Measurement and Analysis of X-ray dosage change by variable kVp using the Photoconductive Property of Amorphous Selenium
Amorphous Selenium 의 광도전특성을 이용한 의료용 X선의 kVp 변화에 따른 선량 측정분석
- Cho, S.Y. (Department of Biomedical Engineering, College of Health, Inje University) ;
- Choi, J.Y. (Department of Biomedical Engineering, College of Health, Inje University) ;
- Lim, C.D. (Department of Biomedical Engineering, College of Health, Inje University) ;
- Nam, S.H. (Department of Biomedical Engineering, College of Health, Inje University)
- Published : 1996.11.15
Abstract
This research is the basical experiment for designing kVp meter using the photo-conductive property of Amorphous Selenium. At first, we charged the surface with A re discharge method, and changed the formed surface potential voltage by x-raying on that surface. As a result the variations of the surface charge increased by the rising kVp, while maintaining a constant exposure time and mA. We become to know that we can apply the photoconductive property of Amorphous Selenium in designing kVp meter.