Effects of Interferents in Blood on the Blood Glucose Measurements by Using Polyphosphazene Diagnostic Membranes

포스파젠 고분자 진단막을 이용한 혈당 측정 시 혈액 중 장애성분이 미치는 영향

  • Kwon, Suk-Ky (Department of Biological and Chemical Engineering, Hongik University)
  • 권석기 (홍익대학교 과학기술대학 바이오화학공학과)
  • Received : 2013.08.08
  • Accepted : 2013.08.26
  • Published : 2013.08.30

Abstract

Diagnostic membranes which were made of polyphosphazenes were prepared for the measurements of blood glucose concentration. The effects of possible interferents in human blood on the glucose concentration measurements were examined. Above those interferents, ascorbic acid (AA) showed 3~9% higher K/S values compared to the standard plasma solution, tolbutamide (TA) showed 11~13% lower K/S values than those of plasma, triglycerides and bovine serum albumin (BSA) showed 20~25% lower K/S values than those of plasma. However, most of the interferents except above materials did not show any serious effects on the blood glucose measurements.

혈액 속의 글루코우즈의 농도를 측정하기 위해 포스파젠 고분자로 만들어진 진단막을 제조하였다. 혈액 속에 존재할 가능성이 있는 장애성분들이 글루코우즈의 농도측정에 미치는 영향을 조사하였다. 그러한 물질들 중, ascorbic acid (AA)는 기준 플라즈마 용액보다 3~9% 정도 높은 K/S 결과치를 나타내었고, tolbutamide (TA)는 기준용액보다 11~13% 정도 낮은 K/S 결과치를 나타내었고, triglycerides와 bovine serum albumin (BSA)은 기준용액 보다 20~25% 정도 낮은 K/S 결과치를 나타내었다. 그러나 위의 물질들을 제외하고는 대부분의 장애성분들이 포스파젠 고분자 진단막을 이용한 혈당측정에 큰 영향을 미치지 않는다는 것을 보여 주었다.

Keywords

References

  1. T. Asano, "Diabetic Food and Life Style", Shufu- To-Seikatsusha Co., Tokyo (2005).
  2. S. Kim, "Life Guide for Diabetes Control", Hyoseong, Seoul (1993).
  3. J. Kim, "Diabetes : Diagnostics and Control", Ohsung, Seoul (1997).
  4. Home Health Management Research Institute, "Diabetes", Kumyoung, Seoul (1992).
  5. T. Asano and T. Norioka, "Diabetes", Nihon Bungeisha, Tokyo (2000).
  6. K. Huh, "Diabetic Health", D and C Media, Seoul (2006).
  7. S. Kwon, "Preparation of Polyphosphazene Diagnostic Membranes for Blood Glucose Measurements", Membrane Journal, 24, 4 (2011).
  8. S. Kwon, "Effects of Testing and Storage Environments on the Blood Glucose Measurements by Using Phosphazene Diagnostic Membranes", Membrane Journal, 22, 5 (2012).
  9. S. Kwon, "A Study on the Preparation of Polyurethane Diagnostic Membrane for Urine Glucose Test", J. Kor. Ind. & Eng. Chem., 5, 6 (1994).
  10. S. Kwon and M. Choi, "The Preparation of Polyacrylonitrile Diagnostic Membranes for Blood Glucose Measurements(2) : Effects of Blood constituens on the Measurements of Glucose Concentration", Membrane Journal, 18, 4 (2008).
  11. R. P. Back, "Biosensor Technology", Marcel Dekker, New York (1990).
  12. S. Kwon, "Basic Studies on the Preparation of Diagnostic Membranes by Using Multi-Layered Gelatin Films to Measure Blood Glucose Level of Diabetics", Membrane Journal, 8, 1 (1998).
  13. S. Kwon and B. Lee, "Studies on the Multi- Layered Gelatin Diagnostic Membranes for Diabetes (1) : Effects of Temperature and Humidity on the Diffusion-Controlled Rates of Glucose", Membrane Journal, 9, 2 (1992).
  14. M. Gordon, "Polymer Membranes", Springer-Verlag, New York (1985)
  15. R. E. Kesting, "Synthetic Polymeric Membranes", Wiley-Interscience, New York (1985).
  16. S. Kwon, "Studies on the Muti-Layered Gelatin Diagnostic Membranes for Diabetes(2) : Effects of Interferents in Blood on the Diffusion-Controlled Rates of Glucose", Membrane Journal, 9, 4 (1999).
  17. S. Kwon, "Studies on the Preparation of Diagnostic Membranes for Blood Glucose Measurements (2) : Effects of Temperature on the Measurements of Glucose Concentration", Hongik Industrial Technology, 13 (2003).
  18. H. R. Allcock and S. Kwon, "Covalent Linkages of Proteins to Surface-Modified Poly(organophosphazenes) : Immobilization of Glucose-6-Phosphate Dehydrogenase and Trypsin", Macromolecules, 19, 1502 (1986). https://doi.org/10.1021/ma00160a002
  19. S. Kwon and C. Jun, "Studies on Polyphosphazenes- Bound Wittig Reactions", J. Kor. Ind. & Eng. Chem, 5, 5 (1994).
  20. H. R. Allcock, M. Gebura, S. Kwon, and T. X. Neenan, "Amphiphilic Polyphosphazenes as Membrane Materials : Influence of Side Group on Radiation Crosslinking, Semipermeability, and Surface Morphology", Biomaterials, 9, 500 (1988). https://doi.org/10.1016/0142-9612(88)90045-2
  21. H. R. Allcock, and S. Kwon, "Hydrophilic Polyphosphazenes as Hydrogels : Radiation Crosslinking and Hydrogel Characteristics of Poly[bis(methoxyethoxy) phosphazene]", Biomaterials, 7, 6 (1986).
  22. S. Kwon, "Studies on the preparation of Diagnostic Membranes fir Blood Glucose Measurements (1) : Model Experiments by Using Microporous Polyurethane Membrane", Hongik Industrial Technology, 12 (2002).
  23. S. Kwon and J. Yu, "The Preparation of Polyacrylonitrile Diagnostic Membranes for Blood Glucose Measurements(3) : Effects of Storage Environments on the Measurements of Glucose Concentration", Membrane Journal, 19, 3 (2009).
  24. S. Kwon and B. Lee, "A Study on the Preparation of Metal-Ion Separation Membrane with Hydrophilic Polyphosphazene", J. Kor. Ind. & Eng. Chem, 10, 3 (1999).
  25. S. Kwon, "Preparation of Polyurethane Diagnostic Membranes for Blood Glucose Measurements(6) : Effects of Hematocrit on the Measurements of Glucose Concentration", Membrane Journal, 20, 3 (2010).
  26. S. Kwon, "Studies on the Preparation of Diagnostic Membranes for Blood Glucose Measurements (4) : Effects of Additives in Blood on the Measurements of Glucose Concentration", Hongik Industrial Technology, 15 (2005).
  27. S. Kwon, I. Park, and D. Yoon, "Studies on the Preparation of Polyurethane Diagnostic Membranes for Blood Glucose Measurements(5) : Effects of Temperature and Humidity on the Measurements of Glucose Concentration", Membrane Journal, 17, 1 (2007).
  28. Y. Kang, "Life Science Encyclopedia", Academy Books, Seoul (2008).
  29. Korean Nursing Society, "Nursing Encyclopedia", Korean Sajeonyeonkusa, Seoul (1996).