Abstract
To evaluate the usefulness of newly designed percutaneous bone biopsy needle for pumpkin's specimen collection. We manufactured three types of biopsy needle with different internal processing which were made of 10 mm-diameter acrylic material. We made the conventional type (Ct) similar to the clinical type then compared the test group. The type a (Ta) made 1 cm-length internal processing from the distal, type b (Tb) made taper, and type c (Tc) made internal processing like spiral configuration. We performed 20 times biopsy to get an 10 mm length specimen from pumpkin's surface and evaluated the success rate of the biopsy, length of the specimen, and determine internal processing type of the most suitable biopsy needle (ANOVA test). Success rates of Ct, Ta, Tb, and Tc were 55%, 80%, 90%, and 100%, respectively. The lengths of the specimen of Ct, Ta, Tb, and Tc were $5.6{\pm}1.1\;mm$, $5.9{\pm}0.87\;mm$, $3.9{\pm}0.77\;mm$, and $9.4{\pm}0.54\;mm$, respectively. All groups were statistically significant (p < 0.05) except the group between Ct and Ta (p = 0.28). Newly designed bone biopsy needle seems to be useful for obtaining enough specimen. Tc may be more effective than other types.
호박을 대상으로 새로 개발된 골생검용 바늘의 유용성을 평가하고자 본 실험을 하였으며, 골생검용 바늘은 10 mm 직경의 아크릴 재료를 사용하여 내부가공을 달리 하여 3가지 형태로 제작하였다. 임상에서 사용되는 구조와 유사한 구조로 재현하여(Conventional type, Ct) 실험군과 비교하였다. Type a(Ta)는 바늘의 원위부 1cm 부분에 내부가공을 하였고, Type b(Tb)는 원위부로부터 동일 길이까지 점점 가늘게 제작하였고, Type c(Tc)는 내부에 동일 길이로 나사산을 제작하였다. 호박 표면으로부터 10 mm되는 부분까지의 생검 채취를 목적으로 20회를 시행하였으며, 시술 성공률, 채취된 표본의 길이 등을 측정하여 가장 적절한 생검용 바늘의 구조를 평가하고자 하였다. 그 결과 생검술의 성공률은 각각 Ct는 55%, Ta는 80%, Tb는 90%, Tc는 100%이었으며, 생검된 표본의 양은 Ct는 $5.6{\pm}1.1\;mm$, Ta는 $5.9{\pm}0.87\;mm$, Tb는 $3.9{\pm}0.77\;mm$, Tc는 $9.4{\pm}0.54\;mm$이었다. 그룹별 통계적 유 의성은 Ct와 Ta(p = 0.28)를 제외한 모든 그룹에서 통계적 의미를 가졌다(p < 0.05). 따라서 새로 개발된 골생검용 바늘은 생검술에 유용하게 사용될 것으로 사료되며, 그중 내부의 나선 가공이된 Tc형의 바늘이 골 생검에 가장 적절한 구조라 예상된다.