A Study on Safety Management Inspection of Diagnostic X-ray System

진단용 엑스선 장치의 안전관리 검사에 관한 연구

  • Lee, Hoo min (Department of Radiologic Technology, Dongnam Health University) ;
  • Kim, Hyeon ju (Department of Radiologic Technology, Dongnam Health University)
  • 이후민 (동남보건대학교 방사선과) ;
  • 김현주 (동남보건대학교 방사선과)
  • Received : 2018.11.23
  • Accepted : 2018.12.31
  • Published : 2018.12.30

Abstract

The purpose of this study is to compare the performance of X-ray generators installed in hospitals and universities and apply the quality control items of diagnostic X-ray generators to recognize the importance of periodic performance management. First, the reproducibility and linearity test results showed that the PAE of the reproducibility evaluation was high for the GX-650 devices that met the acceptance criteria in all the experimental conditions and lacked the periodic quality control. In the linearity evaluation, when the tube voltage was set to 100 kVp, It was measured to deviate from the error. In addition, it was found that the PAE in the low-accuracy evaluation results relative to an X-ray tube voltage and tube current of the device low occurrence frequency. The HVL experiment was included in all of the devices at the HVL by tube voltage. Therefore, it is necessary to recognize the importance of quality control of all devices rather than hospital and laboratory, and to manage the device performance by actively managing the device, and to establish a short - term quality control system like special medical devices.

본 연구는 병원 및 대학교 등에 설치된 X선 발생장치를 대상으로 진단용 방사선 발생장치의 품질관리 항목을 적용하여 성능 비교를 하여 주기적인 성능관리의 중요성을 인식시켜 성능유지 및 관리에 도움을 주고자 한다. 우선 재현성 및 직선성 실험결과 모든 실험조건에서 합격기준에 부합되었으며 주기적인 정도관리가 부족한 GX-650장치의 경우 재현성 평가의 백분율 오차가 높았으며 직선성 평가에서는 관전압을 100 kVp로 설정 시 0.105로 백분율 오차에서 벗어나게 측정되었다. 또한 관전압 및 관전류의 정확도 평가결과 상대적으로 X선 발생횟수가 낮은 장치에서 백분율 오차가 낮은 것을 알 수 있었다. 반가층 실험은 모든 장치에서 관전압 별 반가층 기준에 모두 포함되었다. 따라서 병원용, 실습용 구분보다는 모든 장치의 정도관리의 중요성을 인식하고 적극적으로 장치의 성능관리를 시행하여 장치를 유지 관리해야 할 것이며 특수의료장비처럼 짧은 주기의 정도관리 제도를 마련해야 할 것으로 사료된다.

Keywords

BSSHB5_2018_v12n7_887_f0001.png 이미지

Fig. 1. Image of Measuring device by rectification type.

BSSHB5_2018_v12n7_887_f0002.png 이미지

Fig. 2. Image of Multi_function meter.

BSSHB5_2018_v12n7_887_f0003.png 이미지

Fig. 3. Image of Measuring HVL.

Table 1. Table of Reproducibility evaluation of exposure dose

BSSHB5_2018_v12n7_887_t0001.png 이미지

Table 2. Table of Linearity evaluation of exposure dose

BSSHB5_2018_v12n7_887_t0002.png 이미지

Table 3. Table of Accuracy evaluation according to the change of tube voltage

BSSHB5_2018_v12n7_887_t0003.png 이미지

Table 4. Table of Accuracy evaluation according to the change of milli Ampere

BSSHB5_2018_v12n7_887_t0004.png 이미지

Table 5. Table of HVL according to the change of tube voltage

BSSHB5_2018_v12n7_887_t0005.png 이미지

References

  1. H. K. Min, "The study for performance, image quality and patient dose of various diagnostic X-ray generation," Journal of Inje University pp. 1-101, 2006.
  2. T. G. Kim, M. W. Cheon, Y. P. Park, "Comparison of Output and Radiation Quality of X-rays according to the Full-Wave Rectification Method and Dual-Voltage Rectification Method of an X-ray Generator," Journal of KIEEME, Vol. 23, No. 7, pp. 534-538, 2010.
  3. H. M. Lee, J. Y, H. J. Kim, "Effects of Contrast Improvement on High Voltage Rectification Type of X-ray Diagnostic Apparatus," Journal of Radiological Science and Technology, Vol. 37, No. 3, pp. 187-193, 2014.
  4. S. S. Kang, "Experimental control of diagnostic imaging device," Vol. 4, No. 3, pp.3-10, 2014.
  5. S. C. Kim, Y I. Kim, Y. H. kIM, "Diagnostic X-ray Equipment and Quality control," Shin Kwang Publishing Co, pp. 348-357, 2006.
  6. J. I. Choi, D. G. Na, H. H. Kim, Y. M. Shin, K. J. An, J. Y. Lee, "Quality Control of Medical Imaging," Journal of The Korean Society of Radiology, Vol. 50, No. 5, pp. 317-331, 2004. https://doi.org/10.3348/jkrs.2004.50.5.317
  7. I. G. You, C. H. Lim, S. H. Lee, M. K. Lee, "Performance Measurement of Diagnostic X Ray System," Journal of the Korean Society of Radiology, Vol. 6, No. 6, pp. 447-454, 2012. https://doi.org/10.7742/jksr.2012.6.6.447
  8. C. J. Gu, K. S Kim, B. K Kim, N. J. An, H. S. Kim, S. G. Kim, S. Im, "Improvement Way for Mobile X-ray Examinations by Rule Revision about Safety Management of Diagnosis Radiation Occurrence System," Journal of Radiological Science and Technology, Vol. 30, No. 1, pp. 53-59, 2007.
  9. J. G. Park, "Medical Clinics' quality Management of X-ray Units in Gyeongbuk Area," International Journal of contents Vol. 10, No. 9, pp. 267-275, 2010.
  10. J. D. Yeo, I. H. Ko, "A Study on Perception by Examines of the Radiology Departmentabout Exposure to Radioactivity," Journal of the Korean Society of Radiology, Vol. 7, No. 5, pp. 321-331, 2013. https://doi.org/10.7742/jksr.2013.7.5.321
  11. B S Kang, H H Park, K R Dong, H Y Yeo "Research for the Quality Control of General X-ray Systems," Journal of Advanced Engineering and Technology, Vol. 6, No. 4, pp. 31-35, 2013.