Dosimetric Validation of the Acuros XB Advanced Dose Calculation Algorithm for Volumetric Modulated Arc Therapy Plans

  • Park, So-Yeon (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Park, Jong Min (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Choi, Chang Heon (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Chun, Minsoo (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Kim, Jung-in (Department of Radiation Oncology, Seoul National University Hospital)
  • Received : 2016.11.22
  • Accepted : 2016.12.07
  • Published : 2016.12.31

Abstract

Acuros XB advanced dose calculation algorithm (AXB, Varian Medical Systems, Palo Alto, CA) has been released recently and provided the advantages of speed and accuracy for dose calculation. For clinical use, it is important to investigate the dosimetric performance of AXB compared to the calculation algorithm of the previous version, Anisotropic Analytical Algorithm (AAA, Varian Medical Systems, Palo Alto, CA). Ten volumetric modulated arc therapy (VMAT) plans for each of the following cases were included: head and neck (H&N), prostate, spine, and lung. The spine and lung cases were treated with stereotactic body radiation therapy (SBRT) technique. For all cases, the dose distributions were calculated using AAA and two dose reporting modes in AXB (dose-to-water, $AXB_w$, and dose-to-medium, $AXB_m$) with same plan parameters. For dosimetric evaluation, the dose-volumetric parameters were calculated for each planning target volume (PTV) and interested normal organs. The differences between AAA and AXB were statistically calculated with paired t-test. As a general trend, $AXB_w$ and $AXB_m$ showed dose underestimation as compared with AAA, which did not exceed within -3.5% and -4.5%, respectively. The maximum dose of PTV calculated by $AXB_w$ and $AXB_m$ was tended to be overestimated with the relative dose difference ranged from 1.6% to 4.6% for all cases. The absolute mean values of the relative dose differences were $1.1{\pm}1.2%$ and $2.0{\pm}1.2%$ when comparing between AAA and $AXB_w$, and AAA and $AXB_m$, respectively. For almost dose-volumetric parameters of PTV, the relative dose differences are statistically significant while there are no statistical significance for normal tissues. Both $AXB_w$ and $AXB_m$ was tended to underestimate dose for PTV and normal tissues compared to AAA. For analyzing two dose reporting modes in AXB, the dose distribution calculated by $AXB_w$ was similar to those of AAA when comparing the dose distributions between AAA and $AXB_m$.

Keywords

References

  1. Otto K: Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 35, 310-317 (2008)
  2. Park JM, Kim IH, Ye SJ, Kim K: Evaluation of treatment plans using various treatment techniques for the radiotherapy of cutaneous Kaposi's sarcoma developed on the skin of feet. J Appl Clin Med Phys 15, 4970 (2014)
  3. Park JM, Kim K, Chie EK, Choi CH, Ye SJ, Ha SW: RapidArc vs intensity-modulated radiation therapy for hepatocellular carcinoma: a comparative planning study. Br J Radiol 85, e323-329 (2012) https://doi.org/10.1259/bjr/19088580
  4. Park JM, Wu HG, Kim JH, Carlson JN, Kim K: The effect of MLC speed and acceleration on the plan delivery accuracy of VMAT. Br J Radiol 88, 20140698 (2015) https://doi.org/10.1259/bjr.20140698
  5. Jin H, Jesseph FB and Ahmad S: A comparison study of volumetric modulated Arc therapy quality assurances using portal dosimetry and MapCHECK 2. Prog Med Phys. 25, 7 (2014)
  6. Mattes MD, Lee JC, Elnaiem S, Guirguis A, Ikoro NC, Ashamalla H: A predictive model to guide management of the overlap region between target volume and organs at risk in prostate cancer volumetric modulated arc therapy. Radiat Oncol J 32, 23-30 (2014) https://doi.org/10.3857/roj.2014.32.1.23
  7. Borgers C: Complexity of Monte Carlo and deterministic dose-calculation methods. Phys Med Biol 43, 517-528 (1998) https://doi.org/10.1088/0031-9155/43/3/004
  8. Gifford KA, Horton JL, Wareing TA, Failla G, Mourtada F: Comparison of a finite-element multigroup discrete-ordinates code with Monte Carlo for radiotherapy calculations. Phys Med Biol 51, 2253-2265 (2006) https://doi.org/10.1088/0031-9155/51/9/010
  9. Vassiliev ON, Wareing TA, McGhee J, Failla G, Salehpour MR, Mourtada F: Validation of a new grid-based Boltzmann equation solver for dose calculation in radiotherapy with photon beams. Phys Med Biol 55, 581-598 (2010) https://doi.org/10.1088/0031-9155/55/3/002
  10. Han T, Mikell JK, Salehpour M, Mourtada F: Dosimetric comparison of Acuros XB deterministic radiation transport method with Monte Carlo and model-based convolution methods in heterogeneous media. Med Phys 38, 2651-2664 (2011) https://doi.org/10.1118/1.3582690
  11. Bush K, Gagne IM, Zavgorodni S, Ansbacher W, Beckham W: Dosimetric validation of Acuros XB with Monte Carlo methods for photon dose calculations. Med Phys 38, 2208-2221 (2011) https://doi.org/10.1118/1.3567146
  12. Fogliata A, Nicolini G, Clivio A, Vanetti E, Mancosu P, Cozzi L: Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm: fundamental characterization in water. Phys Med Biol 56, 1879-1904 (2011) https://doi.org/10.1088/0031-9155/56/6/022
  13. Lloyd SA and Ansbacher W: Evaluation of an analytic linear Boltzmann transport equation solver for high-density inhomogeneities. Med Phys 40, 011707 (2013)
  14. Kron T, Clivio A, Vanetti E, Nicolini G, Cramb J, Lonski P et al.: Small field segments surrounded by large areas only shielded by a multileaf collimator: comparison of experiments and dose calculation. Med Phys 39, 7480-7489 (2012) https://doi.org/10.1118/1.4762564
  15. Han T, Mourtada F, Kisling K, Mikell J, Followill D, Howell R: Experimental validation of deterministic Acuros XB algorithm for IMRT and VMAT dose calculations with the Radiological Physics Center's head and neck phantom. Med Phys 39, 2193-2202 (2012) https://doi.org/10.1118/1.3692180
  16. Han T, Followill D, Mikell J, Repchak R, Molineu A, Howell R et al.: Dosimetric impact of Acuros XB deterministic radiation transport algorithm for heterogeneous dose calculation in lung cancer. Med Phys 40, 051710 (2013) https://doi.org/10.1118/1.4802216
  17. Kathirvel M, Subramanian S, Clivio A, Arun G, Fogliata A, Nicolini G et al.: Critical appraisal of the accuracy of Acuros-XB and Anisotropic Analytical Algorithm compared to measurement and calculations with the compass system in the delivery of RapidArc clinical plans. Radiation Oncology 8 (2013)
  18. Huang B, Wu L, Lin P, Chen C: Dose calculation of Acuros XB and Anisotropic Analytical Algorithm in lung stereotactic body radiotherapy treatment with flattening filter free beams and the potential role of calculation grid size. Radiat Oncol 10, 53 (2015) https://doi.org/10.1186/s13014-015-0357-0
  19. Hirata K, Nakamura M, Yoshimura M, Mukumoto N, Nakata M, Ito H et al.: Dosimetric evaluation of the Acuros XB algorithm for a 4 MV photon beam in head and neck intensity-modulated radiation therapy. J Appl Clin Med Phys 16, 5222 (2015)
  20. Rana S, Rogers K, Lee T, Reed D, Biggs C: Dosimetric impact of Acuros XB dose calculation algorithm in prostate cancer treatment using RapidArc. J Can Res Ther 9, 430-435 (2013) https://doi.org/10.4103/0973-1482.119328
  21. Zhen H, Hrycushko B, Lee H, Timmerman R, Pompos A, Stojadinovic S et al.: Dosimetric comparison of Acuros XB with collapsed cone convolution/superposition and anisotropic analytic algorithm for stereotactic ablative radiotherapy of thoracic spinal metastases. J Appl Clin Med Phys 16, 5493 (2015)
  22. Liu HH: Dm rather than Dw should be used in Monte Carlo treatment planning. For the proposition. Med Phys 29, 922-923 (2002) https://doi.org/10.1118/1.1473137
  23. Walters BR, Kramer R, Kawrakow I: Dose to medium versus dose to water as an estimator of dose to sensitive skeletal tissue. Phys Med Biol 55, 4535-4546 (2010) https://doi.org/10.1088/0031-9155/55/16/S08
  24. Rana S and Pokharel S: Dose-to-medium vs. dose-to-water: Dosimetric evaluation of dose reporting modes in Acuros XB for prostate, lung and breast cancer. Int J Cancer Ther Oncol 2, 020421 (2014) https://doi.org/10.14319/ijcto.0204.21
  25. Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L: Dosimetric evaluation of Acuros XB Advanced Dose Calculation algorithm in heterogeneous media. Radiat Oncol 6, 82 (2011) https://doi.org/10.1186/1748-717X-6-82
  26. Kan MW, Leung LH, Yu PK: Dosimetric impact of using the Acuros XB algorithm for intensity modulated radiation therapy and RapidArc planning in nasopharyngeal carcinomas. Int J Radiat Oncol Biol Phys 85, e73-80 (2013) https://doi.org/10.1016/j.ijrobp.2012.08.031
  27. Kathirvel M, Subramanian S, Clivio A, Arun G, Fogliata A, Nicolini G et al.: Critical appraisal of the accuracy of Acuros-XB and Anisotropic Analytical Algorithm compared to measurement and calculations with the compass system in the delivery of RapidArc clinical plans. Radiat Oncol 8, 140 (2013) https://doi.org/10.1186/1748-717X-8-140
  28. Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L: Critical appraisal of Acuros XB and Anisotropic Analytic Algorithm dose calculation in advanced non-small-cell lung cancer treatments. Int J Radiat Oncol Biol Phys 83, 1587-1595 (2012) https://doi.org/10.1016/j.ijrobp.2011.10.078
  29. Robinson D: Inhomogeneity correction and the analytic anisotropic algorithm. J Appl Clin Med Phys 9, 112-122 (2008) https://doi.org/10.1120/jacmp.v9i2.2786
  30. Liu HW, Nugent Z, Clayton R, Dunscombe P, Lau H, Khan R: Clinical impact of using the deterministic patient dose calculation algorithm Acuros XB for lung stereotactic body radiation therapy. Acta Oncologica 53, 324-329 (2014) https://doi.org/10.3109/0284186X.2013.822552
  31. Fogliata A, Vanetti E, Albers D, Brink C, Clivio A, Knoos T et al.: On the dosimetric behaviour of photon dose calculation algorithms in the presence of simple geometric heterogeneities: comparison with Monte Carlo calculations. Phys Med Biol 52, 1363-1385 (2007) https://doi.org/10.1088/0031-9155/52/5/011
  32. Beam Configuration Reference Guide. Varian medical system (2009)

Cited by

  1. Comparison of dosimetric and radiobiological parameters on plans for prostate stereotactic body radiotherapy using an endorectal balloon for different dose-calculation algorithms and delivery-beam mod vol.70, pp.4, 2016, https://doi.org/10.3938/jkps.70.424
  2. Feasibility of a 3D-printed anthropomorphic patient-specific head phantom for patient-specific quality assurance of intensity-modulated radiotherapy vol.12, pp.7, 2016, https://doi.org/10.1371/journal.pone.0181560
  3. Calculation of absorbed dose in radiotherapy by solution of the linear Boltzmann transport equations vol.64, pp.2, 2019, https://doi.org/10.1088/1361-6560/aaf0e2