Spatial dose distribution and exposure dose during lumbar lateral test

요추 측면 검사 시 공간선량 분포와 피폭선량

  • Received : 2014.01.05
  • Accepted : 2014.02.24
  • Published : 2014.03.31

Abstract

During the lateral x-ray testing of lumbar, in order to obtain the optimal image for diagnosis and to minimize the exposure dose, a glass dosimeter and spatial dose measuring meter was used to measure and evaluate the exposure dose and spatial dose distribution of each organs. The exposure dose of the organs have increased as they were closer to the X-ray tube and when the radiation field was completely opened, the exposure dose was increased. In addition, scattered rays have increased as the distance got closer to the subject and with the distance of more than 200cm, 95% of scattered rays was reduced. Such results can anticipate the exposure dose of patients during the lumbar x-ray test in the future and it can be proposed as a data for determining the testing methods and expected to be widely used as an important basic data for reducing the medical exposure dose.

요추 측면 X선 검사 시 피폭선량을 최소화 하고 진단에 최적의 영상을 얻기 위하여 유리선량계와 공간선량 측정계를 이용하여 장기별 피폭선량과 공간선량분포를 측정 평가 하였다. X선관과 가까울수록 장기의 피폭선량이 증가하였으며 조사야를 완전히 열었을 때 피폭선량이 증가되었다. 또한 피사체와 거리가 가까울수록 산란선이 증가하였으며 200 cm 이상 거리를 두면 95% 이상의 산란선이 감소되었다. 이 결과는 향후 요추 X선 검사 시 환자의 피폭선량을 예측하고 검사 방법을 결정하는데 자료로 제시되어 의료피폭선량을 감소 하는데 중요한 기초자료로 많은 활용이 있을 것으로 기대된다.

Keywords

References

  1. C. G. Kim, "Exposure dose Reduction using Pb banding of own manufacturing", The Society of Digital Policy & Management, 11(6), pp. 269-273. 2013.
  2. C. G. Kim, "Radiation dose reduction effectiveness of a male gonadal shield during 128-MDCT using Glass Detector", The Society of Digital Policy & Management, 11(7), pp. 237-242. 2013. https://doi.org/10.14400/JDPM.2013.11.12.237
  3. "Special medical device health insurance requisition present condition", Health Korea News, 2010.
  4. ICRP, 2007. "Recommendations of the International Commission on Radiological Protection", ICRP Publication 103, 2007.
  5. S.G. Kweon, Y.M. Kim, C.G. Kim, H.Y. Yang, S.Y. Yang, E.O. Han, "Radiation Control & Magement", pp. 79-86, 2012.
  6. C. J. KOTRE, "X-ray absorptiometry of the breast using mammographic exposure factors", The Britsh Journal of Radiology, 83 pp. 515-523, 2010. https://doi.org/10.1259/bjr/68799159
  7. M, Y. Kim, H. S. Kim "The evaluation of Radiation Dose by Exposure Method in Digital Magnification Mammography", The Korea Journal of Radiological Science, 35(4) pp. 293-298, 2012.
  8. M, SAUDERS, A. BUDDEN, F. MACIVER, M. TEUNIS, H. WARREN-FORWARD, Dose implications of fluoroscopy-guided positioning for lumbar spine spine examinations prior to acquiring plain film radiographs, The Britsh Journal of Radiology, 78 pp. 130-134, 2005. https://doi.org/10.1259/bjr/50195548
  9. K. K. L. FUNG, W. B. GILBOY, "Anode heel effect" on patient dose in lumbar spine radiography, The Britsh Journal of Radiology, 73 pp. 531-536, 2000. https://doi.org/10.1259/bjr.73.869.10884750
  10. D. ZWEERS, J. GELEIJNS, N.J.M. AARTS, L.J. HARDAM, J.S. LAMERIS, F.W. SCHULTZ, L.J. SCHULTZ KOOL, Patient and staff radiation dose in fluoroscopy-guided TIPS procedures and dose reduction, using delicated fluoroscopy exposure settings, The Britsh Journal of Radiology, 71 pp. 672-676, 1998. https://doi.org/10.1259/bjr.71.846.9849393
  11. C. G. Kim, The evaluation of the radiation dose and the image quality during MDCT using Glass Rod Detector, The Korea Society of Digital Policy and Magement journal, 10(2) pp. 249-254, 2012.
  12. M. Y. Jung, D. C. Kweon, S. I. Kwon, "Effectiveness of Bismuth Shield to Reduce Eye Lens Radiation Dose Using the Photoluminescence Dosimetry in Computed Tomography", The Korea Society of Radiological Sceience journal, Vol. 32. No. 3, pp. 307-312, 2009.
  13. P. K. Cho, "Distribution of the Scatter Ray in Fluoroscopy X-ray Room", JKCA, Vol. 11., No. 10, pp. 349-354, 2011.
  14. S. C. Kim, D. Y. Won, C. H. Park, K. R. Dong, "Bone Densitometry Spatial Distribution of Radiation Dose Measurement", Korean Journal Digit Imaging Med, Vol. 13. No. 2, pp. 59-62, 2011.
  15. S. K. Kim, S. H. Son, "The Measurement and Analysis by Free Space Scatter Dose Distribution of Diagnostic Radiology Mobile Examination Area", Journal of the Korean Society for Digital Imaging in Medicine, Vol. 11(1)., No. 05, pp. 5-13, 2009.
  16. R. H. Clarke and J. Valentin, The History of ICRP and the Evolution of Its Policies, in ICRP Publication 109,Ann. ICRP 39(1), 2008.
  17. M. Y. Jung, D. C. Kweon, S. I. Kwon, "Effectiveness of Bismuth Shield to Reduce Eye Lens Radiation Dose Using the Photoluminescence Dosimetry inComputed Tomography", The Korea Society of Radiological Sceience journal, Vol. 32. No. 3, pp. 307-312, 2009.
  18. S. K. Kim, S. H. Son, "The Measurement and Analysis by Free Space Scatter Dose Distribution of Diagnostic Radiology Mobile Examination Area", Journal of the Korean Society for Digital Imaging in Medicine, Vol. 11(1)., No. 05, pp. 5-13, 2009.