty estimation of absolute radiometric calibration to more realistically reflect the actual conditions of field campaigns.

Author Contributions

Conceptualization: Kim SH, Jeong JH; Data curation: Choi JH, Baek IK; Methodology: ChoiJH, Choi YY; Formal analysis: Choi JH, Baek IK; Validation: Choi YY, Jeong JH; Project administration: Kim SH, Choi YY, Jeong JH; Funding acquisition, Supervision: Choi YY, Jeong JH; Writing–original draft: Kim SH, Choi JH; Writing–review & editing: All authors.

Conflicts of Interest

No potential conflict of interest relevant to this article was reported.

Funding

This research was supported by funding from the Korean government (KASA, Korea AeroSpace Administration) (grant number RS-2020-NR055936).

Data Availability Statement

The data used in this study cannot be provided due to security reasons.

Acknowledgments

None.

Supplementary Materials

None.

참고문헌

  1. Ångström, A. (1929). On the atmospheric transmission of sun radiation and on dust in the air. Geografiska Annaler, 11(2), 156–166. https://doi.org/10.1080/20014422.1929.11880498
  2. Berk, A., Conforti, P., Kennett, R., Perkins, T., Hawes, F., & van den Bosch, J. (2014). MODTRAN® 6: A major upgrade of the MODTRAN radiative transfer code. Proceedings of SPIE 9088, Algorithms and Technologies for Multispectral, Hyperspectral, and Ultraspectral Imagery XX, 90880H. https://doi.org/10.1117/12.2050433
  3. Bouvet, M., Thome, K., Berthelot, B., Bialek, A., Czapla-Myers, J., Fox, N.P., Goryl, P., Henry, P., Ma, L., & Woolliams, E. R.(2019). RadCalNet: A radiometric calibration network for earth observing imagers operating in the visible to shortwave infrared spectral range. Remote Sensing, 11(20), 2401. https://doi.org/10.3390/rs11202401
  4. Brion, J., Chakir, A., Charbonnier, J., Daumont, D., Parisse, C., & Malicet, J. (1998). Absorption spectra measurements for the ozone molecule in the 350–830 nm region. Journal of Atmospheric Chemistry, 30, 291–299. https://doi.org/10.1023/A:1006036924364
  5. Bruegge, C. J., Conel, J. E., Green, R. O., Margolis, J. S., Holmgren, E. A., Toon, G. C., & Young, S. J. (1992). Water vapor column abundance retrievals during FIFE. Journal of Geophysical Research: Atmospheres, 97(D17), 18759–18768. https://doi.org/10.1029/92JD01050
  6. Bu, J. O., Han, J. H., Jeong, J., Hwang, S. O., & Won, D. (2024). A comparison between ground-based measurements of AOD products in South Korea: Case study of PFR and skyradiometer. Particulate and Aerosol Research, 20(4), 139–147. https://doi.org/10.11629/jpaar.2024.20.4.139
  7. Campmany, E., Bech, J., Rodríguez-Marcos J., Sola,Y., & Lorente, J. (2010). A comparision of total precipitable water measurements from radiosonde and sunphotometers. Atmospheric Research, 97(3), 385–392. https://doi.org/10.1016/j.atmosres.2010.04.016
  8. Choi, Y., Ghim, Y. S., & Sohn, B.J.(2011).Comparison of aerosol optical properties from different models of skyradiometer. Atmosphere, 21(3), 311–317. https://doi.org/10.14191/Atmos.2011.21.3.311
  9. Eck, T. F., Holben, B. N., Reid, J. S., Dubovik, O., Smirnov, A., O'Neill, N. T., Slutsker, I., & Kinne, S.(1999). Wavelength dependence of the optical depth of biomass burning, urban, and desert dust aerosols. Journal of Geophysical Research: Atmospheres, 104(D24), 31333–31349. https://doi.org/10.1029/1999JD900923
  10. Elliott, W. P., & Gaffen, D. J. (1991). On the utility of radiosonde humidity archives for climate studies. Bulletin of the American Meteorological Society, 72(10), 1507–1520. https://doi.org/10.1175/1520-0477(1991)072%3C1507:OTUORH%3E2.0.CO;2
  11. Free, M., Seidel, D. J., Angell, J. K., Lanzante, J., Durre, I., & Peterson, T. C. (2005). Radiosonde atmospheric temperature products for assessing climate (RATPAC): A new dataset and comparisons with other upper-air data. Journal of Geophysical Research: Atmospheres, 110, D22101. https://doi.org/10.1029/2005JD006169
  12. Gao, B. C., Montes, M. J., Davis, C. O., & Goetz, A. F. H. (2009). Atmospheric correction algorithms for hyperspectral remote sensing data of land and ocean. Remote Sensing of Environment, 113(S1), S17–S24. https://doi.org/10.1016/j.rse.2007.12.015
  13. Giles, D. M., Sinyuk, A., Sorokin, M. G., Schafer, J. S., Smirnov, A., Slutsker, I., Eck, T. F., Holben, B. N., Lewis, J. R.,Campbell, J. R., Welton, E.J., & Lyapustin, A. I. (2019).Advancements in the AERONET version 3 aerosol retrieval algorithm. Remote Sensing, 11(18), 2166. https://doi.org/10.5194/amt-12-169-2019
  14. Giles, D. M., Sinyuk, A., Smirnov, A., Holben, B. N., Slutsker, I., Eck, T. F., Giles D. M., Slusker I., Korkin, S., Schafer, J. S., Smirnov, A., Sorokin, M., & Lyapustin, A. (2020). The AERONET version 3 aerosol retrieval algorithm, associated uncertainties and applications.Atmospheric Measurement Techniques, 13(6), 3375–3411. https://doi.org/10.5194/amt-13-3375-2020
  15. Gómez-Amo, J. L., Estellés, V., di Sarra, A., Pedrós, R., & Utrillas, M. (2012). Operational considerations to improve total ozone measurements with a Microtops II ozone monitor. Atmospheric Measurement Techniques, 5, 759–770. https://doi.org/10.5194/amt-5-759-2012
  16. Gong, S.(2018). Evaluation of maritime aerosol optical depth and precipitable water vapor content from the MICROTOPS II sun photometer. Optik, 169, 1–7. https://doi.org/10.1016/j.ijleo.2018.05.025
  17. Gorshelev, V., Serdyuchenko, A., Weber, M., Chehade, W., & Burrows, J. P. (2014). High spectral resolution ozone absorption cross-sections – Part 1: Measurements in the 200–1100 nm region. Atmospheric Measurement Techniques, 7, 609–623. https://doi.org/10.5194/amt-7-609-2014
  18. Holben, B. N., Eck, T. F., Slutsker, I., Tanré, D., Buis, J. P., Setzer, A., Vermote, E., Reagan, J. A., Kaufman, Y. J., Nakajima, T., Lavenu, F., Jankowiak, I., & Smirnov, A. (1998). AERONET—A federated instrument network and data archive for aerosol characterization. Remote Sensing of Environment, 66(1), 1–16. https://doi.org/10.1016/S0034-4257(98)00031-5
  19. Ichoku, C., Levy, R., Kaufman, Y. J., Remer, L. A., Li, R. R., Martins, V. J., Holben, B. N., Abuhassan, N., Slutsker, I., Eck, T. F., & Pietras, C. (2002). Analysis of the performance characteristics of the five-channel Microtops II sun photometer for measuring aerosol optical thickness and precipitable water vapor. Journal of Geophysical Research: Atmospheres, 107(D13), AAC5-1–AAC5-17. https://doi.org/10.1029/2001JD001302
  20. Jin, C., Ahn, H., Seo, D., & Choi, C. (2020).Radiometric calibration and uncertainty analysis of KOMPSAT-3A using the reflectance-based method. Sensors, 20(9), 2564. https://doi.org/10.3390/s20092564
  21. Kasten, F., & Young, A. T. (1989). Revised optical air mass tables and approximation formula.Applied Optics, 28(22), 4735–4738. https://doi.org/10.1364/AO.28.004735
  22. Komhyr, W. D., Grass, R. D., & Leonard, R. K. (1989). Dobson spectrophotometer 83: A standard for total ozone measurements, 1962–1987. Journal of Geophysical Research, 94(D7), 9847–9861. https://doi.org/10.1029/JD094iD07p09847
  23. Lee, K. H., & Lee, K. T. (2016). Aerosol optical thickness measurements from the Microtops-II multi-wavelength radiometer. Journal of the Korean Society for Atmospheric Environment, 32(1), 57–66. https://doi.org/10.5572/KOSAE.2016.32.1.057
  24. Lin, C. A., Zhang,Y., Heath, G., Henze, D.K., Sengupta, M., & Lu, C. H. (2023). Improvement of aerosol optical depth data forl ocalized solar resource assessment. Solar Energy, 249, 457–466. https://doi.org/10.1016/j.solener.2022.11.047
  25. Liu, Y. K., Ma, L. L., Wang, N., Qian, Y. G., Zhao, Y. G., Qiu, S., Gao, C. X., Long, X. X., & Li, C. R. (2020). On-orbit radiometric calibration of the optical sensors on-board SuperView-1 satellite using three independent methods. OpticsExpress, 28(8), 11085–11105.https://doi.org/10.1364/OE.388387
  26. Luers, J. K., & Eskridge, R. E. (1995). Temperature corrections for the VIZ and Vaisala radiosondes. Journal of Applied Meteorology and Climatology, 34(6), 1241–1253. https://doi.org/10.1175/1520-0450(1995)034%3C1241:TCFTVA%3E2.0.CO;2
  27. Maxar Technologies. (2020, Feb 26). Why you should care about absolute radiometric calibration, Vantor. https://www.vantor.com/blog/why-you-should-care-about-absolute-radiometric-calibration
  28. Morys, M., Mims, F. M., Hagerup, S.,Anderson, S. E., Baker,A., & Kia, J. (2001). Design, calibration, and performance of MICROTOPS II handheld ozone monitor and sun photometer.Journal of Geophysical Research: Atmospheres, 106, 14573–14582. https://doi.org/10.1029/2001JD900103
  29. Noh, Y. (2021). A long-term trend of annual average aerosol optical properties in the Korean Peninsula using AERONET sun/sky radiometer data. Journal of Korean Society for Atmospheric Environment, 37(3), 456–465. https://doi.org/10.5572/KOSAE.2021.37.3.456
  30. Park, M., Lee, K., & Choi, M. (2023). Spectral and spatial dependencies in the validation of satellite-based aerosol optical depth from the geostationary ocean color imager using the aerosol robotic network. Remote Sensing, 15(14), 3621. https://doi.org/10.3390/rs15143621
  31. Parlange, M. B., & Brutsaert, W. (1989). Regional roughness of the land surface and its effects on the atmospheric boundary layer. Boundary-Layer Meteorology, 48, 69–81. https://doi.org/10.1007/BF00121783
  32. Purnama, D. R., Zulistyawan K. A., Pradita, N., Haryanto, Y. D., Riama, N. F.,& Briliano, B. (2023).Retrieval and evaluation of maritime AOD and PWV from ship-based MICROTOPS II sun photometers and their seasonal variability using MODIS measurements over the Indonesian Throughflow region. In S. Lestari et al. (Eds.), Proceedings of the International Conference on Radioscience, Equatorial Atmospheric Science and Environment and Humanosphere Science: INCREASE 2023 (pp. 35–46). Springer. https://doi.org/10.1007/978-981-97-0740-9_4
  33. Qiao, C., Liu, S., Huo, J., Mu, X., Wang, P., Jia, S., Fan, X., & Duan M.(2023). Retrievals of precipitable water vapor and aerosol optical depth from direct sun irradiance measurements. Atmospheric Measurement Techniques, 16(6), 1539–1549. https://doi.org/10.5194/amt-16-1539-2023
  34. Revel, C., Gascon, F., Lacherade, S., Fougnie, B., & Clerc, S. (2019). Sentinel-2A and 2B absolute calibration monitoring. European Journal of Remote Sensing, 52(1), 122–137. https://doi.org/10.1080/22797254.2018.1562311
  35. Schmid, B., Michalsky, J. J., Slater, D. W., Barnard, J.C., Halthore, R. N., Liljegren, J. C., Holben, B. N., Eck, T. F., Livingston, J. M., Russell, P. B., Ingold, T., & Slutsker, I. (2001). Comparison of column water vapor measurements from solar transmittance methods. Applied Optics, 40(12), 1886–1896. https://doi.org/10.1364/AO.40.001886
  36. Smirnov, A., Holben, B. N., Eck, T. F., Dubovik, O., & Slutsker, I. (2000).Cloud-screening and quality control algorithms for the AERONET database. Remote Sensing of Environment, 73(3), 337–349. https://doi.org/10.1016/S0034-4257(00)00109-7
  37. Solar Light Company. (2001). MICROTOPS II user's guide (version 5.31). Solar Light Company.
  38. Solar Light Company. (2016). MICROTOPS II ozone monitor and sunphotometer user manuals. Solar Light Company. https://www.solarlight.com/user-manuals?utm_source=chatgpt.com
  39. Solar Light Company. (2023). Model 520 and 540 MICROTOPS II series – Atmospheric instrumentation product information. Solar Light Company. https://www.solarlight.com/product-category/products/atmospheric-instrumentation
  40. Tang, H., Xie, J., Chen, W., Zhang, H., & Wang, H. (2023). Absolute radiometric calibration of ZY3-02 satellite multispectral imager based on irradiance-based method. Remote Sensing, 15(2), 448. https://doi.org/10.3390/rs15020448
  41. Vermote, E. F., El Saleous, N. Z., & Justice,C. O. (2002). Atmospheric correction of MODIS data in the visible to middle infrared: First results. Remote Sensing of Environment, 83(1–2), 97–111. https://doi.org/10.1016/S0034-4257(02)00089-5
  42. Wallace, J. M., & Hobbs, P. V. (2006). Atmospheric science: An introductory survey (2nd ed.). Academic Press. https://doi.org/10.1088/0031-9112/29/6/037
  43. Yeom, J., Ko, J., Hwang, J., Lee, C. S., Choi, C. U., & Jeong, S. (2018). Updating absolute radiometric characteristics for KOMPSAT-3 and KOMPSAT-3A multispectral imaging sensors using well-characterized pseudo-invariant tarps and MICROTOPS II. Remote Sensing, 10(5), 697. https://doi.org/10.3390/rs10050697
  44. Zhai, P., & Eskridge, R. E. (1996). Analyses of inhomogeneities in radiosonde temperature and humidity time series. Journal of Climate, 9(4), 884–894. https://doi.org/10.1175/1520-0442(1996)009<0884:AOIIRT>2.0.CO;2