DOI QR코드

DOI QR Code

Drift Self-compensating Type Flux-meter for Automatic Magnetic Flux Measurement

  • Ga, E.M. (Hannam University) ;
  • Son, D. (Hannam University) ;
  • Bak, J.G. (Korea Basic Science Institute) ;
  • Lee, S.G. (Korea Basic Science Institute)
  • Published : 2003.12.01

Abstract

In magnetic flux measurement, output voltage drift of electronic integrator is an essential problem. In this work, we have developed a new kind of Miller type integrator using a sample and hold amplifier. Input bias current was measured and this value was hold in the sample and hold amplifier, after that input bias current of Miller integrator was compensated automatically using the value which holds in the sample and hold amplifier. Developed flux-meter shows the drift of flux-meter are smaller than 10$^{-5}$ Wb/min in full scale of 10$^{-2}$, and we could also measure multi-channel magnetic flux simultaneously.

Keywords

References

  1. Rev. Sci. Instrum. v.68 no.1 E.T.Strait (et al.) https://doi.org/10.1063/1.1147835
  2. 17th IEEE/NPSS Symposium on Fusion Engineering K.Kurihara;Y.kawamata
  3. Elsevier Science B.V. Fusion Technology S.G.Lee (et al.)

Cited by

  1. Active LR Integrator Circuit for Drift-Free Magnetoelastic Transducer vol.133, pp.3, 2018, https://doi.org/10.12693/APhysPolA.133.745