Sung, Hong-Mo;Shin, Jae-Woo;Lee, Hyun-Sook;Hwang, Sung-Ho;Yoon, Young-Ro
181
The purpose of this study is to predict the defibrillation success of a ventricular Fibrillation ECG signal using time-frequency analysis. During CPR, coronary perfusion pressure and electrocardiogram were measured. Parameters extracted from time-frequency domain were served as predictor of resuscitation success. Time frequency distribution(TFD) of ECG signals was estimated from the smoothed pseudo Wigner-Ville distribution(SPWVD). Median frequency, peak frequency, 1/f slope, frequency band ratios$(2{\sim}4Hz,\;4{\sim}6Hz,\;6{\sim}8Hz,\;8{\sim}10Hz,\;10{\sim}12Hz,\;12{\sim}15Hz)$ were extracted from each TFD as function of time. Paired t-test was used to determine the differences in ROSC and non-ROSC groups. In the statistical results, we selected four significant parameters - median frequency, 1/f slope, $2{\sim}4Hz$ band ratio, $8{\sim}10Hz$ band ratio. We made an attempt to predict defibrillation success by combining features extracted from time frequency distribution. Independent t-test was used to determine the differences ROSC and non-ROSC groups. Consequently, we selected four significant parameters-median frequency, 1/f slope, $2{\sim}4Hz$ band ratio, $8{\sim}10Hz$ band ratio. The relationship between coronary perfusion pressure and ECG parameters was analyzed with linear regression analysis. R-square value was 55%. 1/f slope and $8{\sim}10Hz$ band ratio had the significant relationship with coronary perfusion pressure.