DOI QR코드

DOI QR Code

Development of Solenoid RF coil for 4.7 T Magnetic Resonance Velocimeter to Improve Resolution

해상도 향상을 위한 4.7 T 자기공명유속계 용 솔레노이드 RF 코일 개발

  • 양병권 (한양대학교 융합기계공학과) ;
  • 조지현 (한국기초과학지원연구원 자기공명연구단) ;
  • 송시몬 (한양대학교 융합기계공학과)
  • Received : 2016.07.28
  • Accepted : 2016.08.19
  • Published : 2016.08.31

Abstract

Magnetic resonance velocimeter (MRV) is a powerful tool to non-invasively measure the velocity of a fluid flow in various fields ranging from medicine to engineering. However, since the demands for accurate measurement in the solid/liquid interface for cardiovascular diseases and porous media increase, the improvement of spatial resolution is required. In this study, a solenoid RF coil is developed for high spatial resolution measurement. The signal-to-noise ratio in solenoid RF coil is increased seventeen times better than that in commercial coil. Moreover, the velocity distribution of Hagen-Poiseuille flow is measured with in-plane resolution of $36{\mu}m$ by $36{\mu}m$ and the accuracy of the measured velocity is compared with theoretical distribution of the laminar flow. Flow rate calculated by MRV is estimated with the flow rate injected by syringe pump.

Keywords

References

  1. Elkins, C. J. and Alley, M. T., 2007, "Magnetic resonance velocimetry: applications of magnetic resonance imaging in the measurement of fluid motion," Experiments in Fluids, Vol. 43(6), pp. 823-858. https://doi.org/10.1007/s00348-007-0383-2
  2. Markl, M., Wegent, F., Zech, F., Bauer, S., Strecker, C., Schumacher, M., Weiller, C., Hennig, J. and Harloff, A., 2010, "In vivo wall shear stress distribution in the carotid artery effect of bifurcation geometry, internal carotid artery stenosis, and recanalization therapy," Circulation Cardiovascular Imaging, Vol. 3(6), pp.647-655. https://doi.org/10.1161/CIRCIMAGING.110.958504
  3. Lee, J., Yang, B., and Song, S., 2015, "3D Visualization of Flow Inside Carpet Under Vacuum Cleaner Nozzle by Magnetic Resonance Velocimetry," The Korean Society of Mechanical Engineers, pp.149-152.
  4. Hoult, D. I. and Richards, R. E., 1976, "The signal-to-noise ratio of the nuclear magnetic resonance experiment," Journal of Magnetic Resonance, Vol. 24(1), pp.71-85. https://doi.org/10.1016/0022-2364(76)90233-X
  5. Elkins, C. J., Alley, M. T. and Eaton, J. K., 2009, "Three-dimensional magnetic resonance velocimetry measurements of turbulence quantities in complex flow," Experiments in Fluids, Vol. 46(2), pp.285-296. https://doi.org/10.1007/s00348-008-0559-4
  6. Sederman, A. and Gladden, L., 2001, "Magnetic resonance visualisation of single-and two-phase flow in porous media," Magnetic resonance imaging, Vol. 19(3), pp.339-343. https://doi.org/10.1016/S0730-725X(01)00246-6
  7. Edelhoff, D., Walczak, L., Henning, S., Weichert, F. and Suter, D., 2013, "High-resolution MRI velocimetry compared with numerical simulations," Journal of Magnetic Resonance, Vol. 235, pp.42-49. https://doi.org/10.1016/j.jmr.2013.07.002
  8. Peck, T. L., Magin, R. L., and Lauterbur, P. C., 1995, "Design and Analysis of Microcoils for NMR Microscopy," Journal of Magnetic Resonance Series B, Vol. 108(2), pp.114-124. https://doi.org/10.1006/jmrb.1995.1112
  9. Mispelter, J., Lupu, M., and Briguet, A., 2006, NMR Probeheads for Biophysical and Biomedical Experiments: Theoretical Principles & Practical Guidelines, Imperial College Press.
  10. Markl, M., 2005, "Velocity encoding and flow imaging," University Hospital Freiburg, Dept of Diagnostic Radiology.
  11. Moser, W. K., Kutter, E. C., Georgiadis, J. G., Buckius, R. O., Morris, H. D. and Torczynski, J. R., 2000, "Velocity measurements of flow through a step stenosis using Magnetic Resonance Imaging," Experiments in Fluids, Vol. 29(5), pp.438-447. https://doi.org/10.1007/s003480000110