Comparison of TIA with ELISA for circulating antibody detection in clonorchiasis

간흡충증에 있어서 항체검출을 위한 Enzyme-linked Immunosorbent Assay와 Thin Layer Immunoassay의 비교

  • Published : 1983.06.01

Abstract

A comparison was made of a new serological method, thin layer immunoassay (TIA), and an established method, enzyme-linked immunosorbent assay (ELISA), in the detection and quantiacation of antibodies in clonorchiasis. Saline extract of Iyophilized Clonorchis sinensis adult worm was used as antigen, and TIA by the method of Elwing et at. (1976) and ELISA by Voller et at. (1974) were performed. Using sera from known clonorchiasis cases,100% of the sera tested were Positive by TIA and 88.35 by ELISA. TIA produced false positive results in 14 out of 36 cases, which were 10 amoebiasis cases, 16 paragonimiasis cases and 10 healthy controls. ELISA. however, produced a small number of false positives, 7 out of 55 cases. There was correlation between Immunoglobulin G level in sera and ELISA value (correlation coefficient, 0.69), whereas no correlation between Immunoglobulin G level and TIA result. The Performance of TIA and ELISA was not correlated in the results using homologous antigen.

간흡충증 환자 혈청에서 Thin layer immunoassay에 의하여 항체를 정량적으로 측정하여 감수성과 특이성을 알아보고, 효소표식면역 법(ELISA)에 의한 성적과 비교하였다. 냉동 건조한 간흡충 성충의 생리식염수 추출물을 항원으로 사용하였고, TIA는 Elwing등(1976)의 방법으로, 효소표식면 역법은 Veiler등(1974)의 방법에 의해 실시하였다. 1. 간홉충중 환자 60명의 혈청에서 TIA zone 크기의 평균은 4.14mm였고, 3mm 이상을 양성으로 판정하였을 때 감수성은 100%, 특이성은 61.1%였다. 2. 효소표식면역 법에 의하면 ELISA치 0.8이상을 양성으로 판정하였을 때 감수성은 88.3%, 특이성은 87.3%였다. 3. 혈청내 IgG 농도는 TIA zone 크기와 상관이 없었으나 좌소표식면역법에 의한 ELISA치와는 상관이 있어 그 계수는 0.69였다. 4. TIA zone 크기는 효소표식면역 법에 의한 ELISA치와 그 분포에 있어 상관이 없었다.