DOI QR코드

DOI QR Code

맞춤형 지대주 각도에 따른 지지골의 유한요소 분석

Finite Element Analysis of Supporting Bone according to Custom Abutment Angles

  • Nam, Min-Gyeong (Catholic University of Pusan, Department of Dental Laboratory Technology, Masan University) ;
  • Kim, Nam-Sic (Catholic University of Pusan, Department of Dental Laboratory Technology, Masan University)
  • 투고 : 2015.08.24
  • 심사 : 2015.09.18
  • 발행 : 2015.09.30

초록

Purpose: The purpose of this study is a finite element analysis of supporting bone according to custom abutment angle. Methods: Implant fixture was selected with a diameter of 4 mm and the length of 13 mm. The fixture and abutment was designed by a combination of the abutment screw clamping force to produce a custom abutment model of $0^{\circ}$, $15^{\circ}$, $25^{\circ}$ and $35^{\circ}$. The loading condition of 176 N was applied to the lingual surface of the crown, near to the incisor edge, and horizontal load. An oblique load of $90^{\circ}$ was applied long axis of the implant fixture analyze the stress of supporting bone. Results: The result of mechanical analysis was observed that the supporting bone stress analysis of the horizontal load, the von Mises stress values (MPa) are given in the order of TH00 (432.6) > TH25 (418.0) > TH15 (417.4) > TH35 (415.8), the oblique load, the von Mises stress values are given in the order of TO00 (459.3) > TO15 (399.6) > TO25 (374.8) > TO35 (343.4) Conclusion: The $35^{\circ}$ abutment over the current clinical tolerance limits will be available for clinical application.

키워드

참고문헌

  1. Ahmad Y, Alireza M, Edwin AM. Implantabutment interface: A comparison of the ultimate force to failure among narrowdiameter implant systems. The journal of prosthetic dentistry, 112(5), 136-142, 2014. https://doi.org/10.1016/j.prosdent.2014.01.020
  2. Burak Y, Louai G., Jeremy DS, Nancy LC, Edwin A M. Load to failure of different titanium abutments for an internal hexagon implant. The journal of prosthetic dentistry, In Press, Corrected Proof, Available online, 2015.
  3. Christopher B. Marchack. A custom titanium abutment for the anterior single-tooth implant. The journal of Prosthetic Dentistry, 76(3), 288-291, 1996. https://doi.org/10.1016/S0022-3913(96)90173-0
  4. Kao HC, Gung YW, Chung TF, Hsu ML. The influence of abutment angulation on micromotion level for immediately loaded dental implants: a 3-D finite element analysis. Int J Oral Maxillofac Implants, 23(4), 623-630, 2008.
  5. Kebin T, Jiang C, Lilin H, Jin Y, Wenxiu H, Dong W. Angled abutments result in increased or decreased stress on surrounding bone of single-unit dental implants: A finite element analysis. Medical Engineering & Physics, 24(10), 2012.
  6. Lee MG. A 3-dimensional Finite Element Analysis of Stress Distribution in the Supporting Bone by Diameters of Dental Implant Fixture. The Journal of Korean academy of dental technology, 26(1), 69-76, 2004.
  7. Lin CL, Wang JC, Ramp LC, Liu PR. Biomaterial response of implant systems placed in the maxillary posterior region under various conditions of angulation, bone density and loading. Int J oral maxillofac implant, 23(1), 56-64, 2008.
  8. Nam MG. Finite Element Stress Analysis in Supporting Bone and Abutment Screw according to the Custom Abutment Angle and Material. Catholic University of Pusan Graduate School, Thesis, 2015.
  9. Tamar B, Raphael P, David S. The influence of abutment angulation on strains and stresses along the implant/bone interface : Comparison between two experimental techniques. The journal of prosthetic dentistry, 79(3), 328-334, 1998. https://doi.org/10.1016/S0022-3913(98)70246-X
  10. Tind W, Wenhe L, Ning D, Chunbo T. Design of a custom angled abutment for dental implants using computer-aided design and nonlinear finite element analysis. Journal of Biomechanics, 46(10), 1941-1946, 2010.
  11. Xavier ES, Jason A, John M, Robert L. Effect of abutment angulation on the strain on the bone around an implant in the anterior maxilla: Afinite element sutdy. The journal of prosthetic dentistry, 92(2), 85-92, 2007.