Quantitative Evaluation of Image Quality using Automatic Exposure Control

AEC 사용 유무에 따른 영상의 화질평가

  • Kim, Gwang-ho (Dept. of Radiology Science, College of Health sciences, Catholic University of Pusan) ;
  • Yoo, Gwan-ju (Dept. of Radiology Science, College of Health sciences, Catholic University of Pusan) ;
  • Kang, Ji-won (Dept. of Radiology Science, College of Health sciences, Catholic University of Pusan) ;
  • Ko, Bong-joo (Dept. of Radiology Science, College of Health sciences, Catholic University of Pusan) ;
  • Kim, Kyoun-lan (Dept. of Radiology Science, College of Health sciences, Catholic University of Pusan) ;
  • Kim, Min-hee (Dept. of Radiology Science, College of Health sciences, Catholic University of Pusan) ;
  • Yoon, Young-woo (Wallace Memorial Baptist Hospital) ;
  • Ye, Soo-young (Dept. of Radiology Science, College of Health sciences, Catholic University of Pusan) ;
  • Choi, Seok-yoon (Dept. of Radiology Science, College of Health sciences, Catholic University of Pusan)
  • 김광호 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 유관주 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 강지원 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 고봉주 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 김경란 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 김민희 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 윤영우 ;
  • 예수영 (부산가톨릭대학교 보건과학대학 방사선학과) ;
  • 최석윤 (부산가톨릭대학교 보건과학대학 방사선학과)
  • Published : 2015.10.26

Abstract

Quantitative Evaluation of Image Quality using Automatic Exposure Control This it leads, the relationship of control and image between of photographing condition, DR proper use method proposing. Chest phantom acquires Chest PA images which it follows in conditional change, Image evaluation factor (SNR (Signal to Noise Ratio), CNR (Contrast to Noise Ratio), PSNR (Peak Signal to Ratio), RMS (Root Mean Square)). It excepted RMS price, SNR, CNR, PSNR the case which uses AEC it came out being high The fact that the quality of image is better there was a case which does not use AEC. but Price it was slight. Image qualitative it was deficient in AEC use presence. Through this, the overall situation and most efficient use of radiation workers is better suited. But Passive AEC use would be appropriate for use rather than proactive AEC.

자동노출제어장치 (AEC : Auto Exposure Control)를 사용하는 경우와 사용하지 않았을 경우, 관전압과 관전류 설정에 따른 영상의 정량적인 평가를 통하여 최적의 영상을 찾고자 한다. 이를 통해 AEC 사용유무 및 촬영조건의 조절과 영상간의 관계를 파악하여, DR의 올바른 사용방법을 제안하고자 한다. 흉부팬텀으로 조건변화에 따른 Chest PA 영상을 획득한 뒤 영상평가인자인 신호 대 잡음비 (SNR), 대조도 대 잡음비 (CNR), 최대신호 대 잡음비 (PSNR), 평균 제곱근 (RMS)을 이용하여 영상평가를 실시한다. 관전압이 감소, 관전류가 증가함에 따라 SNR, CNR, PSNR은 증가하는 경향을 보였으며, RMS는 감소하는 경향을 나타내었다. 또한 고관전압, 저관전류 영역에서는 AEC를 사용하지 않았을 경우가 AEC를 사용하는 경우에 비해 SNR, PSNR이 높게 나타났으며, RMS는 낮게 나타났다. 그리고 저관전압, 고관전류영역에서는 AEC가 SNR, PSNR이 높게 나타났으며, RMS는 낮게 나타났다. 촬영조건의 변화와 AEC의 사용유무에 따라 영상의 질에서 차이를 확인할 수 있었다. 전체적인 상황과 방사선 종사자의 효율적인 업무를 위해 AEC 사용이 더 적합함을 알 수 있었다. 그러나 종사자의 무조건적인 AEC의 사용보다는 환자의 상태 및 촬영 자세 등 여러 조건을 고려한 능동적인 AEC 사용자세로 최적의 영상의 질을 얻도록 권장한다.