A study on measurement apparatus for ferroelectricity in ferroelectrics

강유전특성 측정장치의 연구개발

  • Published : 1997.07.21

Abstract

This paper is to study and develope a measurement apparatus for ferroelectricity. The apparatus consists of wave generation part, high voltage amplifier part, measurement part, data acquisition part and the related controll circuits. Single or double excitation wave is digitalized and sent to the external RAM of wave generation part by personal computer. These datas saved in the RAM are converted to analog excitation wave through D/A converter. The frequency of excitation wave is depend on the read-out speed of the RAM by clock pulse. Such generated wave is applied to high voltage amplifier as a input voltage. The output of high voltage amplifier is applied to ferroelectrics and the response is obtained from the charge amplifier of measurement part. The response is sampled and converted to digital datas through AID converter. These digital datas are automatically saved in the external RAM of acquisition part. The computer takes the digital datas and calculates the electric displacement D, the electric field and the dielectric constant $\varepsilon$. We tested for PZT ceramic sample and could observed the D-E hysteresis lops and ${\varepsilon}_s$-E hysteresis loops with good forms.

Keywords