Evaluation of the clinical accuracy of six portable blood glucose meters in dogs

  • Shin, Min-Keun (Veterinary Teaching Hospital, College of Veterinary Medicine, Chungbuk National University) ;
  • Kim, Hakhyun (Veterinary Teaching Hospital, College of Veterinary Medicine, Chungbuk National University) ;
  • Yun, Taesik (Veterinary Teaching Hospital, College of Veterinary Medicine, Chungbuk National University) ;
  • Kang, Ji-Houn (Western Veterinary Medical Center) ;
  • Kang, Byeong-Teck (Veterinary Teaching Hospital, College of Veterinary Medicine, Chungbuk National University)
  • Received : 2020.03.19
  • Accepted : 2020.07.10
  • Published : 2020.09.30

Abstract

Portable blood glucose meters (PBGMs) are widely used because of their practicality. However, the accuracy of PBGMs has frequently been questioned. The objectives of this study were to evaluate factors that might interfere with measurements made using PBGMs, and to assess the clinical utility of 6 PBGMs. The glucose concentrations measured using the PBGMs were compared with those obtained using a reference method. The agreement between the measured values was assessed using Spearman correlation analysis, Passing-Bablok regression analysis, Bland-Altman plots, and consensus error grid analysis. Mann-Whitney and Kruskal-Wallis tests were performed to identify the parameters affecting glucose measurement. The results indicated that all of the PBGMs tested perform adequately for use in veterinary practice. In most cases, measurements made using PBGM corresponded well with the blood glucose values obtained using the reference method. Error grid analysis revealed that most of the PBGM values fell within zones A and B. However, some measurements of blood glucose concentrations < 80 mg/dL fell into zone C. PCV, and triglyceride and total protein concentration, significantly affected the output of some of the PBGMs. Therefore, clinicians should be aware of the characteristics of the PBGM that they use.

Keywords

References

  1. Wess G, Reusch C. Evaluation of five portable blood glucose meters for use in dogs. J Am Vet Med Assoc 2000;216:203-209. https://doi.org/10.2460/javma.2000.216.203
  2. Cohen TA, Nelson RW, Kass PH, Christopher MM, Feldman EC. Evaluation of six portable blood glucose meters for measuring blood glucose concentration in dogs. J Am Vet Med Assoc 2009;235:276-280. https://doi.org/10.2460/javma.235.3.276
  3. Domori A, Sunahara A, Tateno M, Miyama TS, Setoguchi A, Endo Y. The clinical utility of two human portable blood glucose meters in canine and feline practice. Vet Clin Pathol 2014;43:55-62. https://doi.org/10.1111/vcp.12115
  4. Tauk BS, Drobatz KJ, Wallace KA, Hess RS. Correlation between glucose concentrations in serum, plasma, and whole blood measured by a point-of-care glucometer and serum glucose concentration measured by an automated biochemical analyzer for canine and feline blood samples. J Am Vet Med Assoc 2015;246:1327-1333. https://doi.org/10.2460/javma.246.12.1327
  5. Paul AE, Shiel RE, Juvet F, Mooney CT, Mansfield CS. Effect of hematocrit on accuracy of two point-of-care glucometers for use in dogs. Am J Vet Res 2011;72:1204-1208. https://doi.org/10.2460/ajvr.72.9.1204
  6. Lane SL, Koenig A, Brainard BM. Formulation and validation of a predictive model to correct blood glucose concentrations obtained with a veterinary point-of-care glucometer in hemodiluted and hemoconcentrated canine blood samples. J Am Vet Med Assoc 2015;246:307-312. https://doi.org/10.2460/javma.246.3.307
  7. Johnson BM, Fry MM, Flatland B, Kirk CA. Comparison of a human portable blood glucose meter, veterinary portable blood glucose meter, and automated chemistry analyzer for measurement of blood glucose concentrations in dogs. J Am Vet Med Assoc 2009;235:1309-1313. https://doi.org/10.2460/javma.235.11.1309
  8. Brito-Casillas Y, Figueirinhas P, Wiebe JC, Lopez-Rios L, Perez-Barreto D, Melian C, Wagner AM. ISO-based assessment of accuracy and precision of glucose meters in dogs. J Vet Intern Med 2014;28:1405-1413. https://doi.org/10.1111/jvim.12397
  9. Neeley WE. Simple automated determination of serum or plasma glucose by a hexokinase-glucose-6 -phosphate dehydrogenase method. Clin Chem 1972;18:509-515. https://doi.org/10.1093/clinchem/18.6.509
  10. Cohn LA, McCaw DL, Tate DJ, Johnson JC. Assessment of five portable blood glucose meters, a point-of-care analyzer, and color test strips for measuring blood glucose concentration in dogs. J Am Vet Med Assoc 2000;216:198-202. https://doi.org/10.2460/javma.2000.216.198
  11. Brunner GA, Ellmerer M, Sendlhofer G, Wutte A, Trajanoski Z, Schaupp L, Quehenberger F, Wach P, Krejs GJ, Pieber TR. Validation of home blood glucose meters with respect to clinical and analytical approaches. Diabetes Care 1998;21:585-590. https://doi.org/10.2337/diacare.21.4.585
  12. Trajanoski Z, Brunner GA, Gfrerer RJ, Wach P, Pieber TR. Accuracy of home blood glucose meters during hypoglycemia. Diabetes Care 1996;19:1412-1415. https://doi.org/10.2337/diacare.19.12.1412
  13. Chan JC, Wong RY, Cheung CK, Lam P, Chow CC, Yeung VT, Kan EC, Loo KM, Mong MY, Cockram CS. Accuracy, precision and user-acceptability of self blood glucose monitoring machines. Diabetes Res Clin Pract 1997;36:91-104. https://doi.org/10.1016/S0168-8227(97)00036-3
  14. Ginsberg BH. Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol 2009;3:903-913. https://doi.org/10.1177/193229680900300438
  15. Rebel A, Rice MA, Fahy BG. Accuracy of point-of-care glucose measurements. J Diabetes Sci Technol 2012;6:396-411. https://doi.org/10.1177/193229681200600228
  16. Gerber KL, Freeman KP. ASVCP guidelines: quality assurance for portable blood glucose meter (glucometer) use in veterinary medicine. Vet Clin Pathol 2016;45:10-27. https://doi.org/10.1111/vcp.12310
  17. Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 1987;10:622-628. https://doi.org/10.2337/diacare.10.5.622
  18. Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care 2000;23:1143-1148. https://doi.org/10.2337/diacare.23.8.1143
  19. Freckmann G, Schmid C, Baumstark A, Rutschmann M, Haug C, Heinemann L. Analytical performance requirements for systems for self-monitoring of blood glucose with focus on system accuracy: relevant differences among ISO 15197:2003, ISO 15197:2013, and current FDA recommendations. J Diabetes Sci Technol 2015;9:885-894. https://doi.org/10.1177/1932296815580160
  20. International Organization for Standardization. ISO 15197:2013. In Vitro Diagnostic Test Systems - Requirements for Blood-Glucose Monitoring Systems for Self-Testing in Managing Diabetes Mellitus. Geneva, Switzerland, International Organization for Standardization, 2013.
  21. Kilpatrick ES, Rumley AG, Myint H, Dominiczak MH, Small M. The effect of variations in haematocrit, mean cell volume and red blood cell count on reagent strip tests for glucose. Ann Clin Biochem 1993;30:485-487. https://doi.org/10.1177/000456329303000513
  22. Karon BS, Griesmann L, Scott R, Bryant SC, Dubois JA, Shirey TL, Presti S, Santrach PJ. Evaluation of the impact of hematocrit and other interference on the accuracy of hospitalbased glucose meters. Diabetes Technol Ther 2008;10:111-120. https://doi.org/10.1089/dia.2007.0257
  23. Tang Z, Lee JH, Louie RF, Kost GJ. Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing. Arch Pathol Lab Med 2000;124:1135-1140. https://doi.org/10.5858/2000-124-1135-EODHLO
  24. Wiener K. The effect of haematocrit on reagent strip tests for glucose. Diabet Med 1991;8:172-175. https://doi.org/10.1111/j.1464-5491.1991.tb01566.x
  25. Tonyushkina K, Nichols JH. Glucose meters: a review of technical challenges to obtaining accurate results. J Diabetes Sci Technol 2009;3:971-980. https://doi.org/10.1177/193229680900300446