When negative electron in x-ray tube is accelerated in to a high speed and then the currency of the electron is blocked by the target, x-ray happens by the conversion of the energy. The real area where the fast accelerated electron collides to a target area is called actual focal spot. When the string focused size is observed at the central ray side, where the direction x-ray comes out, the size seems to be reduced. This focus is called effective focal spot. According to radiation angle of x-rays tube, the degree of the negative pole side presents higher value than inclination, the amount of exposed radiation that patient receives differs by the angle of positive pole, which means effective focal spot is the variable. This paper presents the correlation between size of effective focal spot and amount of exposed radiation to the patient by it, and effective research for homogenized dose dispersion by the size of effective focal spot. In conclusion, following the focal size, effective range which was -8cm ~ 0 cm on average, was found and average dose rate was 0.019 R/min. Through this range, for patients with small radiation exposure, image with good density and resolution in aspect of diagnosing will be able to be obtained.
Kim, Ye-Seul;Park, Hye-Suk;Park, Su-Jin;Kim, Hee-Joung
Progress in Medical Physics
/
v.23
no.1
/
pp.26-32
/
2012
The magnification technique has recently become popular in bone radiography, mammography and other diagnostic examination. However, because of the finite size of X-ray focal spot, the magnification influences various imaging properties with resolution, noise and contrast. The purpose of study is to investigate the influence of magnification and focal spot size on digital imaging system using eDQE (effective detective quantum efficiency). Effective DQE is a metric reflecting overall system response including focal spot blur, magnification, scatter and grid response. The adult chest phantom employed in the Food and Drug Administration (FDA) was used to derive eDQE from eMTF (effective modulation transfer function), eNPS (effective noise power spectrum), scatter fraction and transmission fraction. According to results, spatial frequencies that eMTF is 10% with the magnification factor of 1.2, 1.4, 1.6, 1.8 and 2.0 are 2.76, 2.21, 1.78, 1.49 and 1.26 lp/mm respectively using small focal spot. The spatial frequencies that eMTF is 10% with the magnification factor of 1.2, 1.4, 1.6, 1.8 and 2.0 are 2.21, 1.66, 1.25, 0.93 and 0.73 lp/mm respectively using large focal spot. The eMTFs and eDQEs decreases with increasing magnification factor. Although there are no significant differences with focal spot size on eDQE (0), the eDQEs drops more sharply with large focal spot than small focal spot. The magnification imaging can enlarge the small size lesion and improve the contrast due to decrease of effective noise and scatter with air-gap effect. The enlargement of the image size can be helpful for visual detection of small image. However, focal spot blurring caused by finite size of focal spot shows more significant impact on spatial resolution than the improvement of other metrics resulted by magnification effect. Based on these results, appropriate magnification factor and focal spot size should be established to perform magnification imaging with digital radiography system.
An optical system for iris pattern recognition of the human eye is designed. The lens system is triplet type, and characterized to minimize longitudinal chromatic aberration and Petzval sum. The distance from object to image is 200 mm-300 mm and the effective focal length is 50 mm. Performance of the imaging system is assessed by calculating the ray-fan and spot diagram for Fraunhofer C, d and F line for object height 0 mm, 4 mm and 6 mm. Furthermore, MTFs are calculated. All of the spot sizes are less than 0.05 mm in diameter. The MTF values are higher than 0.5 in the spatial frequency range up to 20 lines/mm for all of the object heights. ights.
We studied the compact anode design to develop 100 kW rotating anode X-ray tube with large focal spot 1.2 mm, small focal spot 0.6 mm and tube voltage 150 kV for large hospital digital radiography using computer thermal simulation. The larger thermal radiation effect in a high vacuum can reduce the temperature of anode so the method to increase the surface area of anode is investigated. The anode has the multi-tier shape at the back side of TZM body of anode and also bigger diameter of anode. The number of multi-tiers was varied from 6 to 15 and the diameter of anode was also varied from ${\Phi}74$ to ${\Phi}82$. From ANSYS transient thermal simulation result, we could obtain $1056.4^{\circ}C$ anode maximum temperature when applying 100 kW input power at 0.1 second on target focal track which is less than $1091^{\circ}C$ of the conventional 75 kW X-ray tube with reduced anode weight by 15.5% than the conventional anode. The compact anode of reduced anode weight is able to improve the unwanted noise when the rotor is rotating at high-speed and also reduce the rotational torque which the cost effective stator-coil is possible. It is believed that the anode with 15 ea multi-tiers using ${\Phi}82$ can satisfy with the specification of the anode heat capacity. From the results of this paper, it has been confirmed that the proposed compact anode can be used as the anode of 100 kW rotating anode X-ray tube for digital radiography.
The purpose of this study was tried to evaluate the filling sodium states inside the fuel rod of sodium-cooled fast reactor by digital medical X-ray. We used the diagnostic X-ray generators in digital radiography (DR). This study have found the optimal conditions by changing the effective focal spot size of X-ray tube and post-processing of the DR method with a tungsten edge plate in order to ensure excellent sharpness At this time, the sharpness and resolution were evaluated using the MTF (modulation transfer function). As a result, this study obtained a spatial resolution of 3.871 lp/mm (0.1 MTF), 3.290 lp/mm (0.5 MTF) when implemented the contrast strengthen post-processing in small focal spot. In this research, the result is able to evaluate the level of sodium inside the fuel rod by using the diagnostic X-ray generators in medical digital radiographic images.
X-ray mammography is the most effective method for the diagnosis of calcified lesions of various breast diseases. To reduce patient dose and to obtain optimal image required for diagnosis, the performance of the mammography system should be maintained continuously. Because the target (anode) angle of the X-ray tube is measured from the central X-ray, the effective angle can be slightly different in view of the position on the detector, which can result in degrading spatial resolution of the imaging within the field of view. In this study, we measured the MTF to examine spatial resolution for positions on the detector in the digital mammography system. For a tungsten wire of $50{\mu}m$ diameter, the highest spatial frequency was obtained. It meant that a wire diameter for measuring MTF through LSF should be small compared to the pixel size of the detector used in the mammography system. The spatial resolution showed slightly different performance according to positions on the detector. The center position gave the best spatial resolution and positions away from the center showed the degraded performance although the difference of the spatial resolution was small. The effective focal spot size of the full width at half maximum also showed similar result. It concluded that the slightly increase of the effective focal spot size gave the degradation of the spatial resolution for positions on the detector.
In this paper, we propose a new type of a planar waveguide lens, a meniscus lens instead of a plano-convex lens used widely for the optical integrated circuits such as optical numerical processors. A plano-convex lens has a spherical aberration caused by the linear boundary. In the proposed meniscus lens, this linear boundary is replaced by a circular-boundary, and the spherical aberration is removed. To test the performance of the proposed lens, a meniscus lens is designed using Fermat principle and fabricated on the Y-cut $LiNbO_3$ substrate. First a $Ti:LiNbO_3$ planar waveguide is fabricated on the $LiNbO_3$ substrate by Ti indiffusion. Then a meniscus lens is fabricated on the planar waveguide by TIPE(Ti indiffued proton-exchange) method. A plano-convex lens is also fabricated on the same substrate to compare the spot size at the focal point. The measured spot size of the meniscus lens is reduced to 59% that of the plano-convex lens. This result shows that the proposed meniscus lens is more effective for optical integrated circuits than the plano-convex lens.
We have preliminarily designed two infrared optical systems of the multi-purpose infrared camera system (MIRIS) which is the main payload of STSAT-3. Each optical system consists of a Cassegrain telescope, a field lens and a 1:1 re-imaging lens system that is essential for providing a cold stop. The Cassegrain telescope is identical for both of two infrared cameras, but the field correction lens and re-imaging lens system are different from each other because of different bands of wavelength. The effective aperture size is 100mm in diameter and the focal ratio is f/5. The total length of the optical system is 300mm and the position of the cold stop is 25mm from the detector focal plane. The RMS spot size is smaller than $40{\mu}m$ over the whole detector plane.
The performance of ultrasonic testing systems for industrial or medical purpose largely depends on the performance of ultrasonic transducers. Generally, the information about an ultrasonic transducer performance characteristics are expressed by the ultrasonic R/F signal back from a reflector and its frequency characteristics in the data sheet provided by manufacturers. In case of focused ultrasonic transducers, the two pieces of information can, however, hardly assure that the focused ultrasonic transducer would produce well-focused C-scan images. Therefore, we propose the measured size of focal spot and the reconstructed shape of effective focal zone in the focused sound field as novel measures for the performance evaluation of the focused ultrasonic transducers. The process of getting the both measures of the transducers is conducted by the implemented software including sound field scanning and virtual 3D reconstruction functions which requires the echo of a point reflector. The proposed method could, otherwise impossible in the existing method, effectively and simply distinguish superior ones among many transducers made in the same specification and be also used to detect the performance degradation due to the aging of the transducers. Eventually, the quality of performance of the ultrasonic testing systems for industrial or medical purpose is secured.
Journal of the Korean Society of Laryngology, Phoniatrics and Logopedics
/
v.32
no.2
/
pp.51-55
/
2021
There are several lasers available for office-based or suspension microlaryngoscopy laser procedures in the treatment of laryngeal diseases. Each has advantages and disadvantages given the depth of penetration per unit of power, absorption in water, spectral absorption characteristics, mode of delivery, safety, and cost. It is important to note that while the proper selection of indication of treatment based on a laser wavelength is critical, of equal importance is selecting the appropriate power setting, focal length (or spot size), and time of exposure. The photoangiolytic lasers precisely target hemoglobin within the microcirculation of the highly vascularized tissue and may have better hemostatic effects and preservation of surrounding normal tissue than the CO2 laser. Although the choice of laser is purely theoretical and cannot be accurately concluded which parameters of laser (wattage and pulse width) were best to use, photoangiolytic laser surgery is safe and effective for specific laryngeal lesions. In this review, indications for photoangiolytic laser procedures for various laryngeal diseases, laser settings and surgical techniques for specific laryngeal lesions including sulcus vocalis, laryngeal dysplasia, and recurrent respiratory papillomatosis will be introduced. Pros and cons of in-office laser surgery using photoangiolytic laser and flexible CO2 laser will also be addressed.
본 웹사이트에 게시된 이메일 주소가 전자우편 수집 프로그램이나
그 밖의 기술적 장치를 이용하여 무단으로 수집되는 것을 거부하며,
이를 위반시 정보통신망법에 의해 형사 처벌됨을 유념하시기 바랍니다.
[게시일 2004년 10월 1일]
이용약관
제 1 장 총칙
제 1 조 (목적)
이 이용약관은 KoreaScience 홈페이지(이하 “당 사이트”)에서 제공하는 인터넷 서비스(이하 '서비스')의 가입조건 및 이용에 관한 제반 사항과 기타 필요한 사항을 구체적으로 규정함을 목적으로 합니다.
제 2 조 (용어의 정의)
① "이용자"라 함은 당 사이트에 접속하여 이 약관에 따라 당 사이트가 제공하는 서비스를 받는 회원 및 비회원을
말합니다.
② "회원"이라 함은 서비스를 이용하기 위하여 당 사이트에 개인정보를 제공하여 아이디(ID)와 비밀번호를 부여
받은 자를 말합니다.
③ "회원 아이디(ID)"라 함은 회원의 식별 및 서비스 이용을 위하여 자신이 선정한 문자 및 숫자의 조합을
말합니다.
④ "비밀번호(패스워드)"라 함은 회원이 자신의 비밀보호를 위하여 선정한 문자 및 숫자의 조합을 말합니다.
제 3 조 (이용약관의 효력 및 변경)
① 이 약관은 당 사이트에 게시하거나 기타의 방법으로 회원에게 공지함으로써 효력이 발생합니다.
② 당 사이트는 이 약관을 개정할 경우에 적용일자 및 개정사유를 명시하여 현행 약관과 함께 당 사이트의
초기화면에 그 적용일자 7일 이전부터 적용일자 전일까지 공지합니다. 다만, 회원에게 불리하게 약관내용을
변경하는 경우에는 최소한 30일 이상의 사전 유예기간을 두고 공지합니다. 이 경우 당 사이트는 개정 전
내용과 개정 후 내용을 명확하게 비교하여 이용자가 알기 쉽도록 표시합니다.
제 4 조(약관 외 준칙)
① 이 약관은 당 사이트가 제공하는 서비스에 관한 이용안내와 함께 적용됩니다.
② 이 약관에 명시되지 아니한 사항은 관계법령의 규정이 적용됩니다.
제 2 장 이용계약의 체결
제 5 조 (이용계약의 성립 등)
① 이용계약은 이용고객이 당 사이트가 정한 약관에 「동의합니다」를 선택하고, 당 사이트가 정한
온라인신청양식을 작성하여 서비스 이용을 신청한 후, 당 사이트가 이를 승낙함으로써 성립합니다.
② 제1항의 승낙은 당 사이트가 제공하는 과학기술정보검색, 맞춤정보, 서지정보 등 다른 서비스의 이용승낙을
포함합니다.
제 6 조 (회원가입)
서비스를 이용하고자 하는 고객은 당 사이트에서 정한 회원가입양식에 개인정보를 기재하여 가입을 하여야 합니다.
제 7 조 (개인정보의 보호 및 사용)
당 사이트는 관계법령이 정하는 바에 따라 회원 등록정보를 포함한 회원의 개인정보를 보호하기 위해 노력합니다. 회원 개인정보의 보호 및 사용에 대해서는 관련법령 및 당 사이트의 개인정보 보호정책이 적용됩니다.
제 8 조 (이용 신청의 승낙과 제한)
① 당 사이트는 제6조의 규정에 의한 이용신청고객에 대하여 서비스 이용을 승낙합니다.
② 당 사이트는 아래사항에 해당하는 경우에 대해서 승낙하지 아니 합니다.
- 이용계약 신청서의 내용을 허위로 기재한 경우
- 기타 규정한 제반사항을 위반하며 신청하는 경우
제 9 조 (회원 ID 부여 및 변경 등)
① 당 사이트는 이용고객에 대하여 약관에 정하는 바에 따라 자신이 선정한 회원 ID를 부여합니다.
② 회원 ID는 원칙적으로 변경이 불가하며 부득이한 사유로 인하여 변경 하고자 하는 경우에는 해당 ID를
해지하고 재가입해야 합니다.
③ 기타 회원 개인정보 관리 및 변경 등에 관한 사항은 서비스별 안내에 정하는 바에 의합니다.
제 3 장 계약 당사자의 의무
제 10 조 (KISTI의 의무)
① 당 사이트는 이용고객이 희망한 서비스 제공 개시일에 특별한 사정이 없는 한 서비스를 이용할 수 있도록
하여야 합니다.
② 당 사이트는 개인정보 보호를 위해 보안시스템을 구축하며 개인정보 보호정책을 공시하고 준수합니다.
③ 당 사이트는 회원으로부터 제기되는 의견이나 불만이 정당하다고 객관적으로 인정될 경우에는 적절한 절차를
거쳐 즉시 처리하여야 합니다. 다만, 즉시 처리가 곤란한 경우는 회원에게 그 사유와 처리일정을 통보하여야
합니다.
제 11 조 (회원의 의무)
① 이용자는 회원가입 신청 또는 회원정보 변경 시 실명으로 모든 사항을 사실에 근거하여 작성하여야 하며,
허위 또는 타인의 정보를 등록할 경우 일체의 권리를 주장할 수 없습니다.
② 당 사이트가 관계법령 및 개인정보 보호정책에 의거하여 그 책임을 지는 경우를 제외하고 회원에게 부여된
ID의 비밀번호 관리소홀, 부정사용에 의하여 발생하는 모든 결과에 대한 책임은 회원에게 있습니다.
③ 회원은 당 사이트 및 제 3자의 지적 재산권을 침해해서는 안 됩니다.
제 4 장 서비스의 이용
제 12 조 (서비스 이용 시간)
① 서비스 이용은 당 사이트의 업무상 또는 기술상 특별한 지장이 없는 한 연중무휴, 1일 24시간 운영을
원칙으로 합니다. 단, 당 사이트는 시스템 정기점검, 증설 및 교체를 위해 당 사이트가 정한 날이나 시간에
서비스를 일시 중단할 수 있으며, 예정되어 있는 작업으로 인한 서비스 일시중단은 당 사이트 홈페이지를
통해 사전에 공지합니다.
② 당 사이트는 서비스를 특정범위로 분할하여 각 범위별로 이용가능시간을 별도로 지정할 수 있습니다. 다만
이 경우 그 내용을 공지합니다.
제 13 조 (홈페이지 저작권)
① NDSL에서 제공하는 모든 저작물의 저작권은 원저작자에게 있으며, KISTI는 복제/배포/전송권을 확보하고
있습니다.
② NDSL에서 제공하는 콘텐츠를 상업적 및 기타 영리목적으로 복제/배포/전송할 경우 사전에 KISTI의 허락을
받아야 합니다.
③ NDSL에서 제공하는 콘텐츠를 보도, 비평, 교육, 연구 등을 위하여 정당한 범위 안에서 공정한 관행에
합치되게 인용할 수 있습니다.
④ NDSL에서 제공하는 콘텐츠를 무단 복제, 전송, 배포 기타 저작권법에 위반되는 방법으로 이용할 경우
저작권법 제136조에 따라 5년 이하의 징역 또는 5천만 원 이하의 벌금에 처해질 수 있습니다.
제 14 조 (유료서비스)
① 당 사이트 및 협력기관이 정한 유료서비스(원문복사 등)는 별도로 정해진 바에 따르며, 변경사항은 시행 전에
당 사이트 홈페이지를 통하여 회원에게 공지합니다.
② 유료서비스를 이용하려는 회원은 정해진 요금체계에 따라 요금을 납부해야 합니다.
제 5 장 계약 해지 및 이용 제한
제 15 조 (계약 해지)
회원이 이용계약을 해지하고자 하는 때에는 [가입해지] 메뉴를 이용해 직접 해지해야 합니다.
제 16 조 (서비스 이용제한)
① 당 사이트는 회원이 서비스 이용내용에 있어서 본 약관 제 11조 내용을 위반하거나, 다음 각 호에 해당하는
경우 서비스 이용을 제한할 수 있습니다.
- 2년 이상 서비스를 이용한 적이 없는 경우
- 기타 정상적인 서비스 운영에 방해가 될 경우
② 상기 이용제한 규정에 따라 서비스를 이용하는 회원에게 서비스 이용에 대하여 별도 공지 없이 서비스 이용의
일시정지, 이용계약 해지 할 수 있습니다.
제 17 조 (전자우편주소 수집 금지)
회원은 전자우편주소 추출기 등을 이용하여 전자우편주소를 수집 또는 제3자에게 제공할 수 없습니다.
제 6 장 손해배상 및 기타사항
제 18 조 (손해배상)
당 사이트는 무료로 제공되는 서비스와 관련하여 회원에게 어떠한 손해가 발생하더라도 당 사이트가 고의 또는 과실로 인한 손해발생을 제외하고는 이에 대하여 책임을 부담하지 아니합니다.
제 19 조 (관할 법원)
서비스 이용으로 발생한 분쟁에 대해 소송이 제기되는 경우 민사 소송법상의 관할 법원에 제기합니다.
[부 칙]
1. (시행일) 이 약관은 2016년 9월 5일부터 적용되며, 종전 약관은 본 약관으로 대체되며, 개정된 약관의 적용일 이전 가입자도 개정된 약관의 적용을 받습니다.