디지털 방사선 적용을 위한 Hybrid 방사선 검출기의 Feasibility 연구

Feasibility study of Hybrid X-ray detecter for Digital X-ray imaging application

  • 최장용 (인제대학교 의료영상연구소) ;
  • 박지군 (인제대학교 의료영상연구소) ;
  • 이채훈 (인제대학교 의료영상연구소) ;
  • 이규홍 (인제대학교 의료영상연구소) ;
  • 최흥국 (인제대학교 의료영상연구소) ;
  • 남상희 (인제대학교 의용공학과)
  • Choi, Jang-Yong (Medical Imaging Research center of Inje University) ;
  • Park, Ji-Koon (Medical Imaging Research center of Inje University) ;
  • Lee, Chae-Hun (Medical Imaging Research center of Inje University) ;
  • Lee, Kyu-Hong (Medical Imaging Research center of Inje University) ;
  • Choi, Heung-Kook (Medical Imaging Research center of Inje University) ;
  • Nam, Sang-Hee (Department of Biomedical Engineering, College of Biomedical Science and Engineering, Inje University)
  • 발행 : 2004.04.24

초록

In this study, the purpose is to verified the feasibility to develope Hybrid x-ray detector in order to resolve problems of direct and indirect x-ray detectors. The properties of X-ray detector depend on absorption of X-ray, charge generation by x-ray photon, leakage current. In this study, CdS was used as photoconductor, and $Y_2O_2S:Tb$ as x-ray phosphor was formed on CdS in order to embody Hybrid structure. And Screen printing was used to form Muli-layer. Characteristics of this specimen were analyzed by using SEM, and XRD. And Photoluminescence spectrum of $Y_2O_2S:Tb$, leakage current, with respect to applied voltages, output charge with respect to applied voltages, and X-ray sensitivity were measured. Also, linearity with respect to dose was measured. Leakage current was similar with direct digital x-ray detector, but sensitivity of the hybrid structure is much better than the single-layer structure.