This paper studied the influences of the main design parameter-the expansion angle and the material properties of the self-expansion anterior cruciate ligament fixation device on the contact condition with the bone and the initial stability of the device. Using finite element analysis, the stress distributions of the ring part of the device and the wall of the bone tunnel were calculated. And the micro-migration of the device by the pull-out force was calculated. From the analysis results, it was found that when designing the self-expansion type anterior cruciate ligament fixation device, it is desirable to use the material having higher Young's modulus and to design the fixation device that all wedges uniformly maintain contact with bone to obtain initial stability after operation.

키워드

참고문헌

  1. Hefti, F. L., Kress, A., Fasel, J., and Morscher, E. W., 1992, "Healing of the Transected Anterior Cruciate Ligament in the Rabbit," Journal of bone and joint surgery, Vol. 73-A, No.3, pp. 373-383.
  2. Aglietti, P., Buzzi, R., Zaccherotti G., and De Biase, D. P., 1994, "Patellar tendon versus doubled semi-tendinosus and gracilis tendons for anterior cruciate ligament reconstruction," The American Journal of Sports Medicine, Vol. 22, No.2, pp. 211-218.
  3. Brand, J., Weiler, A., Cabom, D. N., Brown, C. H. Jr., and Johnson, D. L., 2000, "Graft Fixation in Cruciate Ligament Reconstruction," The American Journal of Sports Medicine, Vol. 28, No.5, pp. 761-774.
  4. A. Weiler, Hoffmann, R. F. G., Bail, H. J., Rehm, O., and Sudkamp, N. P., 2002, "Tendon healing in a bone tunnel: Part II: Histological analysis after biodegradable interference fit fixation in a model of anterior cruciate ligament reconstruction," Arthroscopy: The Journal of Arthroscopic & Related Surgery, Vol. 18, No.2, pp. 124-125 .
  5. Robbe, R. and Paletta, G. A. Jr., 2004, "Soft-Tissue Graft in Anterior Cruciate Ligament Reconstruction," Operative Techniques in Sports Medicine, Vol. 12, No.4, pp. 188-194.
  6. Shelboume, K. D. and Nitz, P., 1990, "Accerated rehabilitation after anterior cruciate ligament reconstruction," The American Journal of Sports Medicine, Vol. 18, No.3, pp. 292-299.
  7. Magen, H. E., Howell, S. M., and Hull, M. L., 1999, "Structural properties of six tibial fixation methods for anterior cruciate ligament soft tissue grafts," The American Journal of Sports Medicine, Vol. 27, No. 1, pp. 35-43.
  8. Park, J. S., Seo, T. I., Bae, J. S., and Yoon, G. S., 2006, "Development and Evaluation for the MicroMovement Structure of Interspinous," The Korea Society of Machine Tool Engineers, Vol. 15, No.3, pp. 127-132.
  9. Kim, J. D., Oh, C. Y, and Kim, C. S., 2008, "The influence on the contact condition and initial fixation stability of the main design parameters of a selfexpansion type anterior cruciate ligament fixation device," Journal of Mechanical Science and Technology, Vol. 22, No. 12, pp. 2301-2309.
  10. Park, J. B. and Bronzino, J. D., 2003, Biomaterials : Principles and applications, CRC Press, Boca Raton, pp. 1-9. 327
  11. Pacific Research Labs, Vashon Island, WA, 2006, "Third generation composite femur," .
  12. Kayabasi, O. and Erzincanli, F., 2006, "Finite element modelling and analysis of a new cemented hip prosthesis," Advances in engineering software, Vol. 37, No.7, pp. 477-483.
  13. Song, J., Gelin, J. C, Barriere, T., and Liu, B., 2006, "Experiments and numerical modelling of solid state sintering for 316L stainless steel components," Journal of materials processing technology, Vol. 177, No. 113, pp. 352-355.
  14. Norman, T. L., Thyagarajan, G., Saligrama, V. C., Gruen T. A., and Blaha, J. D., 2001, "Stem surface roughness alters creep induced subsidence and 'taper-lock' in a cemented femoral hip prothesis," Journal of Biomechanics, Vol. 34, No. 10, pp. 1325-1333.
  15. Zhong, Z. C, Wei, S. H., Wang, J. P., Feng, C. K., Chen, C. S., and Yu, C. H., "Finite element analysis of the lumber spine with a new cage using a topology optimization method," Medical Engineering & Physics, Vol. 28, No. 1, pp. 90-98.