Evaluation of the Space Scattered Dose According to the Position of the Radiation Workers in Mammography Room

유방촬영 시 방사선관계종사자의 위치에 따른 공간선량평가

  • Lee, Dong-Yeon (Department of Radiation Oncology, Dongnam Inst. of Radiological & Medical Science) ;
  • Lee, Jin-Soo (Department of Radiology, Inje University Heaundae Paik Hospital)
  • 이동연 (동남권원자력의학원 방사선종양학과) ;
  • 이진수 (인제대학교 해운대백병원 영상의학과)
  • Received : 2016.06.10
  • Accepted : 2016.09.09
  • Published : 2016.09.30

Abstract

This study was conducted to evaluate the dose of the space to the controller located within the mammography room conducted a research on ways to the reduction exposure to the radiation workers. Results, the dose of 6.18 mGy/year was measured when there is no difference in the hilar area of the controller position, the dose of 2.35E-11 mGy/year was measured when installing the Shielding door. In addition, when the direction of the X-ray tube anode be heading this direction controller, low average level measured was 0.30 mGy/year. Based on this study, the mammography should be considered when installing the anode and cathod directions. And, by installing the shielding door, it must be able to completely separate shooting space and control room. This is the best way radiation protection method in radiation workers.

본 연구는 유방촬영실 내에서 제어기가 위치하는 공간에 대한 선량평가를 실시하여 방사선관계종사자에 대한 피폭을 저감화 할 수 있는 방안에 대한 연구를 진행하였다. 연구결과 제어기가 위치하는 공간에 차폐문이 없을 경우 6.18 mGy/year의 선량이 측정되었으며, 차폐문을 설치할 경우 2.35E-11 mGy/year의 선량이 측정되었다. 또한 X선관의 양극 방향이 제어기를 향하고 있을 경우, 평균 0.30 mGy/year 정도 낮게 측정되었다. 이를 바탕으로 유방촬영 장치 설치 시 양극과 음극 방향을 고려하여야 하며, 촬영공간과 제어공간을 완벽히 분리할 수 있는 차폐문을 설치하는 것이 방사선관계종사자에게 있어서 가장 좋은 방사선 방어 방법으로 판단된다.

Keywords

References

  1. Seyoung Jang, Wanggyun Oh, Jongbae Park, Gyehwan Jin : The energy spectrum and phantom image quality according to mammography target-filter combinations, Journal of the Korean Society of Radiology, Vol. 7, No. 1, 51-55, 2013 https://doi.org/10.7742/jksr.2013.7.1.051
  2. BC Centre for Disease Control : Guideline for Determining the X-ray Shielding Requirements for a Mammographic Facility, 2014
  3. Inja Lee, Kyeyeon Park, Sungsoo Kim : Research on the Actual Condition of Mammography and Space Scattered Dose in Mammography Room, Korean Society of Radiological Science, Vol. 29, No. 1, 21-28, 2006
  4. Noriah Jamal, Siti Selina Abdul Hamid, Humairah Samad Cheung et al. : A survey on mean grandular dose from full-field digital mommography system, operate using Mo/Mo and W/Rh target/filter combination, IAEA, INIS, Vol. 45, Issue 43, 2013
  5. M.B. Tavakoli, N. Kolghi, Z. Shahi, Z. Shaneh : Evaluation of the relation between breast grandular absorbed dose and radiographic quality in mammography, Iran. J. Radiant. Vol. 6, No. 2, 77-82, 2008
  6. Su Jeong Yu, Sang Wook Lim, Sun Young Ma, et al. : Invivo dosimetry for mammography with and without lead apron using the glass dosimeters, Progress in Medical Physics, Vol. 26, No. 2, 93-98, 2015 https://doi.org/10.14316/pmp.2015.26.2.93
  7. Los Alamos National Laboratory : User's manual version 2.5.0, LANL, 2005
  8. Korea Food and Drug Administration : A study on guidance for test and method of radiation protection, 2012
  9. Jin Young Park, Sung Min Baek : A study on the actual conditions and characteristic of mammographic units in some area, Journal of the Korean Society of Radiology, Vol. 6, No. 2, 121-127, 2012 https://doi.org/10.7742/jksr.2012.6.2.121
  10. A.A. Mowlavi : X-ray spectra calculation for different target-filter of mammograms using MCNP code, Iran. J. Radiant. Vol. 3, No. 3, 129-133, 2005
  11. Seong Kwan Choi : Relationship between the distribution of space doses in the X-ray rooms and the inverse square law of distance, The Korea Contents Association, Vol. 13, No. 8, 301-307, 20