Performance Assessment on Temperature Calibration Capability of the Calibration Laboratories Using High-Precision Platinum Resistance Thermometers

고정밀 백금저항온도계를 이용한 교정기관의 온도교정능력 수행평가

  • Gam, Kee Sool (Center for Thermometry, Div. of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Lee, Young Hee (Center for Thermometry, Div. of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Yang, Inseok (Center for Thermometry, Div. of Physical Metrology, Korea Research Institute of Standards and Science)
  • 감기술 (한국표준과학연구원 기반표준본부 온도센터) ;
  • 이영희 (한국표준과학연구원 기반표준본부 온도센터) ;
  • 양인석 (한국표준과학연구원 기반표준본부 온도센터)
  • Received : 2013.08.13
  • Accepted : 2013.09.10
  • Published : 2013.11.29

Abstract

Calibration capabilities for thermometer calibration by comparison method were assessed using high-precision industrial platinum resistance thermometers (IPRT). It was found in the performance assessment that out of 31 laboratories who participated, 28 laboratories resulted magnitude of En number less than 1 at every calibration points they submitted results in the range from 50 to $500^{\circ}C$. The results of about 75% of the laboratories showed the difference from the assigned values less than 1/10 of the tolerance level of the class B IPRT. This indicates that the participating calibration laboratories performed with satisfactory level that was enough to calibrate IPRTs to significant precision. The sensors used in this work were manufactured and chosen by the criteria of long-term instability less than 4 mK and hysteresis less than 8 mK in the temperature range used in this work. Furthermore, the change in the resistance of the sensors in the calibration temperature range were less than the uncertainty of the calibration, 25 mK (k=2).

Keywords

References

  1. Korean Laboratory accreditation scheme, KOLASG-004 : 2012.
  2. Korean Laboratory accreditation scheme, KOLASR-001 : 2012.
  3. Korean Laboratory accreditation scheme, KOLASR-003 : 2012.
  4. Korean Agency for Technology and Standards, KS Q ISO/IEC 17043 : 2011, "Conformity assessment - General requirements for proficiency testing", 2011.
  5. K. S. Gam, S. H. Yoo, S. M. Kim, and I. S. Lee, "Experimental method and evaluation of the calibration capability for the national calibration centers using the platinum resistance temperature sensors", J. Sensor Sci. & Tech., Vol. 14, No. 4, pp. 231-236, 2005. https://doi.org/10.5369/JSST.2005.14.4.231
  6. K. S. Gam, Y.-G. Kim, and S. G. Kim, "Measurement of traceability error for calibration services center using type S thermocouples", J. Sensor Sci. & Tech., Vol. 4, No. 1, pp. 43-50, 1995.
  7. K. S. Gam, I. Yang, and Y. G. Kim, "Thermal hysteresis in thin-film platinum resistance thermometers", Int. J. Thermophys, Vol. 32, No. 11, pp . 2388-2396, 2011. https://doi.org/10.1007/s10765-011-1041-8
  8. International Electrotechnical Commission, IEC 60751 ed2.0, "Industrial platinum resistance thermometers and platinum temperature sensors", 2008.
  9. I. Yang, Y.-G. Kim, K. S. Gam, and Y. H. Lee, "Improved interpolating equation for industrial platinum resistance thermometer", J. Sensor Sci. & Tech., Vol. 21, No. 2, pp. 109-113, 2012. https://doi.org/10.5369/JSST.2012.21.2.109